Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000307790), received between Jun 2, 2022 and Jun 6, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Migration or Expulsion of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Migration or Expulsion of Device2
- Use of Device Problem2
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Embolism/Embolus1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 6, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT A MICRO VASCULAR PLUG WAS IMPLANTED 54 DAYS POST EXPIRY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22159587 ↗MalfunctionMar 19, 2024Use of Device Problem; Use of Device Problem
A PHYSICIAN IMPLANTED A MICRO VASCULAR PLUG, THE DEVICE HAD PASSED IT'S USE BY DATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18928871 ↗MalfunctionApr 4, 2023Migration or Expulsion of Device; Migration or Expulsion of Device
A PHYSICIAN USED A MICRO VASCULAR PLUG FOR TREATMENT OF A PATIENT. IT WAS REPORTED THE PLUG MIGRATED DISTALLY AFTER DEPLOYMENT. THE CATHETER WAS FLUSHED BEFORE INSERTING THE PLUG. CONTINUOUS FLUSH WAS MAINTAINED DURING THE PROCEDURE. THE CATHETER WAS 2.7 NON MEDTRONIC CATHETER . TARGET VESSEL DIAMETER WAS 3MM. THE ANATOMY WAS NON-TORTUOUS, QUITE STRAIGHT FORWARD. IT WAS A STRAIGHT VESSEL AND THE PLUG WAS FULLY OPEN AND DEPLOYED BEFORE DETACHMENT. THE PLUG REMAINS IN THE PATIENT. NO INJURY TO THE PATIENT AS A RESULT OF THE DIFFICULTY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16671918 ↗MalfunctionSep 24, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
PHYSICIAN WAS ATTEMPTING TO USE A MICROVASCULAR PLUG DURING PROCEDURE. RESISTANCE WAS FELT WHILE ADVANCING THE PLUG THROUGH THE INTRODUCER SHEATH OR CATHETER. IT WAS REPORTED THAT PHYSICIAN ADVANCED MVP-5Q INTO TERUMO 2.7F 130CM MICROCATHETER AND IT DEPLOYED JUST AFTER HUB. THE DELIVERY WIRE WAS NOT MANIPULATED OR ROTATED BEFORE OR DURING DELIVERY OF THE PLUG THROUGH THE CATHETER. THE USER DID NOT ROTATE THE TORQUER IN THE INCORRECT DIRECTION ACCIDENTALLY DURING PREPARATION. ADVANCEMENT WAS SMOOTH THROUGH INTRODUCER SHEATH THEN THE PLUGGED JAMMED IN MICROCATHETER PROXIMALLY. THE REASON THE MVP GOT STUCK WAS BECAUSE THE PROGREAT 2.7F IS TOO SMALL FOR THE MVP5-Q. THE MICRO CATHETER WAS IN THE PATIENT WHEN PLUG DETACHED. PUSHABLE COILS WERE USED TO EMBOLIZE AND COMPLETE THE PROCEDURE. NO INTERVENTION REQUIRED FOR DEVICE REMOVAL. NO VESSEL DAMAGE NOTED. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15475260 ↗InjuryJun 2, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
DURING A PROCEDURE MICRO VASCULAR PLUG WAS IMPLANTED. IT WAS REPORTED THAT THE PATIENT WAS DISCHARGED AND SENT HOME. PATIENT WAS TOLD THAT THE DEVICE HAD DETACHED AND WAS FLOATING AROUND INSIDE. THE MVP EMBOLIZED AND THEN MIGRATED INTO THE PULMONARY ARTERY. THE MVP DEVICE WAS EXPLANTED.
Patient problems: Embolism/Embolus
MAUDE report 14573233 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page