CD HORIZON® Spinal System RBT033302Recalls & reported problems

Medtronic Sofamor Danek, Inc. · DRIVER RBT033302 RG 5.5/6.0 VOYAGER LONG · Class 2

All models of CD HORIZON Spinal System (342)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000011284), received between Jun 4, 2024 and Mar 3, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Device Handling Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 3, 2025Break; Device Handling Problem; Break; Device Handling Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE SURGEON SHEARED THE DRIVER WHEN ATTEMPTING TO SCREW IN A SCREW IN THE PATIENT. THE SURGEON USED POWER EASE ON ALL SCREWS. THE SCREW HOLES WERE DRILLED AND TAPPED BEFORE INTRODUCING THE IMPLANTING SCREWS. HOWEVER, DUE TO THE PATIENT'S INCREASED BONE DENSITY, THE POWER EASE HAD A HARD TIME SCREWING IN ALL SIX SCREWS IN THE CASE. THE SCOPE WAS FROM L4 TO S1. HOWEVER, ON S1 ON THE RIGHT, THE SURGEON STOPPED THE POWER EASE IN THE MIDDLE OF SCREWING IN THE SCREW. DAVID SAID THAT SINCE THE SURGEON STOPPED THE POWER EASE BEFORE THE SCREW WAS FULLY INSERTED, THE POWER EASE WAS UNABLE TO CONTINUE SCREWING IN THE SCREW. THE SURGEON AT THAT POINT INTRODUCED THE DRIVER WITH A HANDLE. HOWEVER, THE SURGEON DID NOT ATTACH THE DRIVER TO THE SCREW BY SCREWING THE DRIVER TO THE SCREW USING THE THREADS. THE SURGEON ONLY ATTACHED THE HEAD OF THE DRIVER TO THE HEAD OF THE SCREW WHICH SHEARED THE HEAD OF THE DRIVER IN THE HEAD OF THE SCREW. THEY WERE NOT ABLE TO REMOVE THE SHEARED DRIVER HEAD FROM THE HEAD OF THE SCREW, SO THE SCREW HAD TO BE REMOVE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21511559 ↗
  • MalfunctionJun 4, 2024Break; Break

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DRIVER USED FOR SPINAL THERAPY. IT WAS REP ORTED THAT THE DEVICE HAS BROKEN. THE TIP WAS BROKEN OFF. PRODUCT WILL BE RETURNED. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19461978 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →