Bio-Medicus™ 96530-119 (UDI 00763000116033)Recalls & reported problems
Medtronic, Inc. · CANNULA 96530-119 BIO-MEDICUS 19FR EA · Class 2
All models of Bio-Medicus (59)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000116033), received between Apr 20, 2023 and Jul 3, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Crack” (1 report).
1 other recall cover devices cleared under the same FDA submission (K153598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Crack1
- Material Deformation1
- Obstruction of Flow1
Patient problems
- Exsanguination1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 3, 2023Obstruction of Flow; Obstruction of Flow
MEDTRONIC RECEIVED INFORMATION THAT DURING USE, THE CUSTOMER REPORTED THE PATIENT HAD BEEN INTUBATED AND VENTILATED FOR 7 DAYS WITH INCREASING VENTILATOR REQUIREMENTS. THE PATIENT WAS CANNULATED WITH A BIO-MEDICUS 19 FR ARTERIAL JUGULAR VENOUS ADULT CANNULA IN THE OPERATING ROOM (OR). CANNULATION ITSELF WAS UNEVENTFUL. THE ANAESTHETIST OBSERVED THE GUIDEWIRE ON TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) AND EVERYONE WAS HAPPY WITH BOTH THE INTERNAL JUGULAR (IJ) AND FEMORAL VENOUS POSITION. IT WAS NOTED ON TEE THAT THERE WAS A LARGE CLOT LOCATED IN THE RIGHT ATRIUM. HEPARIN WAS GIVEN PRIOR TO CANNULA INSERTION. THE GUIDEWIRE IN THE IJ WAS INSERTED OVER THE CENTRAL LINE WHICH THEN WAS REMOVED. THE BIO-MEDICUS DILATORS WERE USED AND A VERY SMALL SKIN INCISION WAS MADE WHILE THE LAST DILATOR WAS IN, JUST PRIOR TO THE FINAL CANNULA INSERTION. CANNULA INSERTION WENT SMOOTHLY AND NO OBSTRUCTION OR RESISTANCE TO INSERTION WAS OBSERVED. IJ LINE WAS CLAMPED WHILE THE FEMORAL VENOUS CANNULA WAS INSTALLED, AGAIN, NO ISSUES WERE NOTED. UPON CONNECTION OF THE IJ CANNULA TO THE EXTRA-CORPOREAL LIFE SUPPORT (ECLS) CIRCUIT CLOT WAS NOTED IN THE IJ CANNULA. THE CANNULATING PHYSICIAN LET THE CANNULA BLEED
Patient problems: Exsanguination
MAUDE report 17245890 ↗MalfunctionApr 20, 2023Crack; Material Deformation; Crack; Material Deformation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THE BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL OR JUGULAR VENOUS CANNULAE, IT WAS REPORTED THAT THE TIP OF THE CONNECTORS ON THE CANNULAE WERE WHITE, STRESSED AND CRACKED. THE CANNULAE WERE NOT USED FOR THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIPS WERE NOT OBSTRUCTED. THE CANNULAE WERE NOT HEATED OR COOLED PRIOR TO USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16776183 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K153598. One submission often covers many models, so check the product and lot details in each record.
- Z-2739-2026 · May 25, 2026 · Open, ClassifiedMedtronic Bio-Medicus Adult Cannula Kit, Product Number/CFN 96530-115; Cardiopulmonary bypass vascular cannula
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY11R00R6Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page