Heli-FX™ EndoAnchor™ System HA-18-114 (UDI 00763000006648)Recalls & reported problems
Medtronic, Inc. · TAA APPLIER DEVICE US EIFU · Class 2 · Implant
All models of Heli-FX EndoAnchor System (13)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000006648), received between Nov 23, 2017 and Oct 24, 2019 — classified as 1 death, 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
1 other recall cover devices cleared under the same FDA submission (K171427, K182957); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Device Dislodged or Dislocated1
- Leak/Splash1
Patient problems
- Cardiac Arrest1
- Cyst(s)1
- Death1
- Device Embedded In Tissue or Plaque1
- Ischemia1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryOct 24, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
HELI-FX ENDOANCHORS WERE IMPLANTED PROPHYLACTICALLY IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 45 MM DIAMETER ABDOMINAL AORTIC ANEURYSM. A NON-MEDTRONIC STENT GRAFT SYSTEM WAS IMPLANTED ON THE SAME DAY. IT WAS REPORTED THAT A LYMPHOCELE WAS OBSERVED APPROXIMATELY 8 DAYS POST INDEX PROCEDURE. THE PATIENT WAS TREATED WITH INCISIONAL DEBRIDEMENT AND THE EVENT WAS RESOLVED. THE INVESTIGATOR ASSESSED THE EVENT AS NOT RELATED TO THE STUDY DEVICES OR PROCEDURE. THE SPONSOR ASSESSED THE EVENT AS PROBABLY RELATED TO THE STUDY PROCEDURE, NOT RELATED TO THE STUDY DEVICE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Patient problems: Cyst(s)
MAUDE report 9232715 ↗InjuryOct 24, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
HELI-FX ENDOANCHORS WERE IMPLANTED PROPHYLACTICALLY IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 45 MM DIAMETER ABDOMINAL AORTIC ANEURYSM. A NON-MEDTRONIC STENT GRAFT SYSTEM WAS IMPLANTED ON THE SAME DAY. IT WAS REPORTED THAT RIGHT LOWER EXTREMITY ISCHEMIA WAS OBSERVED ON THE DAY OF THE INDEX PROCEDURE. THE PATIENT WAS TREATED WITH A LOWER EXTREMITY ARTERIAL INTERVENTION AND THE EVENT WAS RESOLVED. THE INVESTIGATOR ASSESSED THE EVENT AS NOT RELATED TO THE STUDY DEVICES OR PROCEDURE. THE SPONSOR ASSESSED THE EVENT AS PROBABLY RELATED TO THE STUDY PROCEDURE, NOT RELATED TO THE STUDY DEVICE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Patient problems: Ischemia
MAUDE report 9232685 ↗MalfunctionMay 15, 2018Leak/Splash; Leak/Splash
A VALIANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 56MM DIAMETER THORACIC AORTIC ANEURYSM. IT WAS PLANNED THAT HELI-FX ENDOANCHORS WOULD ALSO BE USED PROPHYLACTICALLY IN THE PROCEDURE DUE TO CONCERN OVER A SHORT DISTAL LANDING ZONE. IT WAS REPORTED THAT DURING THE INDEX PROCEDURE THE FIRST VALIANT STENT GRAFT WAS PLACED IN A STRAIGHT SEGMENT OF THE THORACIC AORTA AND DEPLOYED WITHOUT ISSUE. A SECOND VALIANT STENT GRAFT WAS THEN DEPLOYED PROXIMAL TO THE CELIAC TRUNK WITHOUT ISSUE. A RELIANT BALLOON WAS USED TO SEAL THE PROXIMAL, OVERLAP AND DISTAL SEAL ZONES. THEN A TOTAL OF 8 ENDOANCHORS WERE PLACED IN THE DISTAL SEAL ZONE AND AN ANGIOGRAM WAS PREFORMED THAT SHOWED A TYPE IB ENDOLEAK. THE PHYSICIAN RE-BALLOONED THE DISTAL SEAL MORE AGGRESSIVELY AND THIS SHOWED THE TYPE IB ENDOLEAK WAS GREATLY REDUCED. THE PHYSICIAN DECIDED TO END THE PROCEDURE. AS PER THE PHYSICIAN, THE CAUSE OF THE EVENT WAS ANATOMY RELATED DUE TO THE SHORT DISTAL SEAL ZONE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT WILL BE MONITORED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7514282 ↗DeathNov 23, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated
A HELI-FX APPLIER AND ENDOANCHORS WERE USED PROPHYLACTICALLY AS AN ACCESSORY DEVICE FOR THE ENDOVASCULAR TREATMENT OF THE PATIENT WITH A VALIANT STENT GRAFT SYSTEM. IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, THE PHYSICIAN PLACED THE HELI- FX GUIDE IN THE APPROPRIATE LOCATION TO DEPLOY AN ENDOANCHOR IN THE RIGHT LATERAL WALL OF THE DESCENDING THORACIC AORTA, JUST ABOVE THE CELIAC. AFTER DEPLOYING THE ENDOANCHOR TO THE FIRST STAGE, IT WAS DETERMINED THAT IT WAS IN A GOOD POSITION. WHEN THE ENDOANCHOR WAS FULLY DEPLOYED, IT APPEARED TO "FLOAT" BEHIND THE EXISTING STENT GRAFT AND SETTLED IN THE DESCENDING ANEURYSM SAC BEHIND THE STENT GRAFT. A SECOND ENDOANCHOR WAS THEN PLACED BY THE PHYSICIAN IN A SIMILAR LOCATION AND IT ALSO APPEARED TO "FLOAT" THROUGH THE STENT GRAFT WALL TO A SIMILAR LOCATION. IT WAS NOTED THAT PLACEMENT OF THE ENDOANCHORS IN THE LEFT ASPECT OF THE STENT GRAFT WAS SUCCESSFUL. AS PER THE PHYSICIAN, THE CAUSE OF THE ENDOANCHOR EVENT WAS PRODUCT RELATED. IT WAS REPORTED THAT, 6 DAYS POST INDEX PROCEDURE, THE PATIENT HAD A CARDIAC EVENT DURING ROUTINE EXERCISE WITH THE REGISTERED NURSE AND DIED AFTER CPR. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED. IT WAS REPO
Patient problems: Cardiac Arrest; Death; Device Embedded In Tissue or Plaque
MAUDE report 7056725 ↗
FDA clearance and classification
Cleared through Special 510(k) K182957 on Nov 21, 2018 (decision: Substantially Equivalent). Submitted by Medtronic Vascular, Inc. as “Heli-FX EndoAnchor System”. The FDA review took 28 days. FDA record ↗
Cleared through Special 510(k) K171427 on Jun 13, 2017 (decision: Substantially Equivalent). Submitted by Medtronic Vascular, Inc. as “Heli-FX Applier, Heli-FX Guide, Ancillary EndoAnchor Cassette”. The FDA review took 29 days. FDA record ↗
The FDA regulates product code OTD (endovascular suturing system) under 21 CFR 870.3460 as a Class II device. 21 CFR 870.3460 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K171427, K182957. One submission often covers many models, so check the product and lot details in each record.
- Z-3227-2017 · Sep 13, 2017 · TerminatedMedtronic Aptus Heli-FX Thoracic EndoAnchor System Heli-FX Guide 42 mm, REF HG-18-90-42, STERILE EO, Rx Only The Heli-FX EndoAnchor System is intended to provide fixation and sea
Common questions about the Heli-FX™ EndoAnchor™ System HA-18-114
Has the Heli-FX™ EndoAnchor™ System HA-18-114 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000006648), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the Heli-FX™ EndoAnchor™ System HA-18-114?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Nov 23, 2017 and Oct 24, 2019. In those reports, the most common specific device problems are “Device Dislodged or Dislocated” (1) and “Leak/Splash” (1); the most common patient problems are “Cardiac Arrest” (1), “Cyst(s)” (1) and “Death” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Heli-FX™ EndoAnchor™ System HA-18-114?
1 report is classified as a death, 2 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Heli-FX™ EndoAnchor™ System HA-18-114 FDA approved?
It was cleared through Special 510(k) K171427 on Jun 13, 2017; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Heli-FX™ EndoAnchor™ System HA-18-114 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Heli-FX™ EndoAnchor™ System HA-18-114, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand Heli-FX™ EndoAnchor™ System, model HA-18-114). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endovascular suturing system devices
- Heli-FX™ EndoAnchor™ System SA-85Medtronic, Inc. · Endovascular suturing system6 deaths70 reports
- Heli-FX™ EndoAnchor™ System HG-18-90-22Medtronic, Inc. · Endovascular suturing system1 recall1 report
- APTUS Heli-FX® SA-85Medtronic, Inc. · Endovascular suturing system8 deaths50 reports
- Heli-FX™ EndoAnchor™ System HG-18-90-42Medtronic, Inc. · Endovascular suturing system1 recall0 reports
- Heli-FX™ EndoAnchor™ System SA-85Medtronic, Inc. · Endovascular suturing system4 deaths45 reports
- Heli-FX™ EndoAnchor™ System SA-85Medtronic, Inc. · Endovascular suturing system6 deaths41 reports
- Heli-FX™ EndoAnchor™ System SG-64Medtronic, Inc. · Endovascular suturing system18 reports
- APTUS Heli-FX® HG-16-62-28Medtronic Vascular, Inc. · Endovascular suturing system6 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page