Skip to content

Heli-FX™ EndoAnchor™ System HG-18-90-32Recalls & reported problems

Medtronic, Inc. · TAA GUIDE DEVICE 32 EIFU · Class 2 · Implant

All models of Heli-FX EndoAnchor System (13)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000025281), received between Apr 17, 2018 and Feb 22, 2021 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K171427, K182957); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1
  • Difficult to Insert1
  • Failure to Unfold or Unwrap1
  • Material Deformation1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 22, 2021Failure to Unfold or Unwrap; Detachment of Device or Device Component; Failure to Unfold or Unwrap; Detachment of Device or Device Component

    AN HELI-FX ENDO-ANCHORING SYSTEM WAS ATTEMPTED TO BE USED ALONG WITH A VALIANT STENT GRAFT WHICH WAS IMPLANTED IN AN ENDOVASCULAR PROCEDURE TO TREAT AN ENDOLEAK IN A NON-MEDTRONIC STENT GRAFT IMPLANTED 2 YEARS AGO IN THE DISTAL PORTION CLOSE TO THE CELIAC TRUNK IT WAS REPORTED THAT DURING THE PROCEDURE, AFTER THE FIRST TWO ENDOANCHORS HAD BEEN IMPLANTED, DEFLECTION ISSUES WERE ENCOUNTERED IT WAS NOTED THAT AFTER THE SECOND ENDOANCHOR WAS IMPLANTED IT APPEARED TO DISLODGE THE RADIOPAQUE MARKER ON THE GUIDE, IT WAS REPORTED THE MARKER COULD NOT BE CAPTURED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE. THE CAUSE OF THE EVENT IS UNKNOWN. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11358699 ↗
  • MalfunctionMay 4, 2018Break; Break

    A HELI-FX APPLIER AND ENDOANCHORS WERE PLANNED TO BE USED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A TYPE IA ENDOLEAK OF A NON-MDT THORACIC STENT GRAFT ALONG WITH PENUMBRA COILS AND ONYX. IT WAS REPORTED THAT HELI-FX GUIDE HG-18-90-32 (LOT NUMBER 0008847767) BROKE DURING THE PROCEDURE, THAT THE CATHETER DIDN¿T TURN, AND THE HANDLE WAS SEPARATED FROM THE CATHETER. THE PHYSICIAN REPORTED THE CAUSE OF THE EVENT TO BE PRODUCT RELATED AND REPORTED LOW QUALITY OF THE PRODUCT ALONG WITH COMPLEX ANATOMY DUE TO ANGULATION. A SECOND HELI-FX GUIDE G-18-90-32 (LOT NUMBER 0008847776) WAS ALSO USED DURING THE PROCEDURE. IT WAS REPORTED THAT THIS HELI-FX GUIDE ALSO BROKE DURING THE PROCEDURE, THAT THE CATHETER DIDN¿T TURN, AND THE HANDLE WAS SEPARATED FROM THE CATHETER. AGAIN, THE PHYSICIAN REPORTED THE CAUSE OF THE EVENT TO BE PRODUCT RELATED AND REPORTED LOW QUALITY OF THE PRODUCT ALONG WITH COMPLEX ANATOMY DUE TO ANGULATION. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7487511 ↗
  • MalfunctionApr 17, 2018Difficult to Insert; Material Deformation; Difficult to Insert; Material Deformation

    A HELI-FX GUIDE, AND HELI-FX APPLIER AND ENDOANCHORS WERE USED PROPHYLACTICALLY IN THE ENDOVASCULAR TREATMENT OF THE PATIENT FOR ANO THER MANUFACTURERS STENT GRAFT. IT WAS REPORTED THAT DURING THE INDEX PROCEDURE AFTER THE IMPLANTATION OF THE FIRST TWO ENDOANCHORS THE PHYSICIAN WAS NOT ABLE TO ADVANCE THE HELI-FX APPLIER THROUGH THE HELI-FX GUIDE WHEN THE GUIDE WAS BENT. AS TWO ENDOANCHORS HAD BEEN SUCCESSFULLY IMPLANTED THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INTERVENTION. AS PER THE PHYSICIAN THE CAUSE OF THE EVENT WAS DUE TO OBSTRUCTION OF THE TIP OF THE GUIDE BY A PART OF THE GUIDE ITSELF WHICH WAS NOTED AFTER REMOVAL OF THE GUIDE FROM THE PATIENT. IT WAS NOTED THAT THE APPLIER WORKED SUCCESSFULLY. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7435520 ↗

FDA clearance and classification

Cleared through Special 510(k) K182957 on Nov 21, 2018 (decision: Substantially Equivalent). Submitted by Medtronic Vascular, Inc. as “Heli-FX EndoAnchor System”. The FDA review took 28 days. FDA record ↗

Cleared through Special 510(k) K171427 on Jun 13, 2017 (decision: Substantially Equivalent). Submitted by Medtronic Vascular, Inc. as “Heli-FX Applier, Heli-FX Guide, Ancillary EndoAnchor Cassette”. The FDA review took 29 days. FDA record ↗

The FDA regulates product code OTD (endovascular suturing system) under 21 CFR 870.3460 as a Class II device. 21 CFR 870.3460 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K171427, K182957. One submission often covers many models, so check the product and lot details in each record.

  • Z-3227-2017 · Sep 13, 2017 · Terminated
    Medtronic Aptus Heli-FX Thoracic EndoAnchor System Heli-FX Guide 42 mm, REF HG-18-90-42, STERILE EO, Rx Only The Heli-FX EndoAnchor System is intended to provide fixation and sea

Common questions about the Heli-FX™ EndoAnchor™ System HG-18-90-32

Has the Heli-FX™ EndoAnchor™ System HG-18-90-32 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00763000025281), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the Heli-FX™ EndoAnchor™ System HG-18-90-32?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Apr 17, 2018 and Feb 22, 2021. In those reports, the most common specific device problems are “Break” (1), “Detachment of Device or Device Component” (1) and “Difficult to Insert” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Heli-FX™ EndoAnchor™ System HG-18-90-32?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Heli-FX™ EndoAnchor™ System HG-18-90-32 FDA approved?

It was cleared through Special 510(k) K171427 on Jun 13, 2017; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Heli-FX™ EndoAnchor™ System HG-18-90-32 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Heli-FX™ EndoAnchor™ System HG-18-90-32, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand Heli-FX™ EndoAnchor™ System, model HG-18-90-32). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Endovascular suturing system devices

All Endovascular suturing system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page