Skip to content

MODULAR POWER POSITIONING SYSTEM 45T-12L-MAXXRecalls & reported problems

Motion Concepts Inc · DROP SEAT - MAXX RAIL · Class 2

All models of MODULAR POWER POSITIONING SYSTEM (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00754014710034), received between Jun 21, 2022 and Nov 14, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Device Handling Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Handling Problem1
  • Misassembly During Maintenance/Repair1
  • Self-Activation or Keying1

Patient problems

  • Bruise/Contusion1
  • Multiple Fractures1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2024Self-Activation or Keying; Self-Activation or Keying

    ON OCTOBER 15, 2024, MEDBLOC (DISTRIBUTOR) NOTIFIED MOTION CONCEPTS LP OF AN INCIDENT INVOLVING ONE OF OUR WHEELCHAIRS. THE PRODUCT HAD BEEN SOLD TO THE END-USER IN THE USA BY A DEALER, AMERICAN SEATING & MOBILITY IN (B)(6), WHO WAS ALSO RESPONSIBLE FOR ITS MAINTENANCE. THE DEALER HAD INSTALLED A THIRD-PARTY HEAD ARRAY ON THE WHEELCHAIR. ACCORDING TO THE END-USER, WHILE USING THE WHEELCHAIR IN AN OFFICE ROOM, THE HEAD ARRAY UNEXPECTEDLY ACTIVATED, CAUSING THE WHEELCHAIR TO ACCELERATE TOWARD A DESK. AS A RESULT, THE END-USER SUSTAINED BRUISING AND SCRATCHES ON THE LEFT LEG BUT DID NOT REQUIRE MEDICAL ASSISTANCE. THE SYSTEM HAD BEEN IN USE FOR APPROXIMATELY FOUR MONTHS AT THE TIME OF THE INCIDENT.

    Patient problems: Bruise/Contusion

    MAUDE report 20689312 ↗
  • InjuryJun 21, 2022Device Handling Problem; Misassembly During Maintenance/Repair; Device Handling Problem; Misassembly During Maintenance/Repair

    ON 26-MAY-2022, MOTION CONCEPTS LP WAS NOTIFIED BY (B)(4), THAT ONE OF THE DEALERS IN USA ((B)(4)) INFORMED THEM OF AN INCIDENT THAT WAS RELATED TO A MOTION CONCEPTS WHEELCHAIR. THE WHEELCHAIR WAS SOLD TO (B)(4) BUT, (B)(4) WAS MAINTAINING IT. THEY NOTIFIED THAT THE END-USER WAS DRIVING HIS WHEELCHAIR WHEN HE SUDDENLY FELL FORWARD AND WHEELCHAIR FELL ON TOP OF THE END-USER. THE END-USER FRACTURED BOTH HIS ANKLES AND HAD TO WEAR CAST FROM KNEE DOWN. THE DEALER CONFIRMED THAT THEY HAD ADJUSTED THE SEAT DEPTH FORWARD BY 2" ON 13-MAY-2022 AND THE END-USER WAS NOT WEARING SEAT BELT OR CHEST STRAP AS THE BUCKLE ON THE SEAT BELT WAS DAMAGED BEFORE THE INCIDENT.

    Patient problems: Multiple Fractures

    MAUDE report 14755855 ↗

FDA clearance and classification

Cleared through 510(k) K150574 on Jun 30, 2015 (decision: Substantially Equivalent). Submitted by Motion Concepts as “Modular Power Positioning System”. The FDA review took 116 days. FDA record ↗

The FDA regulates product code ITI (wheelchair, powered) under 21 CFR 890.3860 as a Class II device. 21 CFR 890.3860 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. FDA interprets “mobility to persons restricted to a sitting position” to mean the device type is intended to provide mobility to individuals who have mobility impairments and require an assistive device for mobility. NOTE: This type of device is not intended to climb stairs (see product code: IMK).

Common questions about the MODULAR POWER POSITIONING SYSTEM 45T-12L-MAXX

Has the MODULAR POWER POSITIONING SYSTEM 45T-12L-MAXX been recalled?

No FDA recall record in our data names this exact device identifier (DI 00754014710034), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the MODULAR POWER POSITIONING SYSTEM 45T-12L-MAXX?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jun 21, 2022 and Nov 14, 2024. In those reports, the most common specific device problems are “Device Handling Problem” (1), “Misassembly During Maintenance/Repair” (1) and “Self-Activation or Keying” (1); the most common patient problems are “Bruise/Contusion” (1) and “Multiple Fractures” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the MODULAR POWER POSITIONING SYSTEM 45T-12L-MAXX?

None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the MODULAR POWER POSITIONING SYSTEM 45T-12L-MAXX FDA approved?

It was cleared through 510(k) K150574 on Jun 30, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the MODULAR POWER POSITIONING SYSTEM 45T-12L-MAXX, and is it still sold?

The FDA's device database lists Motion Concepts Inc as the company responsible for it (brand MODULAR POWER POSITIONING SYSTEM, model 45T-12L-MAXX). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Wheelchair, Powered devices

All Wheelchair, Powered devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page