BIO-MEDICUS® 96880-025Recalls & reported problems

Medtronic, Inc. · CANNULA 96880-025 BIO-MED VEN 25F 17L EA · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00763000242367), received between Jul 7, 2022 and Jul 24, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation5
  • Mechanical Problem3
  • Obstruction of Flow2
  • Break1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Foreign Body In Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 24, 2025Obstruction of Flow; Material Deformation; Obstruction of Flow; Material Deformation

    IT WAS REPORTED THAT DURING USE OF THE BIO-MEDICUS MULTI-STAGE FEMORAL VENOUS CANNULA, THE SURGEON OBSERVED THAT THE CANNULA MATERIAL WAS SOFTER AT THE TUBING CONNECTION END COMPARED TO PREVIOUS EXPERIENCES WITH THE SAME TYPE OF CANNULA. THE END OF THE CANNULA KINKED, RESULTING IN OBSTRUCTED VENOUS FLOW. THE SURGEON WAS ABLE TO MANEUVER THE CANNULA SO THAT IT WAS NO LONGER KINKED, ALLOWING IT TO BE USED FOR THE REMAINDER OF THE CASE. THE SURGEON NOTED THAT THE CANNULA APPEARED TO BE MADE OF A DIFFERENT MATERIAL THAN PREVIOUSLY USED, AS PRIOR CANNULAE HAD NOT KINKED IN THIS MANNER. THE CUSTOMER QUERIED IF THE CANNULAE MATERIAL HAS CHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE BUT THE PATIENT¿S BLOOD WAS HEATED AND COOLED AND RUN THROUGH THE CANNULA. THE KINK WAS NOT IN THE LOCATION OF THE SUTURES. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS POSITIONED AT THE FEMORAL VEIN DURING USE. THE CANNULA WAS IN USE FOR 8 HOURS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22586881 ↗
  • MalfunctionJul 3, 2025Mechanical Problem; Material Deformation; Mechanical Problem; Material Deformation

    IT WAS REPORTED THAT PRIOR TO USE OF A BIO-MEDICUS MULTISTAGE FEMORAL VENOUS CANNULA, THE DEVICE FELT SOFTER THAN USUAL. THE SURGEON OPENED ANOTHER BOX TO CONFIRM IF THE CANNULA FELT SOFTER AT THE CLAMPING SITE THAN WHAT HE IS USED TO. THE SURGEON OBSERVED THE CANNULA WAS SOFTER THAN USUAL AND SHOWED HOW EASILY IT KINKED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22393273 ↗
  • MalfunctionJul 1, 2025Mechanical Problem; Obstruction of Flow; Material Deformation; Mechanical Problem; Obstruction of Flow; Material Deformation

    IT WAS REPORTED THAT DURING USE OF A BIOMEDICUS MULTISTAGE FEMORAL VENOUS CANNULA, THE CANNULA WAS OBSERVED TO BE VERY SOFT AT THE END, WHICH COULD BE CLAMPED, AND THE CANNULA KINKED, RESULTING IN RESTRICTED VENOUS DRAINAGE. THE SURGEON ADJUSTED THE INSERTION DEPTH OF THE CANNULA TO REMOVE THE KINK. THE SURGEON QUESTIONED WHETHER THERE HAD BEEN A MATERIAL CHANGE MAKING THE CANNULA SOFTER. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22368873 ↗
  • InjurySep 24, 2024Break; Material Deformation; Break; Material Deformation

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS MULTI-STAGE FEMORAL VENOUS CANNULAE WITH INSERTION KIT, IT WAS REPORTED THAT DURING DE-CANNULATION, THE TIP OF THE CANNULA BROKE IN THE VESSEL, AND EMIGRATE RIGHT UP IN THE ATRIUM. USE OF DEVICE WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20289786 ↗
  • MalfunctionJul 3, 2023Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE, THE CUSTOMER REPORTED THE BIO-MEDICUS MULTISTAGE FEMORAL VENOUS CANNULA'S GUIDEWIRE WOULD NOT PASS. THE DEVICE WAS REPLACED TO COMPLETE THE CASE. THERE WAS NO ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17245253 ↗
  • MalfunctionFeb 28, 2023Material Twisted/Bent; Material Twisted/Bent

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS MULTI-STAGE FEMORAL VENOUS CANNULA WITH INSERTION KIT, NO ISSUE WITH THE GUIDEWIRE WAS NOTICED WHEN REMOVING THE WIRE FROM THE PACKAGING. WHEN THE GUIDEWIRE WAS DEPLOYED, THE TIP WAS BENDING. THE CUSTOMER CHECKED THE WIRE AND FOUND THAT IT WAS BENDING AT THE TIP WHICH THE CUSTOMER STATED WAS NOT NORMAL. THE WIRE WAS STILL ABLE TO BE USED AND WHEN THE CANNULA BODY WAS DEPLOYED OVER THE WIRE THERE WAS NO FURTHER ISSUE USING THE CANNULA WITH THE WIRE. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE DAMAGE WAS NOT NOTICED WHEN THE DEVICE WAS REMOVED FROM THE PACKAGING. THE GUIDEWIRE WAS NOT REMOVED WHILE THE INTRODUCER NEEDLE REMAINED IN PLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16452943 ↗
  • MalfunctionJul 7, 2022Material Deformation; Material Deformation

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A BIO-MEDICUS MULTI-STAGE FEMORAL VENOUS CANNULA WITH INSERTION KIT, IT WAS REPORTED THAT THE TIP OF THE INTRODUCER WAS BENT. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE CUSTOMER STATED THAT THE CANNULA WAS NORMAL AND THERE WAS NO HARM TO THE PATIENT. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULA WAS USED ACCORDING TO IFU STATEMENTS, NO HEATING OR COOLING WAS APPLIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14945074 ↗

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices