NA HY8N48R13Recalls & reported problems
Medtronic, Inc. · CUSTOM PACK HY8N48R13 HEART LUN · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000200336), received between Jun 12, 2023 and Jun 12, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K171308); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak3
- Material Separation3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 12, 2023Fluid/Blood Leak; Material Separation; Fluid/Blood Leak; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF THIS CUSTOM TUBING PACK A LEAK WAS OBSERVED FROM THE TUBING AT THE CONNECTION. THE CUSTOMER STATED THAT THERE WAS A ZIP TIE BUT NO ADHESIVE AT THE SITE. THE CUSTOMER CONTINUED TO USE THE PACK AFTER OBSERVING THE LEAK. THEY CLAMPED TO CONTROL THE LEAK. THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THE EVENT. ADDITIONAL INFORMATION: THE AMOUNT OF BLOOD LOST DUE TO THE LEAK WAS ASKED BUT IS UNKNOWN. NO BLOOD TRANSFUSION WAS REQUIRED DUE TO THE LEAK. THERE WAS NO DAMAGE TO THE TUBING OBSERVED, JUST BONDING DID NOT BOND AND THE ZIP TIE WAS LOOSE. THERE WAS NO VISIBLE AIR IN THE SYSTEM/TUBING. THE EXACT LOCATION OF THE LEAK IN THE CUSTOM TUBING PACK SPECIFICATION WAS LINE 5.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17106397 ↗MalfunctionJun 12, 2023Fluid/Blood Leak; Material Separation; Fluid/Blood Leak; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF THIS CUSTOM TUBING PACK A LEAK WAS OBSERVED FROM THE TUBING AT THE CONNECTION. THE CUSTOMER STATED THAT THERE WAS A ZIP TIE BUT NO ADHESIVE AT THE SITE. THE CUSTOMER CONTINUED TO USE THE PACK AFTER OBSERVING THE LEAK. THEY CLAMPED TO CONTROL THE LEAK. THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THE EVENT. ADDITIONAL INFORMATION: THE AMOUNT OF BLOOD LOST DUE TO THE LEAK WAS ASKED BUT IS UNKNOWN. NO BLOOD TRANSFUSION WAS REQUIRED DUE TO THE LEAK. THERE WAS NO DAMAGE TO THE TUBING OBSERVED, JUST BONDING DID NOT BOND AND THE ZIP TIE WAS LOOSE. THERE WAS NO VISIBLE AIR IN THE SYSTEM/TUBING. THE EXACT LOCATION OF THE LEAK IN THE CUSTOM TUBING PACK SPECIFICATION WAS LINE 5.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17106388 ↗MalfunctionJun 12, 2023Fluid/Blood Leak; Material Separation; Fluid/Blood Leak; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF THIS CUSTOM TUBING PACK A LEAK WAS OBSERVED FROM THE TUBING AT THE CONNECTION. THE CUSTOMER STATED THAT THERE WAS A ZIP TIE BUT NO ADHESIVE AT THE SITE. THE CUSTOMER CONTINUED TO USE THE PACK AFTER OBSERVING THE LEAK. THEY CLAMPED TO CONTROL THE LEAK. THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THE EVENT. ADDITIONAL INFORMATION: THE AMOUNT OF BLOOD LOST DUE TO THE LEAK WAS ASKED BUT IS UNKNOWN. NO BLOOD TRANSFUSION WAS REQUIRED DUE TO THE LEAK. THERE WAS NO DAMAGE TO THE TUBING OBSERVED, JUST BONDING DID NOT BOND AND THE ZIP TIE WAS LOOSE. THERE WAS NO VISIBLE AIR IN THE SYSTEM/TUBING. THE EXACT LOCATION OF THE LEAK IN THE CUSTOM TUBING PACK SPECIFICATION WAS LINE 5.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17106378 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171308. One submission often covers many models, so check the product and lot details in each record.
- Z-0535-2018 · Jan 3, 2018 · TerminatedMedtronic Custom Perfusion Systems, indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY2V10R33Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page