BIO-MEDICUS® 96880-019Recalls & reported problems

Medtronic, Inc. · CANNULA 96880-019 BIO-MED VEN 19F 17L EA · Class 2

All models of Bio-Medicus (59)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000242343), received between Aug 21, 2024 and May 13, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Mechanical Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem2
  • Crack1
  • Material Deformation1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2025Crack; Mechanical Problem; Crack; Mechanical Problem

    IT WAS REPORTED THAT DURING USE OF A BIO-MEDICUS MULTI-STAGE FEMORAL VENOUS CANNULA WITH INSERTION KIT, A GUIDE WIRE (0.97MM X 180CM) FOR THE CANNULA WAS BEING USED TO CANNULATE A PATIENT FOR VENOVENOUS (VV) EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) (CV SURGERY) THROUGH THE RIGHT FEMORAL VEIN, RE-WIRING THROUGH A PREVIOUS DIALYSIS SITE. THE WIRE WAS INSERTED WITHOUT RESISTANCE INITIALLY, BUTRESISTANCE WAS NOTED DURING INSERTION AND REMOVAL. UPON REMOVAL, THE WIRE WAS FOUND TO BE SHREDDED AND SIGNIFICANTLY DAMAGED. ULTIMATELY, THIS DAMAGED THE PATIENT'S VENOUS SYSTEM (EXTERNAL ILIAC AND FEMORAL VEIN), RESULTING IN HEMORRHAGIC SHOCK THAT REQUIRED MASSIVE TRANSFUSION AND INTERVENTIONAL RADIOLOGY FOR MANAGEMENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22013931 ↗
  • InjuryAug 21, 2024Mechanical Problem; Material Deformation; Mechanical Problem; Material Deformation

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS MULTI-STAGE FEMORAL VENOUS CANNULA WITH INSERTION KIT, IT WAS REPORTED THAT IN THE INITIAL PROCEDURE, THE ATTENDING SURGEON WAS HOLDING DIGITAL PRESSURE ON THE INNOMINATE VEIN WHILE ANOTHER SURGEON ATTEMPTED RIGHT VENOUS CANNULIZATION. THE SURGEON COULD NOT ADVANCE THE WIRE OR REMOVE THE WIRE THROUGH THE SHEATH SO THE ENTIRE DEVICE WAS REMOVED. THE TEAM DID NOT VISUALISE THE WIRE PROTRUDING FROM THE TIP OF THE CANNULA. THE REMAINING WIRE WAS RETURNED TO THE WIRE HOOP TO CONTAIN ON THE FIELD. LATER IN THE CASE THE COUNT WAS OVER BY TWO NEEDLES AND A CHEST AND LEFT GROIN X-RAY WAS PERFORMED TO CLEAR THE COUNT. THERE WAS NO RSI SEEN BY THE RADIOLOGIST. THE CASE WAS POSTED TO REMOVE THE WIRE. THE WIRE WAS REMOVED BY VASCULAR SURGERY. THE SURGEON STATED THAT THE WIRE DID NOT NEED TO GO TO PATHOLOGY. THE CUSTOMER STATED THIS WAS AN EMERGENCY CASE. THE SCRUB TEAM DID RECOGNIZE THE WIRE LOOKED DIFFERENT BUT TO THE DEGREE THEY SPOKE UP IS NOT CLEAR. THE SURGEON WAS AWARE OF POTENTIAL WIRE PIECE RETAINED.

    Patient problems: Insufficient Information

    MAUDE report 20023427 ↗

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices