NA BB10S02R6Recalls & reported problems
Medtronic, Inc. · CUSTOM PACK BB10S02R6 3/8 ECCPK · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000299002), received between Aug 10, 2022 and Mar 25, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
1 other recall cover devices cleared under the same FDA submission (K171308); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2025Material Separation; Material Separation
IT WAS REPORTED THAT PRIOR TO USE A CUSTOM TUBING PACK HAD A LOOSE/SEPARATED COMPONENT AT THE PIGTAIL CONNECTION AT THE LUER. THE LO OSE/SEPARATED COMPONENT DID NOT FALL INTO THE PATIENT. THE DEVICE WAS REPLACED TO COMPLETE THE CASE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21686124 ↗MalfunctionAug 10, 2022Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, THE CUSTOMER REPORTED THAT DURING SET UP AND PRIMING, A LARGE PUDDLE OF CRYSTALLOID WAS FOUND BENEATH THE CIRCUIT ON THE FLOOR THE NEXT DAY. THE LEAK WAS TRACED TO THE PRE-BYPASS FILTER. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION STATING THAT THE LEAK ORIGINATED FROM THE COMPONENT AND THERE WAS A SMALL DRIP COMING OUT OF THE SEAM AT THE TOP OF THE FILTER. THERE WERE NO VISIBLE AIR IN THE SYSTEM/TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15201735 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171308. One submission often covers many models, so check the product and lot details in each record.
- Z-0535-2018 · Jan 3, 2018 · TerminatedMedtronic Custom Perfusion Systems, indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY11R00R6Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page