SimuForm™ 7800RR30 (UDI 00763000271497)Recalls & reported problems
Medtronic, Inc. · Ring 7800RR30 SimuForm SR 30mm GLB · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00763000271497), received between Jun 7, 2022 and Jul 3, 2026 — classified as 11 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information3
- Intravalvular regurgitation3
- Adverse Event Without Identified Device or Use Problem2
- Patient Device Interaction Problem2
- Fluid/Blood Leak1
- Inadequacy of Device Shape and/or Size1
- Incomplete Coaptation1
Patient problems
- Insufficient Information5
- Mitral Valve Insufficiency/ Regurgitation3
- Non specific EKG/ECG Changes3
- Valvular Insufficiency/ Regurgitation1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryJul 3, 2026Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM SIMUFORM SR ANNULOPLASTY HEART RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS REPORTED AS MITRAL REGURGITATION (MR). AFTER THE RING WAS IMPLANTED, A SALINE TEST SHOWED POOR REPAIR WITH REGURGITATION, AND THE DEVICE WAS THEREFORE REMOVED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Mitral Valve Insufficiency/ Regurgitation
MAUDE report 25703818 ↗InjuryJun 5, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE IMPLANT OF THIS 30 MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 25407717 ↗InjuryMar 16, 2026Fluid/Blood Leak; Incomplete Coaptation
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE REASON FOR REPLACEMENT OF THIS MITRAL ANNULOPLASTY RING WAS DUE TO COAPTATION. THE DEVICE WAS FULLY SUTURED AND TESTED TO REMOVAL. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Valvular Insufficiency/ Regurgitation
MAUDE report 24606176 ↗InjuryJan 6, 2026Insufficient Device Problem Information; Intravalvular regurgitation
ADDITIONAL INFORMATION REPORTED THAT THE REASON FOR INTERVENTION WAS MITRAL REGURGITATION (MR) AND THE ANNULOPLASTY PRODUCT WAS FULLY SUTURED AND TESTED PRIOR TO REMOVAL. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation
MAUDE report 23974815 ↗InjuryNov 11, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT 8 DAYS POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY HEART RING, IT WAS EXPLANTED AND REPLACED WITH A 29MM MITR AL BIOPROSTHETIC VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 23522470 ↗InjuryAug 1, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM SIMUFORM ANNULOPLASTY HEART RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN COMPETITORS PRODUCT. THE REASON FOR THE REPLACEMENT WAS REPORTED AS REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Mitral Valve Insufficiency/ Regurgitation
MAUDE report 22684854 ↗InjuryJul 15, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 12 DAYS POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME (SSS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 19743038 ↗InjuryJul 10, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 6 MONTHS POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING AND 23MM AORTIC BIOPROSTHETIC VALVE, A REPLACEMENT PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME (SSS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 19708489 ↗InjuryMar 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION THAT 1 DAY POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, AN ELECTROCARDIOGRAM (ECG) NOTED 2ND DEGREE ATRIOVENTRICULAR (AV) BLOCK . IT WAS REPORTED THAT ONE DAY LATER, THE PATIENT WAS TREATED BY STOPPING THEIR MEDICATION (AMIODARONE) AND PACED WITH AN EPICARDIAL PACEMAKER FOR 3 DAYS. IT WAS REPORTED THAT THE EVENT RESOLVED AND THE PATIENT RECOVERED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 18906177 ↗InjuryDec 7, 2022Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 28MM MITRAL ANNULOPLASTY RING OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 15933969 ↗InjuryJun 7, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT ONE DAY POST IMPLANT OF THIS 30MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A NON-MEDTRONIC PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 14622391 ↗
Other RING, ANNULOPLASTY devices
- CG FUTURE™ 638BL32Medtronic, Inc. · RING, ANNULOPLASTY1 death71 reports
- PROFILE 3D® 680R28Medtronic, Inc. · RING, ANNULOPLASTY1 death67 reports
- CG FUTURE™ 638BL30Medtronic, Inc. · RING, ANNULOPLASTY1 death65 reports
- CG FUTURE™ 638RL32Medtronic, Inc. · RING, ANNULOPLASTY59 reports
- Tri-Ad™ 900SFC228Medtronic, Inc. · RING, ANNULOPLASTY4 deaths56 reports
- CG FUTURE™ 638BL34Medtronic, Inc. · RING, ANNULOPLASTY51 reports
- Contour 3D® 690R30Medtronic, Inc. · RING, ANNULOPLASTY1 death50 reports
- CG FUTURE™ 638RL30Medtronic, Inc. · RING, ANNULOPLASTY1 death47 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page