Heli-FX™ EndoAnchor™ System HG-18-90-22Recalls & reported problems
Medtronic, Inc. · TAA GUIDE DEVICE 22 EIFU · Class 2 · Implant
All models of Heli-FX EndoAnchor System (13)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000006655), received between Sep 22, 2017 and Sep 22, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Sep 13, 2017Class IITerminated
It was determined that the deflection length indicated on the Guide catheter handle does not match the label on the box and sterile packaging for two lots.
Z-3227-2017 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Markings/Labelling Problem1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2017Device Markings/Labelling Problem; Device Markings/Labelling Problem
A HELI-FX GUIDE CATHETER WAS INSERTED INTO A PATIENT PROPHYLACTICALLY DURING THE IMPLANT OF A VALIANT STENT GRAFT SYSTEM IN THE THORACIC AORTA. IT WAS REPORTED THAT THE PHYSICIAN OPENED THE PACKAGE FOR HELI-FX GUIDE CATHETER AND NOTED THAT LABEL ON THE BOX AND STERILE PACKAGING WAS LABELLED AS THORACIC HG-18-90-22MM HOWEVER THE HELI-FX DEVICE CATHETER LABEL WAS 42MM. THE PHYSICIAN ELECTED TO USE THE DEVICE EVEN THOUGH THE LABELLING WAS DIFFERENT. IT WAS OBSERVED THAT THE DEVICE WOULD NOT DEFLECT AS MUCH AS THE PHYSICIAN INTENDED HOWEVER PHYSICIAN DID MANAGE TO PLACE A FEW OF THE HELI-FX APTUS ENDOANCHORS. A SECOND THORACIC HG-18-90-22MM, ALSO FROM THE SALES REP¿S TRUNK STOCK, WITH THE SAME LOT NUMBER WAS OPENED. HOWEVER, THE SAME LABELING ISSUE WAS ON THE DEVICE AND THE PHYSICIAN ELECTED TO NOT USE THIS DEVICE. THE HOSPITAL HAD A HELI-FX APTUS ABDOMINAL GUIDE (SG-64) 22MM AND THE PHYSICIAN WAS ABLE TO USE THIS DEVICE TO COMPLETE DEPLOYMENT OF THE HELI-FX APTUS ENDOANCHORS. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6885698 ↗
FDA clearance and classification
Cleared through Special 510(k) K182957 on Nov 21, 2018 (decision: Substantially Equivalent). Submitted by Medtronic Vascular, Inc. as “Heli-FX EndoAnchor System”. The FDA review took 28 days. FDA record ↗
Cleared through Special 510(k) K171427 on Jun 13, 2017 (decision: Substantially Equivalent). Submitted by Medtronic Vascular, Inc. as “Heli-FX Applier, Heli-FX Guide, Ancillary EndoAnchor Cassette”. The FDA review took 29 days. FDA record ↗
The FDA regulates product code OTD (endovascular suturing system) under 21 CFR 870.3460 as a Class II device. 21 CFR 870.3460 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Heli-FX™ EndoAnchor™ System HG-18-90-22
Has the Heli-FX™ EndoAnchor™ System HG-18-90-22 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-3227-2017 (Class II), began on Sep 13, 2017: It was determined that the deflection length indicated on the Guide catheter handle does not match the label on the box and sterile packaging for two lots. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Heli-FX™ EndoAnchor™ System HG-18-90-22?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Sep 22, 2017. That report lists device problem “Device Markings/Labelling Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Heli-FX™ EndoAnchor™ System HG-18-90-22?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Heli-FX™ EndoAnchor™ System HG-18-90-22 FDA approved?
It was cleared through Special 510(k) K171427 on Jun 13, 2017; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Heli-FX™ EndoAnchor™ System HG-18-90-22 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Heli-FX™ EndoAnchor™ System HG-18-90-22, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand Heli-FX™ EndoAnchor™ System, model HG-18-90-22). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endovascular suturing system devices
- Heli-FX™ EndoAnchor™ System SA-85Medtronic, Inc. · Endovascular suturing system6 deaths70 reports
- APTUS Heli-FX® SA-85Medtronic, Inc. · Endovascular suturing system8 deaths50 reports
- Heli-FX™ EndoAnchor™ System HG-18-90-42Medtronic, Inc. · Endovascular suturing system1 recall0 reports
- Heli-FX™ EndoAnchor™ System SA-85Medtronic, Inc. · Endovascular suturing system4 deaths45 reports
- Heli-FX™ EndoAnchor™ System SA-85Medtronic, Inc. · Endovascular suturing system6 deaths41 reports
- Heli-FX™ EndoAnchor™ System SG-64Medtronic, Inc. · Endovascular suturing system18 reports
- APTUS Heli-FX® HG-16-62-28Medtronic Vascular, Inc. · Endovascular suturing system6 reports
- Heli-FX™ EndoAnchor™ System HG-16-62-28Medtronic, Inc. · Endovascular suturing system5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page