FUSION Compact ™ 9735602RRecalls & reported problems

Medtronic Navigation, Inc. · COMPUTER 9735602R FUSION COMP RFB LOADED · Class 2

All models of FUSION Compact (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000099787), received between Feb 24, 2022 and Apr 3, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Freezes or Fails to Launch1
  • Application Program Problem1
  • Communication or Transmission Problem1
  • Device Sensing Problem1
  • Display or Visual Feedback Problem1
  • Failure to Power Up1
  • Mechanical Problem1
  • Power Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2025Mechanical Problem; Application Program Problem; Mechanical Problem; Application Program Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE SITE WAS HAVING ISSUES WHERE THE UNIT OUT BOOT INTO A "LOCALIZER FAULTED" ERROR, REBOOTING WOULD RESOLVED THE ISSUE EACH TIME. THE SITE WAS GETTING IMPATIENT REGARDING THIS ISSUE AS IT WAS A DAILY OCCURRENCE. THE REPRESENTATIVE WAS UNABLE TO CHECK FOR THE LED NUMBER STATUS ON THE AXIEM BOX EACH TIME AS THE SITE REBOOTED THE UNIT RIGHT AWAY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21763528 ↗
  • MalfunctionMar 19, 2025Display or Visual Feedback Problem; Failure to Power Up; Power Problem; Unexpected Shutdown; Application Program Freezes or Fails to Launch; Display or Visual Feedback Problem; Failure to Power Up; Power Problem; Unexpected Shutdown; Application Program Freezes or Fails to Launch

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT UPON BOOTUP, THE SYSTEM POWER BUTTON TURNED GREEN AND THE MEDTRONIC LOGO FLASHED ON THE SCREEN, BUT THEN THE SYSTEM WOULD SHUT OFF. THE SCREEN WOULD BLACK AND THE POWER BUTTON HAD NO LIGHTS. THE USER TRIED MULTIPLE REBOOTS AND DIFFERENT OUTLETS, BUT IT DID NOT RESOLVE THE ISSUE. A MEDTRONIC REPRESENTATIVE (REP) LATER CALLED FOR ADDITIONAL TROUBLESHOOTING. THE REP TESTED THE VOLTAGE ON SEVERAL POWER OUTLETS AND NOTED 120 VOLTS. THE REP ALSO SWAPPED SYSTEM POWER CABLES WITH A KNOWN WORKING NAVIGATION SYSTEM AND NOTED THAT THIS SYSTEM STILL WOULD NOT POWER ON, EVEN WITH A KNOWN WORKING POWER CABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21644132 ↗
  • MalfunctionJan 2, 2025Communication or Transmission Problem; Communication or Transmission Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE SYSTEM DISPLAYED INVERTED IMAGES EVERY TIME IT RECEIVED A DIGITAL INTERCOMMUNICATION OF MEDICINE (DICOM) UPLOAD. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21059654 ↗
  • MalfunctionFeb 24, 2022Device Sensing Problem; Device Sensing Problem

    ON 2022-FEB-02 (B)(4) (REP, HCP): MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE. IT WAS REPORTED THAT THE SITE WAS HAVING INTERMITTENT TRACKING DIFFICULTY DURING A CASE.  TROUBLESHOOTING STEPS WERE PERFORMED.  THE MANUFACTURER REPRESENTATIVE WALKED THE SITE THROUGH THE TRACKING DETAILS, ENSURING THAT NO METAL WAS IN THE FIELD.  THEY ALSO ADJUSTED THE EMITTER PLACEMENT, AND CHECKED THE STATUS OF THE EMITTER BOX.  IT WAS NOTED THAT THE PROBABLE CAUSE OF THE ISSUE WAS DUE TO A MALFUNCTIONING EMITTER.  ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE. IT WAS REPORTED THAT THE TROUBLESHOOTING STEPS THAT WERE PERFORMED AT THE TIME OF THE EVENT WERE NOT ABLE TO RESOLVE THE ISSUE. IT WAS NOTED THAT THE EMITTER WOULD BE RETURNED FOR ANALYSIS. THERE WAS LESS THAN AN HOUR DELAY TO THE PROCEDURE AND NO IMPACT TO PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13596219 ↗