NA TL3A08R49Recalls & reported problems
Medtronic, Inc. · CUSTOM PACK TL3A08R49 P / T · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000280208), received between Jun 16, 2023 and Jun 21, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K171308); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 21, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT A CONNECTOR WITH LUER LEAKED THROUGHOUT THE CASE. THE CUSTOM TUBING PACK SPECIFICATIONS SHOW THAT THIS CONNECTOR SHOULD BE SOLVENT BONDED AND THE CUSTOMER WAS SKEPTICAL THAT IT WAS, BECAUSE IT WAS LEAKING AND AFTER ONE OF THE CASES THAT HAD A LEAK THE PERFUSIONIST WAS ABLE TO "PUSH THE TUBING FURTHER ONTO CONNECTOR". THIS CONNECTOR IS ON LINE 7 OF THE CUSTOM TUBING PACK SPECIFICATIONS. THE CUSTOM PACK WAS USED FOR THE ENTIRETY OF THE CASE. APPROXIMATELY 50 CC OF BLOOD WAS LOST DUE TO THE LEAK. NO BLOOD TRANSFUSION WAS REQUIRED. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17169805 ↗MalfunctionJun 21, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT A CONNECTOR WITH LUER LEAKED THROUGHOUT THE CASE. THE CUSTOM TUBING PACK SPECIFICATIONS SHOW THAT THIS CONNECTOR SHOULD BE SOLVENT BONDED AND THE CUSTOMER WAS SKEPTICAL THAT IT WAS, BECAUSE IT WAS LEAKING AND AFTER ONE OF THE CASES THAT HAD A LEAK THE PERFUSIONIST WAS ABLE TO "PUSH THE TUBING FURTHER ONTO CONNECTOR". THIS CONNECTOR IS ON LINE 7 OF THE CUSTOM TUBING PACK SPECIFICATIONS. THE CUSTOM PACK WAS USED FOR THE ENTIRETY OF THE CASE. APPROXIMATELY 50 CC OF BLOOD WAS LOST DUE TO THE LEAK. NO BLOOD TRANSFUSION WAS REQUIRED. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17169767 ↗MalfunctionJun 21, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT A CONNECTOR WITH LUER LEAKED THROUGHOUT THE CASE. THE CUSTOM TUBING PACK SPECIFICATIONS SHOW THAT THIS CONNECTOR SHOULD BE SOLVENT BONDED AND THE CUSTOMER WAS SKEPTICAL THAT IT WAS, BECAUSE IT WAS LEAKING AND AFTER ONE OF THE CASES THAT HAD A LEAK THE PERFUSIONIST WAS ABLE TO "PUSH THE TUBING FURTHER ONTO CONNECTOR". THIS CONNECTOR IS ON LINE 7 OF THE CUSTOM TUBING PACK SPECIFICATIONS. THE CUSTOM PACK WAS USED FOR THE ENTIRETY OF THE CASE. APPROXIMATELY 50 CC OF BLOOD WAS LOST DUE TO THE LEAK. NO BLOOD TRANSFUSION WAS REQUIRED. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17169740 ↗MalfunctionJun 16, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT A CONNECTOR WITH LUER LEAKED THROUGHOUT THE CASE. THE CUSTOM TUBING PACK SPECIFICATION SHOW THAT THIS CONNECTOR SHOULD BE SOLVENT BONDED AND THE CUSTOMER WAS SKEPTICAL THAT IT WAS, BECAUSE IT WAS LEAKING AND AFTER ONE OF THE CASES THAT HAD A LEAK THE PERFUSIONIST WAS ABLE TO "PUSH THE TUBING FURTHER ONTO CONNECTOR". THIS CONNECTOR IS ON LINE 7 OF THE CUSTOM TUBING PACK SPECIFICATION. THE CUSTOM PACK WAS USED FOR THE ENTIRETY OF THE CASE. APPROXIMATELY 50 CC OF BLOOD WAS LOST DUE TO THE LEAK. NO BLOOD TRANSFUSION WAS REQUIRED. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17142442 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171308. One submission often covers many models, so check the product and lot details in each record.
- Z-0535-2018 · Jan 3, 2018 · TerminatedMedtronic Custom Perfusion Systems, indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY11R00R6Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page