Skip to content

MVP™ MVP-5QRecalls & reported problems

Medtronic, Inc. · PLUG MVP-5Q MICRO VASCULAR · Class 2 · Implant

All models of MVP (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00763000307790), received between Jun 2, 2022 and Jun 6, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration or Expulsion of Device2
  • Use of Device Problem2
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Embolism/Embolus1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 6, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT A MICRO VASCULAR PLUG WAS IMPLANTED 54 DAYS POST EXPIRY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22159587 ↗
  • MalfunctionMar 19, 2024Use of Device Problem; Use of Device Problem

    A PHYSICIAN IMPLANTED A MICRO VASCULAR PLUG, THE DEVICE HAD PASSED IT'S USE BY DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18928871 ↗
  • MalfunctionApr 4, 2023Migration or Expulsion of Device; Migration or Expulsion of Device

    A PHYSICIAN USED A MICRO VASCULAR PLUG FOR TREATMENT OF A PATIENT. IT WAS REPORTED THE PLUG MIGRATED DISTALLY AFTER DEPLOYMENT. THE CATHETER WAS FLUSHED BEFORE INSERTING THE PLUG. CONTINUOUS FLUSH WAS MAINTAINED DURING THE PROCEDURE. THE CATHETER WAS 2.7 NON MEDTRONIC CATHETER . TARGET VESSEL DIAMETER WAS 3MM. THE ANATOMY WAS NON-TORTUOUS, QUITE STRAIGHT FORWARD. IT WAS A STRAIGHT VESSEL AND THE PLUG WAS FULLY OPEN AND DEPLOYED BEFORE DETACHMENT. THE PLUG REMAINS IN THE PATIENT. NO INJURY TO THE PATIENT AS A RESULT OF THE DIFFICULTY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16671918 ↗
  • MalfunctionSep 24, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    PHYSICIAN WAS ATTEMPTING TO USE A MICROVASCULAR PLUG DURING PROCEDURE. RESISTANCE WAS FELT WHILE ADVANCING THE PLUG THROUGH THE INTRODUCER SHEATH OR CATHETER. IT WAS REPORTED THAT PHYSICIAN ADVANCED MVP-5Q INTO TERUMO 2.7F 130CM MICROCATHETER AND IT DEPLOYED JUST AFTER HUB. THE DELIVERY WIRE WAS NOT MANIPULATED OR ROTATED BEFORE OR DURING DELIVERY OF THE PLUG THROUGH THE CATHETER. THE USER DID NOT ROTATE THE TORQUER IN THE INCORRECT DIRECTION ACCIDENTALLY DURING PREPARATION. ADVANCEMENT WAS SMOOTH THROUGH INTRODUCER SHEATH THEN THE PLUGGED JAMMED IN MICROCATHETER PROXIMALLY. THE REASON THE MVP GOT STUCK WAS BECAUSE THE PROGREAT 2.7F IS TOO SMALL FOR THE MVP5-Q. THE MICRO CATHETER WAS IN THE PATIENT WHEN PLUG DETACHED. PUSHABLE COILS WERE USED TO EMBOLIZE AND COMPLETE THE PROCEDURE. NO INTERVENTION REQUIRED FOR DEVICE REMOVAL. NO VESSEL DAMAGE NOTED. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15475260 ↗
  • InjuryJun 2, 2022Migration or Expulsion of Device; Migration or Expulsion of Device

    DURING A PROCEDURE MICRO VASCULAR PLUG WAS IMPLANTED. IT WAS REPORTED THAT THE PATIENT WAS DISCHARGED AND SENT HOME. PATIENT WAS TOLD THAT THE DEVICE HAD DETACHED AND WAS FLOATING AROUND INSIDE. THE MVP EMBOLIZED AND THEN MIGRATED INTO THE PULMONARY ARTERY. THE MVP DEVICE WAS EXPLANTED.

    Patient problems: Embolism/Embolus

    MAUDE report 14573233 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page