This Medtronic Midas Rex surgical drill attachment is used with a powered surgical drill. The 24 reports mostly describe worn or broken bearings, parts detaching or falling off, and difficulty assembling the attachment; most mention no patient involvement. There were no deaths or injuries, and no recalls name this device. Reports increased from 2 in the previous 12 months to 22 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00763000084806), received between Mar 17, 2025 and Jun 22, 2026 — classified as 24 malfunctions.

The most frequently reported specific device problem is “Naturally Worn” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 6 reportsJun 2026: 14 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Naturally Worn13
  • Detachment of Device or Device Component7
  • Difficult to Insert5
  • No Apparent Adverse Event2
  • Overheating of Device1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions23
  • Insufficient Information1

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 22, 2026Naturally Worn

    IT WAS REPORTED THAT THE ATTACHMENT HAD BEARING PROBLEM. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25579246 ↗
  • MalfunctionJun 19, 2026Naturally Worn

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF WORN. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25556770 ↗
  • MalfunctionJun 16, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE ATTACHMENT HAD PARTS FALL OFF. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25515442 ↗
  • MalfunctionJun 13, 2026Naturally Worn

    IT WAS REPORTED THAT THE BEARING WAS DAMAGED PRIOR TO USE. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING BROKEN WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25494849 ↗
  • MalfunctionJun 10, 2026Naturally Worn

    IT WAS REPORTED THAT ATTACHMENT BALL BEARING WAS BROKEN. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHED WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25455466 ↗
  • MalfunctionJun 10, 2026Detachment of Device or Device Component; Naturally Worn

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF WORN. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25453401 ↗
  • MalfunctionJun 9, 2026Naturally Worn

    IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT HAD BEARING ISSUE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25442910 ↗
  • MalfunctionJun 8, 2026Difficult to Insert; Naturally Worn

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF DIFFICULTY WITH ASSEMBLY. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25426359 ↗
  • MalfunctionJun 6, 2026Naturally Worn

    IT WAS REPORTED THAT POST OPERATIVELY, THE ATTACHMENT BEARING WAS DEFECTIVE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25418523 ↗
  • MalfunctionJun 5, 2026Naturally Worn

    IT WAS REPORTED THAT THE BEARINGS OF THE ATTACHMENT WERE FOUND BROKEN PRIOR TO USE. AT THE TIME OF REPORT, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25408708 ↗
  • MalfunctionJun 5, 2026Naturally Worn

    IT WAS REPORTED THAT THE DEVICE BEARING WAS SPOILED AFTER THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARINGS MISALIGNED WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25404619 ↗
  • MalfunctionJun 5, 2026Naturally Worn

    IT WAS REPORTED THAT PRIOR TO USE, ATTACHMENT BEARINGS WERE DAMAGED. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25403857 ↗
  • MalfunctionJun 4, 2026Difficult to Insert

    IT WAS REPORTED THAT PRIOR TO USE, ATTACHMENT WITH THE REPORT OF NOT ATTACHING TO HOSE. IT WAS REPORTED THAT THERE WAS NO PATIENT IN VOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389573 ↗
  • MalfunctionJun 1, 2026Difficult to Insert

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF TOOL CANNOT BE INSERTED. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25355380 ↗
  • MalfunctionMay 29, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT PRIOR TO USE, ATTACHMENT PARTS FALL OUT. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25328584 ↗
  • MalfunctionMay 29, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE DURING PREOP THE THREADS IN THE INTERNAL CHANNEL OF THE NIPPLE APPEARB BROKEN. NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25327351 ↗
  • MalfunctionMay 25, 2026Detachment of Device or Device Component; No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: Insufficient Information

    MAUDE report 25268405 ↗
  • MalfunctionMay 25, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE DEVICE BALL WAS DETACHED WAS ISSUE PRIOR TO USE. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING FRACTURE WERE FOUND DURING EVALUATION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25268007 ↗
  • MalfunctionMay 25, 2026Difficult to Insert

    IT WAS REPORTED THAT PRIOR TO USE THE TOOL WAS NOT INSERTING INTO THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION RECEIVED STATING THAT THE DISTAL BEARINGS GOT DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25267623 ↗
  • MalfunctionMay 4, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST, AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. AT THE TIME OF REPORT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25065584 ↗
  • MalfunctionApr 9, 2026Naturally Worn

    IT WAS REPORTED THAT THE BEARING WAS BROKEN PRIOR TO THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARINGS ARE DETACHED FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24827462 ↗
  • MalfunctionMar 26, 2026Naturally Worn

    IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT BEARING WAS BROKEN .NO PATIENT INVOLVEMENT OR IMPACT TO PATIENT OR STAFF OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698602 ↗
  • MalfunctionSep 17, 2025Overheating of Device; Physical Resistance/Sticking; Overheating of Device; Physical Resistance/Sticking

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING AND LOCK OR ADJUSTMENT RING STICKING.IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23070504 ↗
  • MalfunctionMar 17, 2025Difficult to Insert; Detachment of Device or Device Component; Difficult to Insert; Detachment of Device or Device Component

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF DETACHMENT OF COMPONENT AND DIFFICULTY WITH ASSEMBLY. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21619777 ↗