2 open recalls name at least one StealthStation Cranial model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls
3

We found 13 models of StealthStation Cranial by Medtronic Navigation, Inc. in the FDA's device registry (GUDID). Together they are named in 2 adverse event reports in the FDA's MAUDE database, received between Mar 2, 2017 and May 5, 2017 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (2 reports).

3 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

StealthStation Cranial recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Apr 11, 2023Open, Classified

    In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury

    Z-1666-2023 · Root cause: Software design · names 5 of these models · Recall details · FDA record ↗

  2. Apr 11, 2023Open, Classified

    During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue injury

    Z-1630-2023 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  3. Sep 26, 2018Terminated

    An incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.

    Z-0552-2019 · Root cause: Device Design · names 6 of these models · Recall details · FDA record ↗

Most reported StealthStation Cranial problems

Device problems

  • Computer Software Problem2
  • Communication or Transmission Problem1

Patient problems

  • No Patient Involvement2

StealthStation Cranial models

13 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent StealthStation Cranial adverse event reports