NA 9734683Recalls & reported problems
Medtronic Navigation, Inc. · TRACKER 9734683 NAVLOCK UNIVERSAL ORANGE · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00763000300265), received between May 12, 2023 and Jul 7, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K171267); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Sensing Problem1
- Imprecision1
- Material Fragmentation1
- Mechanical Problem1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 7, 2025Material Fragmentation; Mechanical Problem; Material Fragmentation; Mechanical Problem
IT WAS REPORTED THAT IT GOT DEFORMED DURING PROBING. THE ISSUE WAS DISCOVERED DURING USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22419202 ↗MalfunctionFeb 26, 2025Use of Device Problem; Use of Device Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT THE POST OF THE INSTRUMENT TRACKER WITH A SPHERE ATTACHED TO IT FELL OFF MID-CASE. THE REPORTED CAUSE OF THE EVENT WAS THAT THE INSTRUMENT WAS BEING USED FOR NAVIGATION OF A THORACIC PROBE AND POST OF TRACKER FELL OFF DURING USE. THERE WAS NO PHYSICAL IMPACT TO THE TRACKER WHEN THIS OCCURRED. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO REPORTED IMPACT ON PATIENT OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21471460 ↗MalfunctionAug 27, 2024Device Sensing Problem; Device Sensing Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT WHEN THEY WERE USING THE TRACKERS THE SPHERES WERE FLICKERING IN AND OUT WHILE NAVIGATING. BOTH THE ORANGE AND GRAY TRACKER, AND BOTH HAD A GEOMETRY ERROR OF .49MM. THE PURPLE AND PASSIVE PLANAR PROBE WERE WORKING AS EXPECTED. THE REPRESENTATIVE RAN A STATUS CHECK ON THE CAMERA AND EVERYTHING WAS WORKING AS EXPECTED. THERE WAS NO REPORTED DELAY AND NO REPORTED IMPACT TO PATIENT OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20077533 ↗MalfunctionMay 12, 2023Imprecision; Imprecision
MEDTRONIC RECEIVED INFORMATION REGARDING A GUIDANCE SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT WHEN TRAINING USING THE NAVIGATION THE MANUFACTURER REPRESENTATIVE FOUND THE TRACKER ASSEMBLED ON THE TAP OR DRIVER WERE NOT ACCURATE BECAUSE THE EXTENSION LINE AND TOOLS DISPLAYED ON SCREEN WERE NOT ALONG THE CENTER OF THE TRAJECTORY. THE BLUE AND BLACK TRACKER WERE GOOD THOUGH. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16924203 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171267. One submission often covers many models, so check the product and lot details in each record.
- Z-1693-2020 · Aug 28, 2019 · TerminatedNavLock Tracker *** Rx Only
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page