The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00749756579032), received between Dec 5, 2024 and Jan 16, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Failure to Seal” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 15, 2025Open, Classified

    The tubing may not ensure that the yankauer stays intact, and secure to the tubing / yankauer connection for a tight seal. This could result in inadequate suction and cause delay in patient care.

    Z-1174-2025 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Failure to Seal2
  • Fitting Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 16, 2025Failure to Seal; Failure to Seal

    A DISTRIBUTING FACILITY REPORTED, "THE CUSTOMER REPORTED THAT THEY HAVE HAD SEVERAL COMPLAINTS AND A SAFETY EVENT PUT IN THEIR SYSTEM ABOUT LOT# CNWKD04-06. THE GREEN END ON THIS LOT NUMBER IS WIDER AND NOT AS DEEP AS LOT NUMBER 19600504. IT WILL NOT STAY ATTACHED TO ANYTHING WHICH IS CAUSING ISSUES IN THEIR PICU".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21168136 ↗
  • MalfunctionJan 15, 2025Fitting Problem; Fitting Problem

    A USER FACILITY REPORTED, "TUBING DOES NOT FIT SECURELY ONTO THEIR ENDOSCOPES. THE 71-C506 DOES NOT STAY SECURE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21160088 ↗
  • MalfunctionJan 8, 2025Failure to Seal; Failure to Seal

    A USER FACILITY REPORTED, "THE SUCTION HANDLE WILL ONLY WORK/SEAL PROPERLY ON ONE END OF THE SUCTION TUBING. THE YANKEUR DOES NOT STAY ON AND THERE IS NO TIGHT SEAL. IT HAS BECOME A SAFETY ISSUE BECAUSE THEY NEED RAPID ACCESS TO RELIABLE SUCTION WITHOUT MISHAPS IN CONNECTIONS.".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21106358 ↗
  • MalfunctionDec 5, 2024Fitting Problem; Fitting Problem

    A USER FACILITY REPORTED, "SEVERAL COMPLAINTS AND A SAFETY EVENT PUT IN OUR SYSTEM ABOUT LOT# CNWKD04-06. THE GREEN END ON THIS LOT NUMBER IS WIDER AND NOT AS DEEP AS LOT NUMBER 19600504. IT WILL NOT STAY ATTACHED TO ANYTHING WHICH IS CAUSING ISSUES IN OUR PICU."

    Patient problems: Insufficient Information