Visia AF MRI™ S VR SureScan™ DVFC3D4 (UDI 00763000060060)Recalls & reported problems
Medtronic, Inc. · ICD-VR DVFC3D4 VISIA AF MRI S US DF4 · Class 3 · Implant
All models of Visia AF MRI S VR SureScan (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic Visia implantable defibrillator is placed in the chest to correct fast heart rhythms with pacing or shocks. Reports mostly describe inappropriate shocks and sensing issues, along with infections that led to removal of the system. There are 24 reports that name its identifier, including 1 death and 17 injuries. It has 1 open recall started in May 2023 for a rare potential for reduced or missing energy during high-voltage therapy.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 23
- Deaths reported
- 1
- Injuries reported
- 16
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00763000060060), received between Oct 29, 2021 and May 7, 2026 — classified as 1 death, 16 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 10, 2023Class IOpen, Classified
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Z-1808-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem10
- Device Sensing Problem5
- Inappropriate/Inadequate Shock/Stimulation5
- Appropriate Device Problem Term/Code Not Available2
- Electrical /Electronic Property Problem2
- Pacing Problem2
- Failure to Deliver Shock/Stimulation1
- Failure to Interrogate1
Patient problems
- Unspecified Infection7
- No Clinical Signs, Symptoms or Conditions5
- Shock from Patient Lead(s)5
- Bacterial Infection2
- Erythema2
- Bradycardia1
- Drug Resistant Bacterial Infection1
- Non specific EKG/ECG Changes1
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
InjuryMay 7, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT THAT THEY EXPERIENCED BEING UNCONSCIOUS AND ENDED UP IN THE HOSPITAL. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 25113225 ↗InjuryFeb 2, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN UNSPECIFIED INFECTION. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS SUBSEQUENTLY EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 24240070 ↗InjuryJan 26, 2026Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED BY THE PATIENT THAT THEY FELT "A BIG SHOCK". THE PATIENT STATED THEY WERE MOVING A DOORWAY GATE IN THEIR HOUSE WHEN THEY FELT A LARGE SHOCK AND SAID THEY "SAW BLUE". THE PATIENT DID NOT KNOW IF THEY HIT SOMETHING WHEN MOVING THE GATE THAT SHOCKED THEM OR IF THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SHOCKED THEM. THE ICD REMAINS IN USE. THE MONITOR IS EXPECTED TO BE REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE HIGH VOLTAGE THERAPY FROM THE ICD FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE THAT WAS DETECTED AS VENTRICULAR TACHYCARDIA (VT). THE PATIENT UNDERWENT AN ABLATION PROCEDURE AND THE ICD WAS EXPLANTED DURING AN UPGRADE PROCEDURE TO A CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D).
Patient problems: Shock from Patient Lead(s)
MAUDE report 24184776 ↗InjuryJul 1, 2025Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT DURING USE OF THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THE PATIENT RECEIVED INAPPROPRIATE HIGH VOLTAGE THERAPY. IT WAS NOTED THAT NOISE AND OVERSENSING WERE OBSERVED ON THE SHOCK EPISODES THAT CORRELATED WITH THE TIME THAT THE PATIENT WAS UNDERGOING A LEFT LOWER EXTREMITY AMPUTATION. THE EPISODES WERE NOT CONSISTENT WITH A LEAD ISSUE. ADDITIONALLY, THE ICD INAPPROPRIATELY DETECTED A VENTRICULAR FIBRILLATION (VF) EPISODE. NO ACTION WAS TAKEN. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 22377393 ↗MalfunctionJan 6, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT PRIOR TO IMPLANT, THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS INTERROGATED AND THE BATTERY STATUS INDICATOR BAR WAS GREY. THE DEVICE WAS NOT USED AND THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21080012 ↗MalfunctionJun 26, 2024Pacing Problem; Pacing Problem
IT WAS REPORTED BY THE PATIENT'S DAUGHTER THAT THEY DID NOT SET THE DEVICE BACK AFTER A MAGNETIC RESONANCE IMAGING (MRI) WAS PERFORMED TWO WEEKS AGO. THE PATIENT EXPERIENCED HEART RATES OF 40 BEATS PER MINUTE (BPM) WHICH WAS BELOW THE THRESHOLD OF 60 BPM. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 19616625 ↗InjuryMar 21, 2024Pacing Problem; Pacing Problem
IT WAS REPORTED THAT APPROXIMATELY ONE YEAR POST-IMPLANT, THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS REMOVED AND REPLACED WITH A BIVENTRICULAR DEFIBRILLATOR SYSTEM AFTER THE PATIENT DEVELOPED LEFT BUNDLE BRANCH BLOCK FROM FREQUENT VENTRICULAR PACING. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 18950967 ↗MalfunctionMar 18, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT AND SPOUSE STATED THAT THEY WERE HEARING SOUNDS FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) ALL DAY YESTERDAY AND ALL DAY TODAY. THE ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18922050 ↗InjuryJan 5, 2024Failure to Deliver Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Failure to Deliver Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THERE WAS POSSIBLE INAPPROPRIATE THERAPY DELIVERED, AS THE VENTRICULAR FIBRILLATION (VF) ARRHYTHMIA APPEARED TO HAVE TERMINATED BEFORE THE SECOND SHOCK WAS DELIVERED. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD POTENTIALLY WITHHELD THERAPY INAPPROPRIATELY FOR A SUPRA VENTRICULAR TACHYCARDIA (SVT) DURING A SEPARATE EPISODE. THE RIGHT VENTRICULAR (RV) LEAD APPEARED TO BE UNDERSENSING DURING MONITORED AND TREATED EPISODES, WHICH RESULTED IN A POSSIBLE DELAY TO DETECTION/THERAPIES. A DECREASE IN R-WAVE AMPLITUDE WAS NOTED ON THE LEAD TRENDS. THE DEVICE AND LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 18454585 ↗InjuryNov 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ALMOST TWO MONTHS POST IMPLANT, THE PATIENT EXPERIENCED AN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) POCKET INFECTION WITH PURULENT DISCHARGE AND REDNESS. THE ICD SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Erythema; Purulent Discharge; Unspecified Infection
MAUDE report 18210429 ↗InjuryOct 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS ERODING THE POCKET, WHICH WAS RED AND SWELLING. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Erythema; Swelling/ Edema
MAUDE report 17876376 ↗MalfunctionJan 5, 2023Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE REMOTE MONITOR HAD NO TELEMETRY WITH THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL AND THE PATIENT WAS REFERRED TO CLINIC. THE ICD AND REMOTE MONITOR REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16109488 ↗DeathNov 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD A SYSTEMIC INFECTION. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Drug Resistant Bacterial Infection; Unspecified Infection
MAUDE report 15826101 ↗InjuryNov 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAS HAD CHRONIC BACTEREMIA SINCE THE IMPLANT OF THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM. THE SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacterial Infection
MAUDE report 15789651 ↗InjuryOct 3, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 15528220 ↗InjuryJun 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT FOLLOWING A NECK SURGERY THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) POCKET DEVELOPED SIGNS OF INFECTION AND THE ICD SYSTEM WAS EXPLANTED DUE TO THE INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 14850799 ↗InjuryJun 9, 2022Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) POSSIBLY DELIVERED THERAPY FOR AN ATRIAL ARRHYTHMIA. IT WAS NOTED THAT INITIALLY THE THERAPIES WERE WITHHELD FOR WAVELET BUT THEN WERE DELIVERED DUE TO HIGH RATE TIMEOUT. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 14645416 ↗InjuryApr 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD'S R-WAVE AMPLITUDE HAD DECLINED SINCE IMPLANT. IT WAS FURTHER REPORTED THAT THE PATIENT WAS BROUGHT IN-CLINIC FOR A TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) PROCEDURE WHICH SHOWED A LARGE VEGETATION ON THE RV LEAD. THE BLOOD CULTURES WERE POSITIVE HOWEVER THE ORGANISM IS UNKNOWN AT THIS TIME. IT WAS FURTHER NOTED THAT THE PATIENT HAS A HISTORY OF IV DRUG USE. THE PHYSICIAN MADE THE DECISION TO INTERVENE SURGICALLY BUT NOT TO EXTRACT THE LEAD OR THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AT THIS TIME. THE PATIENT IS BEING TREATED WITH ANTIBIOTICS AND IS BEING MONITORED. THE LEAD AND THE ICD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacterial Infection; Unspecified Infection
MAUDE report 14087027 ↗InjuryMar 31, 2022Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE PATIENT WAS RECEIVING INAPPROPRIATE SHOCKS FROM THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WHILE IN NORMAL SINUS RHYTHM. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 13966709 ↗MalfunctionMar 8, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT PRIOR TO IMPLANT, THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS INTERROGATED AND THE BATTERY STATUS INDICATOR BAR WAS GREY. THE DEVICE WAS NOT USED AND THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13694057 ↗InjuryJan 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 13258822 ↗InjuryDec 29, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD SEPSIS AND THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED AS A RESULT. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Sepsis
MAUDE report 13113759 ↗MalfunctionOct 29, 2021Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS INTERROGATED PRIOR TO IMPLANTED AND IT WAS NOTED THAT SEVERAL MOTHS PRIOR THE DEVICE HAD DETECTIONS/THERAPIES HAD BEEN TURNED ON AND THEN TURNED BACK OFF. THE DEVICE WILL NOT BE USED. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12722524 ↗
Other Implantable cardioverter defibrillator (non-CRT) devices
- EMBLEM™ MRI S-ICD A219Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 recalls36 deaths15,404 reports
- EMBLEM™ S-ICD A209Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 recall7 deaths6,788 reports
- EMBLEM™ S-ICD 3501Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)7 deaths4,936 reports
- Cobalt™ XT DR MRI SureScan™ DDPA2D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)3 recalls30 deaths1,603 reports
- ENDOTAK RELIANCE® G 0185Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,632 reports
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- Evera MRI™ XT DR SureScan® DDMB1D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)1 recall35 deaths1,426 reports
- Durata™ 7122St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)27 deaths1,377 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page