Bio-Medicus™ 96570-123Recalls & reported problems

Medtronic, Inc. · CANNULA 96570-123 BIO-MEDICUS 23FR EA · Class 2

All models of Bio-Medicus (59)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000116118), received between Apr 19, 2023 and Jan 8, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack1
  • Misassembled During Installation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 8, 2026Crack

    IT WAS REPORTED THAT AFTER INSERTING THE BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL/JUGULAR VENOUS CANNULA INTO THE FEMORAL VEIN DURING AN ECMO (EXTRACORPOREAL MEMBRANE OXYGENATION) PROCEDURE, A 1/2" CRACK WAS IDENTIFIED ON THE CONNECTOR OF THE CANNULA. THE CRACK WAS NOT ATTRIBUTED TO THE CANNULATION PROCESS OR THE USE OF CLAMPS, AND THE PACKAGING WAS UNDAMAGED PRIOR TO USE. THE DAMAGE WAS NOT LOCATED NEAR THE SUTURES. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24003205 ↗
  • InjuryApr 19, 2023Misassembled During Installation; Misassembled During Installation

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS NEXTGEN FEMORAL ARTERIAL OR JUGULAR VENOUS CANNULA, IT WAS REPORTED THAT THE VEIN NEEDLE HAD A WRONG CONNECTOR, 1/2-3/8 PART INSTEAD OF 3/8-3/8. IT COULD THEREFORE NOT CONNECT TO THE HOSE SETS ON THE EXTRA-CORPOREAL MEMBRANE OXYGENATION (ECMO) MACHINE, SO IT TOOK APPROXIMATELY 5 MINUTES BEFORE THE PATIENT WAS CONNECTED TO ECMO. THE CORRECT CONNECTOR HAD TO BE FOUND, AND THE WRONG CONNECTOR HAD TO BE CUT OUT DURING THE STERILE PROCEDURE WHILE THE VENOUS CANNULA WAS IN THE PATIENT'S LEFT GROIN. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16766728 ↗

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices