ELEVATE™ Spinal System NAV2134Recalls & reported problems

Medtronic Sofamor Danek, Inc. · INSERTER NAV2134 DRIVE SHAFT · Class 2

All models of ELEVATE Spinal System (33)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000141424), received between Feb 1, 2022 and Jan 5, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 5, 2026Break

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING INSERTERS. IT WAS REPORTED THAT INSERTERS WERE BROKEN AS HANDLES BROKE ON BOTH. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23960365 ↗
  • MalfunctionDec 26, 2023Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING L5/S1 POSTERIOR FUSION AND TLIF FOR L5/S1 DEGENERATIVE DISC DISEASE. IT WAS REPORTED THAT THE TORQUE HANDLE RELEASED WHEN REACHED MAXIMUM TORQUE DUE TO THIS THE TIP OF THE INSERTER SHEARD OFF INSIDE THE IMPLANT IN THE PATIENT. STEPS WERE TAKEN TO REMOVE THE SHEARED COMPONENT. THE SURGEON THEN USED A PENCIL TIP HIGH SPEED DRILL TO SECTION AND REMOVE THE RETAINED TIP IN PIECES. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18391331 ↗
  • MalfunctionFeb 1, 2022Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING TLIF DUE TO UNK NOWN INDICATION. IT WAS REPORTED THAT WHEN EXPANDING THE IMPLANT THE TIP OF THE TORQUE-LIMITING INNER DRIVER SHAFT SHEARED OFF INTO THE IMPLANT. TIP IS LODGED INTO THE IMPLANT AND WOULDN'T COME OUT. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 13411870 ↗

FDA clearance and classification

Cleared through 510(k) K163581 on Apr 14, 2017 (decision: Substantially Equivalent). Submitted by Medtronic Sofamor Danek USA, Inc. as “Navigated ELEVATE Inserter, Navigated Disc Prep Instruments”. The FDA review took 115 days. FDA record ↗

The FDA regulates product code OLO (orthopedic stereotaxic instrument) under 21 CFR 882.4560 as a Class II device. 21 CFR 882.4560 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ELEVATE™ Spinal System NAV2134

Has the ELEVATE™ Spinal System NAV2134 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00763000141424), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ELEVATE™ Spinal System NAV2134?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Feb 1, 2022 and Jan 5, 2026. In those reports, the most common specific device problem is “Break” (3); the most common patient problem is “Device Embedded In Tissue or Plaque” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ELEVATE™ Spinal System NAV2134?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ELEVATE™ Spinal System NAV2134 FDA approved?

It was cleared through 510(k) K163581 on Apr 14, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the ELEVATE™ Spinal System NAV2134, and is it still sold?

The FDA's device database lists Medtronic Sofamor Danek, Inc. as the company responsible for it (brand ELEVATE™ Spinal System, model NAV2134). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.