Skip to content

NA BB10V62RRecalls & reported problems

Medtronic, Inc. · CUSTOM PACK BB10V62R FUSION PK · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000233372), received between Dec 15, 2025 and Feb 24, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K171308); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Delivered as Unsterile Product4
  • Packaging Problem3
  • Material Separation1
  • Misassembled During Installation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Delivered as Unsterile Product; Packaging Problem

    IT WAS REPORTED THAT THE OUTER PACKAGING FOR THESE CUSTOM TUBING PACKS WERE DAMAGED PRIOR TO USE. DEVICES IN THE SAME BOX/SHIPPING CONTAINERS WERE NOT DAMAGED. THE CUSTOM TUBING PACKS WERE NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION VIA ANALYSIS OF THE RETURNED DEVICES THAT THERE WAS ALSO DAMAGE TO THE STERILE BARRIER IN ALL OF THE CUSTOM TUBING PACKS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24428367 ↗
  • MalfunctionFeb 24, 2026Delivered as Unsterile Product; Packaging Problem

    IT WAS REPORTED THAT THE OUTER PACKAGING FOR THESE CUSTOM TUBING PACKS WERE DAMAGED PRIOR TO USE. DEVICES IN THE SAME BOX/SHIPPING CONTAINERS WERE NOT DAMAGED. THE CUSTOM TUBING PACKS WERE NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION VIA ANALYSIS OF THE RETURNED DEVICES THAT THERE WAS ALSO DAMAGE TO THE STERILE BARRIER IN ALL OF THE CUSTOM TUBING PACKS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24428341 ↗
  • MalfunctionFeb 24, 2026Delivered as Unsterile Product; Packaging Problem

    IT WAS REPORTED THAT THE OUTER PACKAGING FOR THESE CUSTOM TUBING PACKS WERE DAMAGED PRIOR TO USE. DEVICES IN THE SAME BOX/SHIPPING CONTAINERS WERE NOT DAMAGED. THE CUSTOM TUBING PACKS WERE NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION VIA ANALYSIS OF THE RETURNED DEVICES THAT THERE WAS ALSO DAMAGE TO THE STERILE BARRIER IN ALL OF THE CUSTOM TUBING PACKS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24428242 ↗
  • MalfunctionDec 15, 2025Delivered as Unsterile Product; Material Separation; Misassembled During Installation; Delivered as Unsterile Product; Material Separation; Misassembled During Installation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THERE WAS NO TAPE ON THE AV LOOP, CAUSING THE TOP COVER TO FALL OFF DURING SET UP, AND MAKING THE DEVICE NON STERILE AND HENCE UNUSABLE. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23804664 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171308. One submission often covers many models, so check the product and lot details in each record.

  • Z-0535-2018 · Jan 3, 2018 · Terminated
    Medtronic Custom Perfusion Systems, indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices

All Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page