Axium Prime Detachable Coil System APB-2-2-HX-ESRecalls & reported problems
Micro Therapeutics, Inc. · COIL APB 2-2 HX-ES V06 · Class 2 · Implant
All models of Axium Prime Detachable Coil System (70)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000313869), received between Dec 4, 2023 and May 5, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Physical Resistance/Sticking2
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2026Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT AN AXIUM COIL ENCOUNTERED RESISTANCE IN THE SHEATH AND CATHETER HUB AND PROXIMAL SECTION. IT WAS REPORTED THAT THE COIL DID NOT PASS THROUGH THE MICROCATHETER HUB. THE CATHETER AND COIL WERE NOT DAMAGED. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). NO PATIENT SYMPTOMS OR COMPLICATIONS WERE ASSOCIATED WITH THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED THAT THE CATHETER RESISTANCE OCCURRED AT THE HUB. THERE WAS NO RESISTANCE IN THE SHEATH. THE CAUSE OF THE RESISTANCE WAS NOT DETERMINED. CONTINUOUS SALINE FLUSH WAS ADMINISTERED AND MAINTAINED AS INDICATED IN THE IFU. DURING PREPARATION, THE INTRODUCER SHEATH WAS NOT HELD VERTICALLY WHEN THE IMPLANT COIL WAS PULLED BACK INSIDE DURING HYDRATION BECAUSE THE COIL DIDN'T EVEN GET PAST THE MICROCATHETER. ADDITIONAL INFORMATION WAS RECEIVED. THE CATHETER WAS A MEDTRONIC PRODUCT (ECHELON). BEFORE ATTEMPTING TO INSERT INTO THE MICROCATHETER DURING PREPARATION, THE INTRODUCER SHEATH WAS HELD VERTICALLY WHEN THE IMPLANT COIL WAS PULLED BACK INSIDE DURING HYDRATION. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE WAS NO KINK/DAMAGE TO THE COIL OR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25088466 ↗MalfunctionApr 5, 2024Detachment of Device or Device Component; Physical Resistance/Sticking; Detachment of Device or Device Component; Physical Resistance/Sticking
MEDTRONIC RECEIVED REPORT THAT DURING PREPARATION OF THE COIL PER THE INSTRUCTIONS FOR USE (IFU), THE COIL PREMATURELY SEPARATED FROM THE PUSHWIRE. IT WAS NOTED THERE HAD BEEN FRICTION OR OTHER DIFFICULT PRIOR TO THE SEPARATION. THE PUSHWIRE WAS NOT BENT OR BROKEN AND THE PHYSICIAN DID NOT ROTATE THE PUSHWIRE. NO ATTEMPT HAD BEEN MADE TO DETACH THE COIL. AS THE EVENT OCCURRED DURING PREPARATION, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19052837 ↗MalfunctionDec 4, 2023Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation
ADDITIONAL INFORMATION WAS RECEIVED REPORTING THAT THE COIL DETACHED OUTSIDE AND WAS DISCARDED. THERE WERE NO DETACHMENT ATTEMPTS BECAUSE THE COIL WAS PRE-DETACHED DUE TO THE WIRE BEING FOLDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18260953 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.