TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx TDXSP2-MCGRecalls & reported problems
Invacare Corporation · Class 2
All models of TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00841447100966), received between Jun 6, 2025 and Aug 7, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 30, 2022Class IIOpen, Classified
The battery wiring harness on affected wheelchairs may become disconnected during shipment from the factory or during wheelchair use traversing large obstacles or after a jolt or impact to the wheelchair. The disconnection of the battery harness results in loss of power to the wheelchair.
Z-1130-2022 · Root cause: Device Design · Recall details · FDA record ↗
- Nov 12, 2021Class IIOpen, Classified
Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury
Z-0445-2022 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method1
- Mechanical Problem1
- Unintended Collision1
Patient problems
- Multiple Fractures1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 7, 2026Unintended Collision; Improper or Incorrect Procedure or Method
The consumer was traveling in her vehicle when she was struck by another automobile. The impact forced her secured wheelchair to break loose from its tie-downs, propelling both the consumer and the wheelchair into the front of the vehicle. This incident resulted in bodily injury to the consumer which included rib fractures, a right femur fracture, a right ankle fracture, and bilateral tibia/fibula fractures. The consumer was admitted to the hospital for 3 months.
Patient problems: Multiple Fractures
MAUDE report 50178342 ↗MalfunctionJun 6, 2025Mechanical Problem; Mechanical Problem
ON A TDXSP2-MCG POWER WHEELCHAIR, THE BRAKE LEVERS ARE SPINNING IN CIRCLES AND NOT CATCHING/ENGAGING FROM DRIVE TO FREE WHEEL MODE. THE USER WENT OVER A BUMP AND SOMEHOW THE COUPLING SWITCHED TO FREE WHEEL WHILE IN THE MIDDLE OF THE STREET. A PERSON GOT OUT OF THEIR CAR AND HELPED PUSH THE USER IN THE WHEELCHAIR ONTO THE SIDEWALK. THERE WAS NO INJURY TO THE USER, HOWEVER, THEY EXPRESSED THEY WERE SLIGHTLY SCARED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22160032 ↗
Other Wheelchair, powered devices
- SmartDrive® MX2+Max Mobility, LLC · Wheelchair, Powered858 reports
- Permobil M3Permobil, Inc. · Wheelchair, Powered1 recall59 reports
- TDX SP2 with LiNX Electronics TDXSP2Invacare Corporation · Wheelchair, powered2 recalls0 reports
- Invacare® TDX SP2 with Captain's Seat TDXSP2VInvacare Corporation · Wheelchair, powered2 recalls0 reports
- Q6 Edge 3Pride Mobility Products Corp · Wheelchair, Powered4 deaths86 reports
- Q6 Edge 2.0 Q6 Edge 2.0 / Q6 Edge 2.0 with iLevelPride Mobility Products Corp · Wheelchair, Powered1 death83 reports
- Permobil M5Permobil, Inc. · Wheelchair, Powered1 recall7 reports
- Invacare® Storm Torque RX™ Power Wheelchair SRX-20MPInvacare Corporation · Wheelchair, powered1 recall6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page