NA BB8C43R12Recalls & reported problems
Medtronic, Inc. · CUSTOM PACK BB8C43R12 INVLOOP · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00763000253882), received between Aug 14, 2024 and Jan 21, 2026 — classified as 1 injury, 9 malfunctions.
The most frequently reported specific device problem is “Delivered as Unsterile Product” (8 reports).
1 other recall cover devices cleared under the same FDA submission (K171308); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Delivered as Unsterile Product8
- Crack1
- Material Deformation1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 21, 2026Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THESE CUSTOM TUBING PACKS, IT WAS REPORTED THAT THE TABLE TRAYS WERE CRACKED. THE CRACKS WERE APPROXIMATELY 6 INCHES. A COMPLETE BREAK DID NOT OCCUR. THE STERILE BARRIER WAS DAMAGED. THERE WERE NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER DAMAGED. THE DEVICES WERE REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24128242 ↗MalfunctionJan 21, 2026Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THESE CUSTOM TUBING PACKS, IT WAS REPORTED THAT THE TABLE TRAYS WERE CRACKED. THE CRACKS WERE APPROXIMATELY 6 INCHES. A COMPLETE BREAK DID NOT OCCUR. THE STERILE BARRIER WAS DAMAGED. THERE WERE NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER DAMAGED. THE DEVICES WERE REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24128213 ↗MalfunctionJan 21, 2026Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THESE CUSTOM TUBING PACKS, IT WAS REPORTED THAT THE TABLE TRAYS WERE CRACKED. THE CRACKS WERE APPROXIMATELY 6 INCHES. A COMPLETE BREAK DID NOT OCCUR. THE STERILE BARRIER WAS DAMAGED. THERE WERE NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER DAMAGED. THE DEVICES WERE REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24128202 ↗MalfunctionJan 21, 2026Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THESE CUSTOM TUBING PACKS, IT WAS REPORTED THAT THE TABLE TRAYS WERE CRACKED. THE CRACKS WERE APPROXIMATELY 6 INCHES. A COMPLETE BREAK DID NOT OCCUR. THE STERILE BARRIER WAS DAMAGED. THERE WERE NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER DAMAGED. THE DEVICES WERE REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24128189 ↗MalfunctionJan 21, 2026Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THESE CUSTOM TUBING PACKS, IT WAS REPORTED THAT THE TABLE TRAYS WERE CRACKED. THE CRACKS WERE APPROXIMATELY 6 INCHES. A COMPLETE BREAK DID NOT OCCUR. THE STERILE BARRIER WAS DAMAGED. THERE WERE NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER DAMAGED. THE DEVICES WERE REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24128183 ↗MalfunctionJan 21, 2026Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THESE CUSTOM TUBING PACKS, IT WAS REPORTED THAT THE TABLE TRAYS WERE CRACKED. THE CRACKS WERE APPROXIMATELY 6 INCHES. A COMPLETE BREAK DID NOT OCCUR. THE STERILE BARRIER WAS DAMAGED. THERE WERE NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER DAMAGED. THE DEVICES WERE REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24128164 ↗MalfunctionJan 20, 2026Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THESE CUSTOM TUBING PACKS, IT WAS REPORTED THAT THE TABLE TRAYS WERE CRACKED. THE CRACKS WERE APPROXIMATELY 6 INCHES. A COMPLETE BREAK DID NOT OCCUR. THE STERILE BARRIER WAS DAMAGED. THERE WERE NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER DAMAGED. THE DEVICES WERE REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24117260 ↗MalfunctionJan 20, 2026Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THESE CUSTOM TUBING PACKS, IT WAS REPORTED THAT THE TABLE TRAYS WERE CRACKED. THE CRACKS WERE APPROXIMATELY 6 INCHES. A COMPLETE BREAK DID NOT OCCUR. THE STERILE BARRIER WAS DAMAGED. THERE WERE NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER DAMAGED. THE DEVICES WERE REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24116509 ↗InjuryOct 2, 2024Obstruction of Flow; Material Deformation; Obstruction of Flow; Material Deformation
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, IT WAS REPORTED THAT THERE WAS A FLOW ISSUE WITH NO FLOW THROUGH THE VALVE AND IT WAS REPORTED THAT THERE WAS A BROKEN 1 WAY VALVE. THE PATIENT HAD PREVIOUSLY BEEN ON HEPARIN ANTI-COAGULANT THERAPY. THE ANTI-COAGULATION WAS ASSESSED WITH THE HMS PLUS INSTRUMENT. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE ONE WAY VALVE WAS NOT WORKING AS NO FLOW WAS PASSING THROUGH IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20352757 ↗MalfunctionAug 14, 2024Crack; Crack
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED A CRACK/BREAK LOCATED ON THE VENT ON THE SUCKER LINE. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE ISSUE WAS A PARTIAL CRACK/BREAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19979631 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171308. One submission often covers many models, so check the product and lot details in each record.
- Z-0535-2018 · Jan 3, 2018 · TerminatedMedtronic Custom Perfusion Systems, indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY11R00R6Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗