Midas Rex PM700-R (UDI 00763000182052)Recalls & reported problems

Medtronic Powered Surgical Solutions · MOTOR PM700-R MR7 PNEUMATIC REFURB · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000182052), received between May 4, 2023 and Feb 19, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device3
  • Difficult to Remove1
  • Electrical /Electronic Property Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Unspecified Tissue Injury1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2025Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21407709 ↗
  • MalfunctionFeb 18, 2025Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21400339 ↗
  • MalfunctionFeb 18, 2025Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR THE DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21393314 ↗
  • InjuryMay 4, 2023Electrical /Electronic Property Problem; Difficult to Remove; Electrical /Electronic Property Problem; Difficult to Remove

    IT WAS REPORTED THAT DURING THE CRANIOTOMY PROCEDURE, THE BURR HOLES WERE DRILLED TO ACCESS THE BRAIN TUMOR USING THE MIDAS REX, USING A CODMAN 14MM DISPOSABLE PERFORATOR ATTACHMENT. FIRST BURR HOLE WAS DRILLED WITHOUT ANY INCIDENT, BUT THE DURING THE SECOND BURR HOLE THE ATTACHMENT GOT STUCK, WHILE LODGING THE PERFORATOR IN THE PATIENT'S SKULL. THE PERFORATOR COULD NOT BE MANUALLY REMOVED AND HAD TO BE CUT OUT. IT WAS ALSO REPORTED THAT INTERVENTION WAS PERFORMED, AND THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT(S). ON FOLLOW UP IT WAS REPORTED THAT THERE WERE NO FRAGMENTS LEFT INSIDE THE PATIENT AND THERE WAS A PROCEDURE DELAY APPROXIMATELY 45 MINUTES TO CALL UP ANOTHER DRILL. IT WAS ALSO REPORTED THAT AFTER OPERATION, THE CONDITION OF THE PATIENT IS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Tissue Injury

    MAUDE report 16866069 ↗

Other MOTOR, DRILL, PNEUMATIC devices

All MOTOR, DRILL, PNEUMATIC devices →