NA 9734590Recalls & reported problems

Medtronic Navigation, Inc. · TRACKER 9734590 NAVLOCK UNIVERSAL GREY · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000300401), received between Jan 24, 2025 and Apr 20, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K171267); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam2
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 20, 2026Material Fragmentation

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED PRIOR TO A PROCEDURE. IT WAS REPORTED THAT THE POST OF THE GREY TRACKER BROKE DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24933075 ↗
  • MalfunctionMar 6, 2026Mechanical Jam

    ADDITIONAL INFORMATION AS RECEIVED. IT WAS REPORTED THAT THE ISSUE WAS LIKELY CAUSED DUE TO PROLONGED USE THE INSTRUMENTS WITH NORMAL WEAR AND TEAR OF PARTS. THIS COULD NOT BE CONFIRMED AS MULTIPLE NEW INSTRUMENTS HAD PREVIOUSLY STOPPED WORKING AFTER BEING USED IN ONLY 1 OR 2 CASES. THE MANUFACTURING REPRESENTATIVE (REP) BELIEVED THAT THERE WAS SOMETHING WITH A DISPOSABLE 2.4 DRILL BIT BEING THE MAIN CAUSE OR CONTRIBUTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24531568 ↗
  • MalfunctionJan 24, 2025Mechanical Jam; Mechanical Jam

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM USED IN A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT THE GREY TRACKER DID NOT ROTATE AND COULD NOT BE REMOVED. AFTER TAPPING WITH A HAMMER, THE TRACKER WAS THEN REMOVED. THE OPERATION WAS PERFORMED BY USING ANOTHER COLORED TRACKER. NO HEALTH DAMAGE. THERE WAS A SURGICAL DELAY OF LESS THAN 1 HOUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21232263 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171267. One submission often covers many models, so check the product and lot details in each record.

  • Z-1693-2020 · Aug 28, 2019 · Terminated
    NavLock Tracker *** Rx Only

Other Orthopedic stereotaxic instrument devices