The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00763000084790), received between Sep 25, 2024 and Jun 30, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Naturally Worn” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Naturally Worn3
- No Apparent Adverse Event2
- Overheating of Device2
- Detachment of Device or Device Component1
- Difficult to Insert1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 30, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ATTACHMENT HAS GENERAL DISSATISFACTION. IT WAS REPORTED THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHMENT WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25663220 ↗MalfunctionJun 15, 2026Naturally Worn
IT WAS REPORTED THAT PRIOR TO USE, BEARING WAS DEFECT. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING MISALIGNED WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25500937 ↗MalfunctionJun 11, 2026Naturally Worn
IT WAS REPORTED THAT, PRIOR TO USE AND DURING PRE-OPERATIVE PREPARATION, A DEVICE EXHIBITED BEARING ISSUES. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25467730 ↗MalfunctionJun 2, 2026Physical Resistance/Sticking
IT WAS REPORTED THAT THE ATTACHMENT WAS NOT WORKING PRIOR TO THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25369221 ↗MalfunctionMay 25, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THERE IS A SMALL SCREW THAT IS BROKEN INSIDE. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO A PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25268133 ↗MalfunctionApr 20, 2026Naturally Worn
IT WAS REPORTED THAT THE DEVICE BEARING WAS DEFECT AFTER THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARINGS MISALIGNED WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24922232 ↗MalfunctionApr 2, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE ATTACHMENT HAS A GENERIC ISSUE.IT WAS REPORTED THAT THERE WAS NO PATINET INVOLVEMENT.ADDITIONAL INFORMATION RECEIVED STATED THAT THE DISTAL BEARINGS GOT DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24764033 ↗MalfunctionMar 20, 2026Difficult to Insert
IT WAS REPORTED THAT THE DEVICE WITH THE REPORT OF UNABLE TO LOCK THE TOOL . IT IS REPORTED THAT THERE IS NO PATIENT INVOLVEMENT .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24646601 ↗MalfunctionFeb 13, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21372510 ↗MalfunctionSep 25, 2024Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. REPAIR IS BEING ESCALATED TO PRODUCT EVENT DUE TO REASON FOR THE RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20299988 ↗
Other MOTOR, DRILL, PNEUMATIC devices
- NA 05.001.080Synthes GmbH · MOTOR, DRILL, PNEUMATIC432 reports
- NA CSR60Synthes GmbH · MOTOR, DRILL, PNEUMATIC221 reports
- Midas Rex® AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC163 reports
- Maestro Air 5407-200-000Stryker Corporation · MOTOR, DRILL, PNEUMATIC112 reports
- Maestro 5400210060Stryker Corporation · MOTOR, DRILL, PNEUMATIC95 reports
- MIDAS REX™ AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC91 reports
- CORE, Maestro 5400200000Stryker Corporation · MOTOR, DRILL, PNEUMATIC75 reports
- Maestro Air 5407-200-00HStryker Corporation · MOTOR, DRILL, PNEUMATIC56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page