Midas Rex AS09SRecalls & reported problems
Medtronic Powered Surgical Solutions · ATT AS09S LEGEND 9CM STRAIGHT SH 3.2MM · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00763000084813), received between Mar 10, 2025 and Sep 29, 2026 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (16 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component16
- Difficult to Insert1
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions17
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50928966 ↗MalfunctionJun 5, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DEVICE, PARTS FALLING OFF. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25406026 ↗MalfunctionJun 2, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT PRIOR TO USE, THE ATTACHMENT PARTS FELL OFF. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25357076 ↗MalfunctionJun 1, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE PARTS WERE FELL OUT PRIOR TO THE PROCEDURE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT FRACTURE OF BEARING FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25345945 ↗MalfunctionMay 29, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT PRIOR TO USE, ATTACHMENT PARTS FALL OFF. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25328816 ↗MalfunctionMay 28, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE DEVICE PARTS FALL OFF PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING FRACTURE WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25314420 ↗MalfunctionMay 27, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE DEVICE WITH THE REPORT OF PARTS FALL OFF. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25298676 ↗MalfunctionMay 27, 2026Difficult to Insert; Detachment of Device or Device Component
IT WAS REPORTED THAT DEVICE, BALL BEARING FELL OUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25297937 ↗MalfunctionMay 27, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING PRIOR TO USE, ATTACHMENT PARTS FALL OUT. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25296490 ↗MalfunctionMay 26, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF BEARING DETACHMENT. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25273764 ↗MalfunctionMay 26, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE DEVICE HAD THE REPORT OF PARTS FALLING OFF. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25273648 ↗MalfunctionDec 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT BEFORE USE, THE PARTS WERE FELLED OFF. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23862881 ↗MalfunctionDec 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT BEARING DETACHMENT IN THE ATTACHMENT. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23769836 ↗MalfunctionJul 14, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF BEARING DETACHMENT. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22499860 ↗MalfunctionJul 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF DETACHMENT OF COMPONENT.IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22392998 ↗MalfunctionJul 2, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF PARTS FALL OFF. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22381357 ↗MalfunctionMar 10, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21558377 ↗
Other MOTOR, DRILL, PNEUMATIC devices
- NA 05.001.080Synthes GmbH · MOTOR, DRILL, PNEUMATIC432 reports
- NA CSR60Synthes GmbH · MOTOR, DRILL, PNEUMATIC221 reports
- Midas Rex® AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC163 reports
- Maestro Air 5407-200-000Stryker Corporation · MOTOR, DRILL, PNEUMATIC112 reports
- Maestro 5400210060Stryker Corporation · MOTOR, DRILL, PNEUMATIC95 reports
- MIDAS REX™ AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC91 reports
- CORE, Maestro 5400200000Stryker Corporation · MOTOR, DRILL, PNEUMATIC75 reports
- Maestro Air 5407-200-00HStryker Corporation · MOTOR, DRILL, PNEUMATIC56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page