Skip to content

StealthStation® Cranial 9735587 (UDI 00763000299927)Recalls & reported problems

Medtronic Navigation, Inc. · SW KIT 9735587 STEALTH CRANIAL PLNG STN · Class 2

All models of StealthStation Cranial (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K153660); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Apr 11, 2023Open, Classified

    In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury

    Z-1666-2023 · Root cause: Software design · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K153660. One submission often covers many models, so check the product and lot details in each record.

  • Z-1995-2020 · Aug 9, 2019 · Terminated
    Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS License or StealthStation S8 Software with Stealth DBS Licen
  • Z-0552-2019 · Sep 26, 2018 · Terminated
    StealthStation Cranial and Synergy Cranial Depth Gauge: StealthStation Cranial Optical Pocket Guide - Instructions for cranial procedures using optical localization StealthStatio

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page