Heli-FX EndoAnchor System recallZ-3227-2017 · Medtronic Vascular, Inc. · initiated Sep 13, 2017
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Vascular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It was determined that the deflection length indicated on the Guide catheter handle does not match the label on the box and sterile packaging for two lots.
FDA root cause category: Labeling Change Control
What the company is doing
The firm initiated their recall on September 13, 2017, by letter. The letter requested the consignee take the following actions: "1. Identify and quarantine all unused affected product as listed in your inventory. 2. Return all unused affected list product in your inventory to Medtronic. Contact Medtronic Customer Service at 1-800-854-3570 to initiate a product return and credit. Your local Medtronic Representative can assist you in the return and replacement of this product as necessary. 3. Complete the enclosed Customer Confirmation Certificate and fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality or scan and email to RS.CFQFCA@medtronic .com. This notice needs to be passed on to all those who need to be aware within your organization where the potentially affected devices have been transferred." For further questions, please call (508) 261-8000.
Recall details
- Recall number
- Z-3227-2017
- Status
- Terminated
- Initiated by the company
- Sep 13, 2017
- Recalling firm
- Medtronic Vascular, Inc.
- Quantity
- 20 units
- Product code
- OTD
- FDA submissions
- K171427
- Distribution
- US Distribution to the states of : AZ, CA, FL, GA, MA, MI, MN and NH.
Product description
Medtronic Aptus Heli-FX Thoracic EndoAnchor System Heli-FX Guide 42 mm, REF HG-18-90-42, STERILE EO, Rx Only The Heli-FX EndoAnchor System is intended to provide fixation and sealing between endovascular aortic grafts and the native artery. The Heli-FX System is indicated for use in patients who endovascular grafts have exhibited migration or endoleak, or are at risk for such complications, in whom augmented radial fixation and or sealing is required or regain or maintain adequate aneurysm exclusion.
Code information (models, lots, serial numbers)
UDI 00763000006679, Lot Number 0008674156 UDI 00763000006655, Lot Number 0008674157
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Heli-FX™ EndoAnchor™ System HG-18-90-22Medtronic, Inc. · Endovascular suturing system1 recall1 report
- Heli-FX™ EndoAnchor™ System HG-18-90-42Medtronic, Inc. · Endovascular suturing system1 recall0 reports
Product lines
Other devices under the same FDA submission
11 devices cleared under K171427 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Heli-FX™ EndoAnchor™ System SA-85Medtronic, Inc. · Endovascular suturing system6 deaths70 reports
- Heli-FX™ EndoAnchor™ System SA-85Medtronic, Inc. · Endovascular suturing system6 deaths41 reports
- Heli-FX™ EndoAnchor™ System SG-64Medtronic, Inc. · Endovascular suturing system18 reports
- Heli-FX™ EndoAnchor™ System HG-16-62-28Medtronic, Inc. · Endovascular suturing system5 reports
- Heli-FX™ EndoAnchor™ System HA-18-114Medtronic, Inc. · Endovascular suturing system1 death4 reports
- Heli-FX™ EndoAnchor™ System HA-18-114Medtronic Vascular, Inc. · Endovascular suturing system4 reports
- Heli-FX™ EndoAnchor™ System HG-18-90-32Medtronic, Inc. · Endovascular suturing system3 reports
- Heli-FX™ EndoAnchor™ System HG-16-62-28Medtronic, Inc. · Endovascular suturing system2 reports
- Heli-FX™ EndoAnchor™ System SG-64Medtronic, Inc. · Endovascular suturing system1 report
- Heli-FX™ EndoAnchor™ System HG-18-90-42Medtronic, Inc. · Endovascular suturing system1 report
- Heli-FX™ EndoAnchor™ System HG-18-90-42Medtronic, Inc. · Endovascular suturing system1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page