N/A 9734682Recalls & reported problems

Medtronic Navigation, Inc. · TRACKER 9734682 NavLock UNIVERSAL VIOLET · Class 2

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00763000054298), received between Aug 30, 2018 and Dec 19, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K171267); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam3
  • Material Integrity Problem2
  • Mechanical Problem2
  • Use of Device Problem2
  • Break1
  • Imprecision1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Patient Involvement2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT ALL 4 COLORS HAD SLIGHT SCRATCHES ON THE INTERNAL SIDE WHICH COULD NOT BE CAPTURED IN AN IMAGE. ALL NAVLOCKS TRACKERS WERE DIFFICULT TO ATTACH OR REMOVE, ESPECIALLY THE GREY ONE WAS HARD AND WAS MOST LIKELY STUCK AS IT CANNOT BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23857414 ↗
  • MalfunctionDec 19, 2025Mechanical Jam; Mechanical Jam

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A CERVICAL SPINE PROCEDURE. IT WAS REPORTED THAT THE USER INSERTED THE INFINITY DRILL BIT INTO THE NAVLOCK TRACKER VIOLET AND CREATED A PILOT HOLE, BUT IT STOPPED ROTATING. ATTACHED IT TO ANOTHER NAVLOCK TRACKER, BUT IT BECAME STUCK AGAIN. THERE WAS NO SURGICAL DELAY. THERE WAS NO IMPACT ON PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23857386 ↗
  • MalfunctionMar 4, 2025Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THERE WERE MULTIPLE INSTRUMENTS THAT REQUIRED REPLACEMENT. THE ORANGE AND VIOLET INSTRUMENT TRACKER SHOWED THE RETAINING CLIP BROKEN OFF. THE THORACIC PROBE TIP WAS BENT AND NO LONGER ABLE TO BE LOGGED INTO THE SOFTWARE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21516034 ↗
  • MalfunctionJun 5, 2023Imprecision; Use of Device Problem; Imprecision; Use of Device Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A GUIDANCE SYSTEM BEING USED DURING A SPINAL PROCEDURE. IT WAS REPORTED THAT THE NAVIGATION WAS MEDIAL AND LATERAL FOR ALL THE INSTRUMENTS BUT ALL SCREWS WERE PLACED ACCURATELY, THE ARM WAS ACCURATE. THEY CHECKED NAVIGATION ACCURACY AND THE SNAPSHOT IMAGE AND ISSUE PERSISTED. THERE WAS NO DELAY AND NO IMPACT TO THE PATIENT OUTCOME. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THEY TESTED THE INSTRUMENTS AND THE NAVLOCKS WERE DAMAGED AND CAUSING THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17065957 ↗
  • MalfunctionJan 26, 2021Use of Device Problem; Use of Device Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE VIOLET INSTRUMENT TRACKER WAS DEFECTIVE. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11232607 ↗
  • MalfunctionJan 23, 2021Mechanical Jam; Mechanical Jam

    MEDTRONIC RECEIVED INFORMATION REGARDING NAVIGATION AND SPINAL INSTRUMENTS, OUTSIDE OF PROCEDURE. IT WAS REPORTED THAT THE VIOLET IN STRUMENT TRACKER AND TAP GOT STUCK, IT WAS NOT POSSIBLE TO REMOVE THE TAP. NO PATIENT WAS PRESENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11220110 ↗
  • MalfunctionOct 16, 2019Mechanical Problem; Material Integrity Problem; Mechanical Problem; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A GUIDANCE SYSTEM BEING USED PRIOR TO A SPINAL PROCEDURE. IT WAS REPORTED THAT THE TRACKERS WERE DAMAGED AND WOULD NOT TRACK. THE MANUFACTURER REPRESENTATIVE REPORTED THAT THE ISSUE HAD OCCURRED ON MULTIPLE OCCASIONS AND SITUATIONS. TROUBLESHOOTING WAS ATTEMPTED, BUT THE ISSUE WAS NOT RESOLVED. THE CAUSE OF NO TRACKING WAS NOT DETERMINED. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Patient Involvement

    MAUDE report 9200167 ↗
  • MalfunctionOct 17, 2018Material Integrity Problem; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM. IT WAS REPORTED THAT INTRA/PERI-OPERATIVELY DURING THE NAVIGATE TASK OF A SACROILIAC AND THORACOLUMBAR PROCEDURE, THE VIOLET NAVIGATED FRAME COMPONENT BECAME STUCK ON A TAP BECAUSE IT WAS DAMAGED. THERE WAS NO DELAY TO CASE. THERE WAS NO REPORTED IMPACT ON PATIENT OUTCOME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7973373 ↗
  • MalfunctionAug 30, 2018Break; Break

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE VIOLET NAVLOCK WAS DAMAGED. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS OBSERVED.

    Patient problems: No Patient Involvement

    MAUDE report 7832676 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171267. One submission often covers many models, so check the product and lot details in each record.

  • Z-1693-2020 · Aug 28, 2019 · Terminated
    NavLock Tracker *** Rx Only

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →