SimuForm™ 7800RR32 (UDI 00763000271503)Recalls & reported problems
Medtronic, Inc. · Ring 7800RR32 SimuForm SR 32mm GLB · Class 2 · Implant
All models of SimuForm (15)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00763000271503), received between Jun 7, 2022 and Jul 9, 2026 — classified as 12 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information4
- Intravalvular regurgitation4
- Incomplete Coaptation3
- Patient Device Interaction Problem3
- Adverse Event Without Identified Device or Use Problem2
- Inadequacy of Device Shape and/or Size1
- Perivalvular Leak1
- Structural Problem1
Patient problems
- Insufficient Information9
- Mitral Valve Insufficiency/ Regurgitation6
- Heart Block1
- Heart Failure/Congestive Heart Failure1
- Non specific EKG/ECG Changes1
- Tachycardia1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 9, 2026Incomplete Coaptation
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE ANNULOPLASTY PRODUCT MALFUNCTION DID NOT MALFUNCTION AND IT WAS FULLY SUTURED AND TESTED PRIOR TO REMOVAL. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 25759971 ↗MalfunctionApr 20, 2026Structural Problem; Incomplete Coaptation
ADDITIONAL INFORMATION RECEIVED CLARIFIED THAT THE REASON FOR INTERVENTION WAS DUE TO COAPTATION. THE ANNULOPLASTY PRODUCT WAS FULLY SUTURED AND TESTED PRIOR TO REMOVAL AND THE DEVICE DID MALFUNCTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 24925741 ↗InjuryMar 12, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE IMPLANT OF THIS 32MM ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 24579202 ↗MalfunctionJan 8, 2026Incomplete Coaptation
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 32MM SIMUFORM HEART RING, IT WAS EXPLANTED AND REPLACED WITH ANOTHER MANUFACTURER'S PRODUCT. THE REASON FOR THE REPLACEMENT WAS REPORTED THAT AFTER IMPLANTATION, IMMEDIATE POSTOPERATIVE TRANSESOPHAGEAL ECHOCARDIOGRAPHY REVEALED THAT THE ANTERIOR RING DID NOT MOVE WITH THE ANTERIOR MITRAL LEAFLET, AFFECTING LEAFLET COAPTATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 24002355 ↗InjuryOct 15, 2025Intravalvular regurgitation; Intravalvular regurgitation
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE ANNULOPLASTY RING WAS FULLY SUTURED AND TESTED PRIOR TO REMOVAL. THE REPLACEMENT DEVICE WAS ALSO CONFIRMED TO BE THE SAME 32MM SIZE OF THE ANNULOPLASTY RING. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Mitral Valve Insufficiency/ Regurgitation
MAUDE report 23278097 ↗InjurySep 11, 2025Intravalvular regurgitation; Intravalvular regurgitation
ADDITIONAL INFORMATION WAS RECEIVED THAT THE VALVE DOES NOT APPEAR TO HAVE ANY MALFUNCTIONS OR DAMAGES PRIOR TO USE AND WAS WAS FULLY SUTURED AND TESTED BEFORE REMOVAL. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Mitral Valve Insufficiency/ Regurgitation
MAUDE report 23027089 ↗InjuryJun 26, 2025Perivalvular Leak; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Perivalvular Leak; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT A 32MM MITRAL ANNULOPLASTY RING WAS IMPLANTED AND EXPLANTED ON THE SAME DAY DURING A PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation
MAUDE report 22338565 ↗InjurySep 5, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 2 YEARS AND 1 MONTH POST IMPLANT OF THIS 32MM MITRAL ANNULOPLASTY RING, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS COMPLETE HEART BLOCK, ATRIAL ARRHYTHMIAS AND SINUS NODE DYSFUNCTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Heart Block; Non specific EKG/ECG Changes
MAUDE report 20148156 ↗InjuryMay 27, 2024Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 1 YEAR POST IMPLANT OF THIS 32MM MITRAL ANNULOPLASTY RING, IT WAS REPORTED THAT AORTIC REGURGITATION (AR) AND MITRAL REGURGITATION (MR) WERE OBSERVED ON A ECHOCARDIOGRAM. A TRANSCATHETER AORTIC VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Mitral Valve Insufficiency/ Regurgitation
MAUDE report 19400025 ↗InjuryApr 25, 2024Patient Device Interaction Problem; Intravalvular regurgitation; Patient Device Interaction Problem; Intravalvular regurgitation
ADDITIONAL INFORMATION RECEIVED FROM THE PHYSICIAN/AUTHOR PROVIDED THE DEVICE UNIQUE DEVICE IDENTIFIER (SERIAL NUMBER (B)(6) AND THE PATIENT'S WEIGHT OF 54 KG. NO FURTHER DETAILS WERE PROVIDED.
Patient problems: Heart Failure/Congestive Heart Failure; Mitral Valve Insufficiency/ Regurgitation
MAUDE report 19188636 ↗InjuryMar 18, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 32MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 18922026 ↗InjuryJan 19, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 32MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 29MM BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS REPORTED AS DUE TO THE PATIENT'S ANATOMY AND MILD TO MODERATE MITRAL REGURGITATION AT THE MEDIAL COMMISSURE. IT WAS NOTED THAT DURING THIS PROCEDURE, AN ASCENDING AORTA REPLACEMENT WITH REPLACEMENT OF THE NON-CORONARY SINUS AND AN AORTIC VALVE REPLACEMENT WAS ALSO PERFORMED. IT WAS REPORTED THAT POST IMPLANT OF THE MITRAL RING AND OF ANOTHER MANUFACTURERS VASCULAR GRAFT AND AORTIC BIOPROSTHETIC VALVE, THE PATIENT EXPERIENCED A JUNCTIONAL RHYTHM FOLLOWED BY VENTRICULAR TACHYCARDIA (V-TACH) WHICH WAS TREATED WITH DEFIBRILLATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Mitral Valve Insufficiency/ Regurgitation; Tachycardia
MAUDE report 18543410 ↗InjuryDec 7, 2022Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 32MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 30MM MITRAL ANNULOPLASTY RING OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 15933911 ↗InjuryAug 30, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 32MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 30MM OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 15317566 ↗InjuryJun 7, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 32MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 30MM MITRAL ANNULOPLASTY RING OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 14622320 ↗
Other RING, ANNULOPLASTY devices
- CG FUTURE™ 638BL32Medtronic, Inc. · RING, ANNULOPLASTY1 death71 reports
- PROFILE 3D® 680R28Medtronic, Inc. · RING, ANNULOPLASTY1 death67 reports
- CG FUTURE™ 638BL30Medtronic, Inc. · RING, ANNULOPLASTY1 death65 reports
- CG FUTURE™ 638RL32Medtronic, Inc. · RING, ANNULOPLASTY59 reports
- Tri-Ad™ 900SFC228Medtronic, Inc. · RING, ANNULOPLASTY4 deaths56 reports
- CG FUTURE™ 638BL34Medtronic, Inc. · RING, ANNULOPLASTY51 reports
- Contour 3D® 690R30Medtronic, Inc. · RING, ANNULOPLASTY1 death50 reports
- CG FUTURE™ 638BL28Medtronic, Inc. · RING, ANNULOPLASTY1 death47 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page