Skip to content

DeRoyal 71-C520Recalls & reported problems

Deroyal Industries, Inc. · Tubing Suction Connector · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00749756579070), received between Jan 8, 2025 and Jan 8, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Seal” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 15, 2025Class IIOpen, Classified

    The tubing may not ensure that the yankauer stays intact, and secure to the tubing / yankauer connection for a tight seal. This could result in inadequate suction and cause delay in patient care.

    Z-1174-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Seal1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 8, 2025Failure to Seal; Failure to Seal

    A USER FACILITY REPORTED, "THE SUCTION HANDLE WILL ONLY WORK/SEAL PROPERLY ON ONE END OF THE SUCTION TUBING. THE YANKEUR DOES NOT STAY ON AND THERE IS NO TIGHT SEAL. IT HAS BECOME A SAFETY ISSUE BECAUSE THEY NEED RAPID ACCESS TO RELIABLE SUCTION WITHOUT MISHAPS IN CONNECTIONS."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21106360 ↗

FDA clearance and classification

The FDA regulates product code BYY (tube, aspirating, flexible, connecting) under 21 CFR 880.6740 as a Class II device. 21 CFR 880.6740 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the DeRoyal 71-C520

Has the DeRoyal 71-C520 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1174-2025 (Class II), began on Jan 15, 2025: The tubing may not ensure that the yankauer stays intact, and secure to the tubing / yankauer connection for a tight seal. This could result in inadequate suction and cause delay in patient care. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the DeRoyal 71-C520?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Jan 8, 2025. That report lists device problem “Failure to Seal” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the DeRoyal 71-C520?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the DeRoyal 71-C520 FDA approved?

Devices of this type (product code BYY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Deroyal Industries, Inc.

Who makes the DeRoyal 71-C520, and is it still sold?

The FDA's device database lists Deroyal Industries, Inc. as the company responsible for it (brand DeRoyal, model 71-C520). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other TUBE, ASPIRATING, FLEXIBLE, CONNECTING devices

All TUBE, ASPIRATING, FLEXIBLE, CONNECTING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page