Bio-Medicus™ 96570-115 (UDI 00763000116071)Recalls & reported problems

Medtronic, Inc. · CANNULA 96570-115 BIO-MEDICUS 15FR EA · Class 2

All models of Bio-Medicus (59)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00763000116071), received between Feb 27, 2023 and Oct 21, 2025 — classified as 1 death, 6 malfunctions.

The most frequently reported specific device problem is “Crack” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack6
  • Material Puncture/Hole2
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 21, 2025Crack; Crack

    IT WAS REPORTED THAT A CRACK MEASURING 1 INCH WAS IDENTIFIED ON THE CONNECTORS OF THE BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL/JUGULAR VENOUS CANNULAE DURING USE. THE DAMAGE WAS NOT LOCATED IN THE AREA OF THE SUTURES. THE DEVICES WERE REPLACED TO COMPLETE THE PROCEDURE. IT WAS UNKNOWN IF THERE WAS ANY COMPLICATIONS AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23347166 ↗
  • MalfunctionOct 21, 2025Crack; Crack

    IT WAS REPORTED THAT A CRACK MEASURING 1 INCH WAS IDENTIFIED ON THE CONNECTORS OF THE BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL/JUGULAR VENOUS CANNULAE DURING USE. THE DAMAGE WAS NOT LOCATED IN THE AREA OF THE SUTURES. THE DEVICES WERE REPLACED TO COMPLETE THE PROCEDURE. IT WAS UNKNOWN IF THERE WAS ANY COMPLICATIONS AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23347113 ↗
  • DeathMar 19, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT DURING USE OF A BIO-MEDICUS CANNULA, THERE WAS A SUDDEN BREAK AT THE LATERAL BRANCH CIRCULATION. THE USE OF THE DEVICE WAS UNSPECIFIED. IT WAS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT OR COMPLICATIONS AS A RESULT OF THE EVENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21641315 ↗
  • MalfunctionFeb 10, 2025Crack; Crack

    IT WAS REPORTED THAT DURING SURGERY THE BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL OR JUGULAR VENOUS CANNULA WAS FOUND TO BE DAMAGED UPON UNPACKING. THE DEVICE WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21342098 ↗
  • MalfunctionJan 30, 2024Material Puncture/Hole; Material Puncture/Hole

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL CANNULA, THE CUSTOMER REPORTED THAT WHILE PERFORMING CARDIAC SURGERY TO ESTABLISH PERIPHERAL SYSTEMIC CIRCULATION, THEY DISCOVERED A CRACK AT THE TIP END OF THE TUBE BODY WHEN USING THE DEVICE. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE. THE CUSTOMER STATED THAT THERE WAS NO DAMAGE IN THE LOCATION OF THE SUTURES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18605296 ↗
  • MalfunctionJan 12, 2024Crack; Material Puncture/Hole; Crack; Material Puncture/Hole

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL OR JUGULAR VENOUS CANNULA, THE CUSTOMER R EPORTED THAT WHILE THE DEVICE WAS BEING PLACED THEY NOTICED THAT UPON OPENING THAT THERE WAS A FISSURE IN THE DISTAL PART. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS NOT HEATED OR COOLED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CUSTOMER OBSERVED THE ISSUE PRIOR TO INSERTION SO THEY REPLACED THE PRODUCT AND THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18501076 ↗
  • MalfunctionFeb 27, 2023Crack; Crack

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL CANNULA, IT WAS REPORTED THAT THE DEVICE HAD A CRACK AT THE INTERFACE. USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16442453 ↗

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices