Medtronic Navigation, Inc. · TRACKER 9734683 NavLock UNIVERSAL ORANGE · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00763000076634), received between Sep 9, 2020 and Jul 28, 2026 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K171267); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem4
  • Mechanical Jam2
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Device Embedded In Tissue or Plaque1
  • No Patient Involvement1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A separate, different tracker was used to complete the case. The issue observed was with an orange tracker, and a gray tracker was used to complete case. When replicating the issue, the same tracker was used.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50061051 ↗
  • MalfunctionFeb 7, 2025Mechanical Jam; Mechanical Jam

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT INSTRUMENTS WERE STUCK WITHIN TRACKERS. IT WAS IMPOSSIBLE TO CLIP THE TRACKER ON OR TO REMOVE IT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21334191 ↗
  • MalfunctionDec 30, 2024Use of Device Problem; Use of Device Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THERE WAS A DEFECTIVE INSTRUMENT DUE TO DAMAGE. THE PROBABLE CAUSE WAS USE OF THE INSTRUMENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21043400 ↗
  • InjuryOct 13, 2021Material Integrity Problem; Material Integrity Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED IN A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT THE HEALTH CARE PROFESSIONAL (HCP) WAS ACCESSING THE PEDICURE WITH THE THORACIC PROBE AND IT BROKE OFF IN THE PATIENT¿S VERTEBRAL BODY. THE HCP WAS HAMMERING ON THE NAVLOCK WHEN IT BROKE. THERE WAS NO REPORTED IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE TIP OF THE PROBE REMAINED IN THE PEDICLE OF THE PATIENT ANATOMY. THE TIP COULD NOT BE REMOVED BECAUSE IT WOULD HAVE CAUSED MORE DAMAGE.

    Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12624206 ↗
  • MalfunctionMay 19, 2021Use of Device Problem; Use of Device Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE INSTRUMENTS WERE DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11849001 ↗
  • MalfunctionJan 23, 2021Mechanical Jam; Mechanical Jam

    MEDTRONIC RECEIVED INFORMATION REGARDING NAVIGATION AND SPINAL INSTRUMENTS, OUTSIDE OF PROCEDURE. IT WAS REPORTED THAT THE ORANGE INSTRUMENT TRACKER WAS DAMAGED. IT WAS REPORTED THAT THE INSTRUMENT TRACKER AND TAP GOT STUCK, IT WAS NOT POSSIBLE TO REMOVE THE TAP. NO PATIENT WAS PRESENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11220178 ↗
  • MalfunctionJan 8, 2021Use of Device Problem; Use of Device Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE ORANGE INSTRUMENT TRACKER WAS DEFECTIVE. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11140351 ↗
  • MalfunctionSep 9, 2020Use of Device Problem; Use of Device Problem

    MEDTRONIC RECEIVED INFORMATION THAT, WHILE OUTSIDE OF A PROCEDURE, THE ORANGE INSTRUMENT TRACKER USED ALONGSIDE THE NAVIGATION SYSTEM WAS DEFECTIVE. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.

    Patient problems: No Patient Involvement

    MAUDE report 10511307 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171267. One submission often covers many models, so check the product and lot details in each record.

  • Z-1693-2020 · Aug 28, 2019 · Terminated
    NavLock Tracker *** Rx Only

Other Orthopedic stereotaxic instrument devices