MIDAS REX LEGEND F2/8TA23Recalls & reported problems
Medtronic Powered Surgical Solutions · TOOL F2/8TA23 LEGEND 8CM 2.3MM TAPER · Class 2
All models of MIDAS REX LEGEND (37)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00763000061890), received between Jan 13, 2023 and Sep 14, 2026 — classified as 1 death, 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Fracture” (9 reports).
9 other recalls cover devices cleared under the same FDA submission (K020069); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture9
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Hemorrhage/Blood Loss/Bleeding2
- Device Embedded In Tissue or Plaque1
- Perforation1
- Pulmonary Dysfunction1
- Unintended Radiation Exposure1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2026Fracture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50762181 ↗MalfunctionJan 2, 2026Fracture
IT WAS REPORTED THAT DURING A CRANIOTOMY TO TREAT CEREBRAL HEMORRHAGE, THE DRILL BLADE BROKE DURING THE OPERATION, RESULTING IN A 10-MINUTE PROCEDURE DELAY. IT WAS REPORTED THAT THE PROCEDURE WAS COMPLETED USING A REPLACEMENT PRODUCT. AT THE TIME OF REPORT, THERE WAS NO KNOWN IMPACT TO A PATIENT. ADDITIONAL DETAILS REGARDING INTERVENTION AND PATIENT IMPACT IS BEING FOLLOWED UP FROM THE FACILITY. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23942254 ↗MalfunctionJan 2, 2026Fracture
IT WAS REPORTED THAT DURING A CRANIOTOMY TO TREAT CEREBRAL HEMORRHAGE, THE DRILL BLADE BROKE DURING THE OPERATION, RESULTING IN A 10-MINUTE PROCEDURE DELAY. IT WAS REPORTED THAT THE PROCEDURE WAS COMPLETED USING A REPLACEMENT PRODUCT. AT THE TIME OF REPORT, THERE WAS NO KNOWN IMPACT TO A PATIENT. ADDITIONAL DETAILS REGARDING INTERVENTION AND PATIENT IMPACT IS BEING FOLLOWED UP FROM THE FACILITY. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23942217 ↗MalfunctionSep 4, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING THE SURGERY THE DRILL BIT BROKE, MAKING IT UNUSABLE. THE DRILL BIT WAS REPLACED, AND THE SURGERY CONTINUED. ON ADDITIONAL FOLLOW UP THERE WAS NO PATIENT IMPACT ASSOCIATED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22963615 ↗MalfunctionFeb 21, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING THE SURGICAL SKULL DRILLING, THE SKULL WAS POLISHED DURING THE OPERATION, AND THE HANDPIECE BLADE WAS HOT. AFTER THE END, TWO BEARING BALLS OF THE HANDPIECE BLADE WERE FOUND TO FALL OFF. ON FOLLOW UP IT WAS REPORTED THAT THE OPERATION WAS SUSPENDED, NEW SURGICAL EQUIPMENT WAS REPLACED, AND THE OPERATION TIME WAS EXTENDED AND THE DETACHED BEARINGS FELL ON THE PATIENT'S SKULL, AND ARE QUICKLY REMOVED BY THE DOCTOR, AND DID NOT FALL INTO THE PATIENT'S CRANIAL CAVITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21431411 ↗MalfunctionDec 29, 2024Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT HAD A BRAIN HEMORRHAGE AND THE CRANIOTOMY WAS PERFORMED TO CLEAR THE HEMATOMA, THE MILLING BLADE BROKE DURING THE CRANIOTOMY. THE BALL DRILL WAS USED TO PERFORM THE CRANIOTOMY, WHICH GREATLY PROLONGED THE CRANIOTOMY TIME AND SIGNIFICANTLY WIDENED THE CRANIOTOMY SUTURE. ON FOLLOW-UP IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT, THERE WAS 15 MINUTES DELAY IN PROCEDURE AND THE DEVICE WAS DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21032310 ↗MalfunctionMay 20, 2024Fracture; Fracture
IT WAS REPORTED THAT THE DRILL TIP WAS BROKEN. IT WAS REPORTED THAT THERE WAS BASICALLY NO HARM. ADDITIONAL INFORMATION REGARDING THE EVENT WAS BEING FOLLOWED-UP. ON FOLLOW-UP, IT WAS REPORTED THAT THE CRANIAL PLATE MILLING PROCEDURE DURING CRANIOTOMY, AND NO PATIENT OR STAFF IMPACT, OPERATION WAS EXTENDED FOR LESS THAN 5 MINUTES, THE INTEGRITY OF THE SECTION WAS CHECKED, AND THE SURGICAL FIELD WAS IRRIGATED AND NO FRAGMENTS LEFT INSIDE THE PATIENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19353849 ↗InjuryOct 27, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING THE PROCEDURE OF INSERTION OF RIGHT FRONTAL EXTERNAL VENTRICULAR DRAIN POSTERIOR FOSSA CRANIOTOMY, THE TIP OF THE TOOL (8-9MM) SNAPPED OFF. AFTER AN X RAY, THE HEALTH CARE PROFESSIONAL WERE ABLE TO ESTABLISH THE TIP WAS STILL 90MM INSIDE THE BRAIN. DUE TO ITS LOCATION, THE HEALTH CARE PROFESSIONAL FELT IT WAS A GREATER RISK TO TRY AND REMOVE THE TIP THAN TO LEAVE IT IN ITS PLACE INSIDE THE PATIENT. IT WAS REPORTED THAT THE PROCEDURE WAS COMPLETE WITH BACKUP PRODUCT. FURTHER, IT WAS REPORTED THAT DEVICE WILL NOT BE RETURNED AS IT WAS DISCARDED.
Patient problems: Device Embedded In Tissue or Plaque; Unintended Radiation Exposure
MAUDE report 18020243 ↗InjuryApr 13, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING CRANIOTOMY SURGERY WAS PERFORMED WITH AN AIR DRILL DUE TO INTRACRANIAL HEMATOMA, THE DRILL BUR BROKE AND THE BROKEN DRILL BUR FELL OFF. IT WAS REPORTED THAT DUE TO BREAKAGE PREVENTED SURGERY FROM PROCEEDING PROPERLY, INCREASED BLEEDING, PROLONGED SURGICAL TIME. IT WAS REPORTED THAT SURGERY WAS COMPLETED WITH BACKUP PRODUCT. ON FURTHER FOLLOW-UP IT WAS REPORTED THAT NO INTERVENTION PERFORMED AND NO ADDITIONAL SURGERY PERFORMED DUE TO THE EVENT. NO PATIENT ABNORMALITIES WERE REPORTED AFTER THE OPERATION.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 16730894 ↗DeathJan 13, 2023Fracture; Fracture
ON FOLLOW UP, IT WAS REPORTED THAT PATIENT WAS OPERATED FOR A MENINGIOMA. AFTER THE BREACH, THERE WAS A LOT OF BLOOD FROM THE DURA MATTER. THEY LOST A LONG TIME DURING THE PROCEDURE AND BACKUP DEVICE WAS USED TO COMPLETE THE PROCEDURE. AFTER 15 DAYS IN INTENSIVE CARE, AND WITH EPILEPSY, HE HAD PULMONARY COMPLICATIONS. THE PATIENT HAS DIED ON THE (B)(6), 2023. IT IS UNKNOWN IF THIS DEATH IS LINKED WITH THE BREACH.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Perforation; Pulmonary Dysfunction
MAUDE report 16165116 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K020069. One submission often covers many models, so check the product and lot details in each record.
- Z-1602-2016 · Apr 8, 2016 · TerminatedMedtronic Midas Rex Dissecting Tool Legend 10 cm 4mm BA DIAM. Intended for use in neurosurgery.
- Z-0321-2016 · Oct 21, 2015 · TerminatedTool Legend 16cm 2.4 mm wire pass for Midas Rex Legend Dissecting Tool. Used by surgeons to drill holes.
- Z-0546-2010 · Nov 30, 2009 · TerminatedMedtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial , craniofacial and sternotom
- Z-2156-2008 · May 30, 2008 · TerminatedMedtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General S
- Z-1574-05 · Sep 1, 2005 · TerminatedMedtronic Midas Rex Classic, GS and Legend dissecting tools manufactured by Medtronic Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, tX 76137 USA, Rx only, S
- Z-1067-05 · Jun 29, 2005 · TerminatedMedtronic MIDAS REX Legend High Speed Pneumatic System, Gold Touch Motor. Manufactured by Medtronic, Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, TX 76137
- Z-0278-05 · Nov 11, 2004 · TerminatedMedtronic Midas Rex Legend High Speed Pneumatic System
- Z-0213-05 · Nov 3, 2004 · TerminatedMedtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic Powered Surgical Solutions, Fort Worth, Texas
- Z-0419-03 · Nov 11, 2002 · TerminatedLegend Lubricant System (Legend Gold, Gold Touch and Platinum motors and Legend Lubricant/Diffuser cartridge)
Other MOTOR, DRILL, PNEUMATIC devices
- NA 05.001.080Synthes GmbH · MOTOR, DRILL, PNEUMATIC432 reports
- NA CSR60Synthes GmbH · MOTOR, DRILL, PNEUMATIC221 reports
- Midas Rex® AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC163 reports
- Maestro Air 5407-200-000Stryker Corporation · MOTOR, DRILL, PNEUMATIC112 reports
- Maestro 5400210060Stryker Corporation · MOTOR, DRILL, PNEUMATIC95 reports
- MIDAS REX™ AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC91 reports
- CORE, Maestro 5400200000Stryker Corporation · MOTOR, DRILL, PNEUMATIC75 reports
- Maestro Air 5407-200-00HStryker Corporation · MOTOR, DRILL, PNEUMATIC56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page