Axium Prime Detachable Coil System APB-3-4-HX-ESRecalls & reported problems
Micro Therapeutics, Inc. · COIL APB 3-4 HX-ES V06 · Class 2 · Implant
All models of Axium Prime Detachable Coil System (70)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000313951), received between Jan 6, 2025 and Jan 2, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Physical Resistance/Sticking1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Hemorrhagic Stroke1
- Ruptured Aneurysm1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJan 2, 2026Unintended Movement
ADDITIONAL INFORMATION RECEIVED. THE CAUSE OF THE CATHETER DISPLACEMENT WAS DETERMINED TO BE TENSION IN THE MICROCATHETER. THERE WAS SIGNIFICANT TENSION BECAUSE, UPON REACHING THE ANEURYSM WITH THE COIL, THE SYSTEM FELT SO RIGID THAT THE COIL WOULD NOT DEPLOY WITHIN THE ANEURYSM, GENERATING TENSION. DURING THE INITIAL ASCENT, NO FRICTION WAS FELT; THE TENSION WAS EXPERIENCED WHEN THE COIL REACHED THE DISTAL SEGMENT.
Patient problems: Hemorrhagic Stroke; Ruptured Aneurysm
MAUDE report 23945405 ↗MalfunctionDec 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
MEDTRONIC RECEIVED A REPORT THAT DURING ANEURYSM EMBOLIZATION, WHILE ADJUSTING THE POSITION OF THE COIL INSIDE THE ANEURYSM, THE COIL WAS WITHDRAWN OUT OF THE ANEURYSM. WHEN IT REACHED THE PROXIMAL SEGMENT OF THE MIDDLE CEREBRAL ARTERY, THE COIL DETACHED AUTOMATICALLY. THE COIL WAS THEN RETRIEVED AND REMOVED FROM THE BODY USING A GUIDEWIRE AND SNARE. IT WAS FOUND THAT THE PROXIMAL DETACHMENT POINT WAS NOT KINKED, INDICATING THAT THE DETACHMENT WAS NOT DUE TO OPERATOR ERROR. A NEW COIL WAS USED TO COMPLETE THE PROCEDURE. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE IFU. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THERE WAS NO SURGICAL OR MEDICINAL INTERVENTION REQUIRED. THE PUSHWIRE WAS NOT BENT OR BROKEN. THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY. THERE WAS COIL PREMATURE DETACHMENT. THE IMPLANT COIL WAS REMOVED FROM THE PATIENT. IT WAS REMOVED FROM THE BODY USING A GUIDEWIRE AND SNARE. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED MIDDLE CEREBRAL ARTERY (MCA) ANEURYSM WITH A MAX DIAMETER OF 3.5 MM AND A 2.2 MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23764549 ↗MalfunctionJan 6, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ADDITIONAL INFORMATION RECEIVED THAT A CONTINUOUS SALINE FLUSH WAS ADMINISTERED AND MAINTAINED AS INDICATED IN THE IFU. THE CAUSE OF THE RESISTANCE WAS NOT DETERMINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21079724 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.