This Medtronic Intellis spinal cord stimulator is an implanted device that delivers pain relief for diabetic neuropathy. Of its 1,119 reports, 345 describe injuries and 774 describe malfunctions, with no deaths. Many reports describe trouble recharging the implant, the controller failing to find it, and the device shifting out of place. Some describe the stimulation no longer being felt. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,118
- Deaths reported
- 0
- Injuries reported
- 357
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,118 adverse event reports that list this exact device (DI 00763000315474), received between Jun 4, 2021 and Sep 28, 2026 — classified as 357 injuries, 761 malfunctions.
The most frequently reported specific device problem is “Charging Problem” (228 reports).
9 other recalls cover devices cleared under the same FDA submission (P840001); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Charging Problem228
- Migration or Expulsion of Device200
- High impedance177
- Insufficient Device Problem Information168
- Failure to Deliver Energy164
- Battery Problem160
- Adverse Event Without Identified Device or Use Problem142
- Overheating of Device135
Patient problems
- No Clinical Signs, Symptoms or Conditions546
- Pain341
- Insufficient Information175
- Electric Shock98
- Burning Sensation74
- Discomfort61
- Unspecified Infection37
- Ambulation or Postural Difficulties36
Recent adverse event reports
The 25 most recent of 1,118 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Intermittent Continuity; Unintended Collision; Inappropriate/Inadequate Shock/Stimulation; Low impedance; Charging Problem
It was reported that an implanted Intellis Pain implantable neurostimulator with two Compact 1x8 leads remained in service when the patient experienced strong, unusual shooting stimulation down the left arm on two occasions while using Group B at the usual preferred intensity without increasing stimulation. The patient questioned whether cervical spine positioning contributed to the sensation. The patient reported a fall onto the right side while walking approximately one month earlier and was unsure whether it was related to medication, complex regional pain syndrome, or both. Impedance testing identified possible short findings at contact 1 (590 ohms) and contact 8 (540 ohms); these contacts were not used in any programming group. The patient also reported a trickle charge screen despite the battery not being depleted. The patient programmer reset after the battery was removed and reinserted, and the charging issue resolved without recurrence. A new Group C was programmed, with stimulation felt in the right arm and a smaller amount in the left arm, which the patient reported as the desired distribution. No other stimulation changes were made. The patient intended to request an X-
Patient problems: Device Overstimulation of Tissue; Electric Shock
MAUDE report 50917320 ↗InjurySep 25, 2026Unauthorized Access to Computer System; Insufficient Device Problem Information
Information was received from a patient (pt) regarding an implantable neurostimulator (INS). The reason for call was Pt stated that they had a couple MRIs done of their lower back and they were seeing a new pain doctor and an orthopedist. Pt stated that before they left the last doctor, they started having a device problem. Pt said that the device didn't seem like it was working. Pt said that they were given donation equipment from HCP before they left HCP. Pt proceeded to just speak of their physicians and whatnot. Agent was not understanding Pt's reason for call. Agent kept trying to politely interject, however Pt just kept speaking louder over agent and wouldn't let agent speak. Pt then said that they had received a cybersecurity incident letter. Pt then said something about the controller and belt. Pt said that they thought their device was being interacted with. Agent reviewed. Pt said that someone picked up on their INS since they could feel it. Pt then said that they just had epidurals. Pt said again that the INS didn't work as well as it should. Pt then said that they were getting steroid shots. Pt said again that the INS wasn't doing the job. Agent kept trying to get the P
Patient problems: Insufficient Information
MAUDE report 50904886 ↗InjurySep 24, 2026Migration or Expulsion of Device
Information was received from a patient (pt) regarding an implantable neurostimulator (INS). The reason for call was pt reported increased discomfort in the lower back beginning in January. Pt stated that ablations were completed, but the discomfort persisted. Pt requested assistance from a manufacturing representative (rep) and reported unsuccessful attempts to contact the rep. Agent redirected pt to the healthcare provider (HCP) and provided the NAS phone number to further address the issue. Additional information was received from the patient reporting that the battery was loose at the top causing pain in their back after they lost 30 pounds. Their battery was replaced to place deeper on (b)(4) 2026. The patient indicated that their L (understood as left) side pain has been greatly reduced. The patient in their spine would be reevaluated on (b)(6) 2026.
Patient problems: Discomfort; Pain
MAUDE report 50890346 ↗InjurySep 18, 2026Failure to Deliver Energy; Data Problem
Information was received from a patient regarding an implantable neurostimulator. The reason for call was a few weeks ago the patient decided to turn their stimulation up a notch. Patient stated that when they pushed the up arrow on the controller, the intensity went to 1.6 and they thought that was too high so they turned it back down to1.2. Patient also stated that they lock the controller and when they open it back up again, the stimulation keeps going back to one. Patient mentioned that they charged the controller to 100% and the stimulator to 100%, but they could not feel the stimulation. Agent walked the patient through checking their settings and the patient was on program 1 at 1.8. The troubleshooting steps that were taken on the call resolved the issue.
Patient problems: Insufficient Information
MAUDE report 50822139 ↗InjurySep 18, 2026Energy Output Problem
Information was received from a patient regarding an implantable neurostimulator (INS). The reason for call was patient reported their implant never worked 100% which they always knew and the pain would come and go. Patient mentioned some days it was good and they were good for a while then all of a sudden they were really hurting. Patient mentioned they could hardly walk this morning and they were in so much pain. Patient mentioned their representative would help them adjust their settings since they didn't know where the wires applied to their 'A, B, C, or D' and reprogram their settings in person. Sometimes their stimulation was up too high and they would try to lower it. Agent redirected patient to their HCP to address the issue.
Patient problems: Ambulation or Postural Difficulties; Device Overstimulation of Tissue; Pain
MAUDE report 50820714 ↗InjurySep 18, 2026Failure to Deliver Energy; High impedance; Low impedance; Patient Device Interaction Problem
It was reported that, following an incident unrelated to the implanted Intellis Pain neurostimulator in which the patient almost fell while exiting the shower but caught themself, the patient no longer felt stimulation and experienced return of pain and loss of pain relief. A low impedance or short issue was reported. Connection check showed red X indicators on electrodes 4, 5, 6, 7, 9, 12, and 13 and values greater than 40,000 with a possible open short involving lead electrodes 0 through 7 during a full in-pain check. The neurostimulator was reprogrammed. The issue was reported as resolved. The patient was alive with no injury. The manufacturer representative (rep) indicated they would send any further information as they become aware.
Patient problems: Pain
MAUDE report 50819013 ↗InjurySep 17, 2026Inappropriate/Inadequate Shock/Stimulation; Battery Problem
Information was received from a patient's representative (caller) regarding an implantable neurostimulator (INS). The caller reported that it's been probably at least 6 weeks since the patient was having electrical shocks and the stimulation doesn't work down to their legs like it supposed to. Pt mentioned they could only feel the stimulation on their back. Caller also mentioned the implant battery wasn't staying long enough and only lasts 5 hours. Pt mentioned they already tried lowering the stimulation down but still the same. Agent reviewed information and redirected them to their healthcare provider (HCP) and manufacturer representative (rep) to further address.
Patient problems: Electric Shock
MAUDE report 50809844 ↗InjurySep 16, 2026Battery Problem
709306287- doesnt shut off properly, 709250118-returned the original dev, 709390778- Can¿t find sweet spot with intensity]\": Information was received from a consumer regarding a patient with an implantable neurostimulator (INS) 2026-08-31 PATLTR (con) \"[Relate event: 709306287- doesnt shut off properly, 709250118-returned the original dev, 709390778- Can¿t find sweet spot with intensity]\": Information was received from a consumer regarding a patient with an implantable neurostimulator (INS). The new device was implanted as the battery of the original device was failing and no longer holding its charge.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50796818 ↗InjurySep 9, 2026Failure to Deliver Energy; Failure to Interrogate; Battery Problem; Communication or Transmission Problem
Information was received from a healthcare provider regarding an implantable neurostimulator. The reason for call was the patient tried to use the controller about 2 weeks ago and it was unsuccessful with connecting to the implant. The remote says unable to connect and the caller indicated that they think the patient had not been using therapy consistently. The issue was not resolved through troubleshooting. Technical Services (TSS) reviewed information about charging an INS. The caller will follow-up with the patient and direct them to contact PS if they need assistance to charge. No symptoms were reported. Additional information was received from the patient (pt). Pt called back and repeated details of call, and that they need to get an MRI. They seemed very confused and unfamiliar with equipment usage. PT says they didn't think the therapy was helping them. During the call pt was able to start Passive Recharge Mode (PRM), and after about 10 minutes they report they are now charging INS with excellent quality. It shows the controller is at 80 percent and INS is low. PT will continue to charge INS, and will call patient and technical services (PARS) back for MRI mode instruction,
Patient problems: Insufficient Information
MAUDE report 50699676 ↗InjurySep 1, 2026Therapy Delivered to Incorrect Body Area
It was reported that a patient noted that when stimulation was increased to obtain lower-extremity relief, intense stimulation or buzzing in the ribs was reported with all programs. Reprogramming was performed using both implanted leads, and rib stimulation continued. Some lower-extremity relief was reported when stimulation was maintained at a low level sufficient to avoid rib stimulation. An impedance check was performed, and all contacts were within normal limits. Thoracic-spine imaging was ordered, and follow-up was planned after imaging was completed. The stimulation-related rib sensation remained ongoing. The manufacturer representative (rep) indicated they would send any further information as they become aware.
Patient problems: Inappropriate Device Stimulation of Tissue
MAUDE report 50546676 ↗InjurySep 1, 2026Failure to Deliver Energy
Information was received from patient (pt) regarding who was implanted with an implantable neurostimulator (INS). It was reported that when they opened the controller this morning to check how much battery power they had, the controller started flashing A1, A2, and A3, and it turned off their stimulation. The patient stated they had not seen this before and that it was their first time experiencing it. The patient mentioned that they used Group A all the time. The patient stated that they were not feeling any stimulation. The patient confirmed that Group A was highlighted in green. Agent had the patient press the white button on top of the controller and select \"Stimulation On\". Pt confirmed that the controller battery was at 100% and the implant was at 80%. Agent had pt reset the controller. Pt confirmed that the controller screen was no longer flashing. Pt stated that they had Groups A, B, and C available. Pt confirmed that the stimulation was on in all programs under Group A. Pt stated that there was no stimulation on Group A and mentioned that it showed intensity 0. Agent guided the pt in adjusting the intensity and switching to different groups. Pt confirmed that the intensi
Patient problems: Insufficient Information
MAUDE report 50542640 ↗InjuryAug 24, 2026Insufficient Device Problem Information
It was reported that the implanted Intellis implantable neurostimulator had stopped working approximately 8 weeks earlier and was no longer helping with pain or providing sufficient pain coverage. The patient reported no falls or accidents that would have affected the device. Lead connections and impedances were checked and were within normal limits. The battery was replaced, and the device was explanted. No evidence that anything was wrong with the device was identified. The issue was resolved, and the patient was alive with no injury. The manufacturer representative (rep) indicated they would send any further information as they become aware.
Patient problems: Pain
MAUDE report 50407618 ↗InjuryAug 20, 2026Inappropriate/Inadequate Shock/Stimulation; Battery Problem; Insufficient Device Problem Information
Information was received from a patient (pt) regarding an implantable neurostimulator (INS). The reason for the call was patient stated as of recently, the INS was not helping as they thought it would to treat their lower back. Patient stated they were not happy with the device. Patient stated they would always feel a zapping constantly from their lower back to their toes, and that's how they knew the device was not working (INS depleted) due to the zapping stopping. Patient stated they used to charge every 3 days, but they met with a representative who changed all their programs, and now they charge every 7-10 days. Patient stated both their flanks still hurt. The patient was redirected to their healthcare provider to further address the issue.
Patient problems: Electric Shock; Pain
MAUDE report 50350975 ↗MalfunctionAug 17, 2026Charging Problem
Information was received from a patient (pt) regarding an external device. The reason for call was patient reported that they noticed 2-3 weeks ago they've been getting "battery needs to be replaced soon" message every time they charge the implant; pt added that they're also experiencing longer charging time and needing to charge the implant more often. Agent attempted to troubleshoot the device and get serial numbers; pt mentioned there's no visible damage with the controller and battery pack but was unable to read the serial number. Agent reviewed the need to troubleshoot the devices. Pt mentioned they will callback once their wife is at home. When asked about healthcare provider (HCP); pt can't remember the name but mentioned Spine and Joint Clinic. Patient and patient rep called back and repeated previously documented information. Patient requested a replacement battery pack for the controller after receiving a "Replace the controller batteries soon" message for at least the past three weeks. Patient was unsure whether the message appeared with or without the recharger. Patient stated the implant used to charge quickly, but they now had to charge it every day and for longer per
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50269815 ↗InjuryAug 5, 2026Adverse Event Without Identified Device or Use Problem
Information was received from a patient regarding an implantable neurostimulator. The reason for call was patient reported they had to have their device removed because they ended up getting bacteria in their back after the surgery. Patient said the back pain they had was right where the stimulator was. Patient also said when they went in to remove the bacteria, there was a lot on the battery and the lead wires. Patient did not know when the issue started.
Patient problems: Bacterial Infection
MAUDE report 50145257 ↗InjuryAug 5, 2026Adverse Event Without Identified Device or Use Problem
It was reported that an implanted Intellis Pain implantable neurostimulator and an implanted MRI lead were removed after the patient reported a return of pain and stated that the system never really worked well for pain relief and that pain felt unchanged compared to the pre-implant baseline. The spinal cord stimulation system was explanted at the patient¿s request due to lack of perceived pain relief. The patient was reported to be alive with no injury, and the issue was reported as resolved. No patient complications have been reported as a result of this event. The manufacturer representative (rep) indicated they would send any further information as they become aware.
Patient problems: Pain
MAUDE report 50141746 ↗MalfunctionJul 30, 2026Display or Visual Feedback Problem; Failure to Deliver Energy; Unintended Collision; Charging Problem; Insufficient Device Problem Information
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50090743 ↗InjuryJul 30, 2026High impedance; Insufficient Device Problem Information
It was reported that a Vectris MRICS Compact Lead 977A260 had high impedance identified on electrodes 8¿15 during the day of surgery, with impedance values reported as greater than 10,000. The patient had a history of many falls as reported by the patient. Impedances were tested with the patient standing, sitting, and lying. A physician planned to replace the lead due to the high impedance, and the product was explanted. The patient was reported alive with no injury. No patient complications have been reported as a result of this event. The manufacturer representative (rep) indicated they would send any further information as they become aware.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50086087 ↗InjuryJul 30, 2026Overheating of Device; Insufficient Device Problem Information
Information was received from the legal representative of a patient who was implanted with multiple implantable neurostimulators (INS). The attorney alleged that the patient wanted to hold the manufacturer accountable for ¿knowingly abusing the federal regulatory process to market and distribute an adulterated spinal cord stimulator system.¿ The attorney alleged that the manufacturer ¿secured FDA clearance for hundreds of significant hardware and firmware modifications through serial PMA supplements rather than the statutorily required new PMA applications, despite knowing these modifications materially altered the device's design and performance.¿ The attorney alleged the patient sustained personal injuries because of their ¿defective spinal cord stimulator device.¿ The attorney noted that patient¿s INS was approved under PMA P840001 and was ¿supplemented through a series of streamlined FDA processes that did not require new clinical testing, including 30-day notices and real-time review pathways.¿ The attorney alleged that MAUDE reports associated with the patient¿s INS model number included ¿more than 2,700 unique adverse events between 2017 and 2022, with recurring patterns of
Patient problems: Burning Sensation; Distress; Inadequate Pain Relief; Insufficient Information; Pain; Unspecified Tissue Injury
MAUDE report 50085082 ↗InjuryJul 29, 2026Failure to Deliver Energy; High impedance
It was reported that an implanted Implantable Neurostimulator had high impedance on electrodes 9, 11, and 15, with impedance values noted as above 40,000, and there was a loss of stimulation. The high impedance was not observed on the day of surgery. Troubleshooting was performed by reprogramming around the affected electrodes. The issue was reported as resolved, and the patient was alive with no injury. No cause was known. No patient complications have been reported as a result of this event. The manufacturer representative (rep) indicated they would send any further information as they become aware.
Patient problems: Insufficient Information
MAUDE report 50068405 ↗InjuryJul 17, 2026Migration or Expulsion of Device
INFORMATION WAS RECEIVED FROM A MANUFACTURER'S REPRESENTATIVE (REP) REGARDING AN IMPLANTABLE NEUROSTIMULATOR. THE REASON FOR CALL WAS REPRESENTATIVE SAID THEY WILL BE COVERING A LEAD REVISION CASE LATER THIS AFTERNOON. REPRESENTATIVE SAID IT LOOKS LIKE ONE OF THE LEADS MIGRATED SOUTH. REPRESENTATIVE ASKED HOW THEY CAN TELL WHICH LEAD IS CONNECTED TO A SPECIFIC PORT ON THE BATTERY. AGENT REVIEWED THAT SHORT OF RUNNING AN IMPEDANCE CHECK WHEN THE LEAD IS DISCONNECTED, THEY WOULD NOT BE ABLE TO TELL WHICH SIDE IS CONNECTED. THE ISSUE WAS NOT RESOLVED THROUGH TROUBLESHOOTING. THE PATIENT WAS REDIRECTED TO THEIR HEALTHCARE PROVIDER TO FURTHER ADDRESS THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25860241 ↗MalfunctionJul 16, 2026Energy Output Problem; Low impedance; Therapeutic or Diagnostic Output Failure
INFORMATION WAS RECEIVED FROM A PATIENT (PT) REGARDING AN IMPLANTABLE NEUROSTIMULATOR (INS). THE REASON FOR CALL WAS PATIENT SAID THEY HAD AN OPERATION ON THEIR BACK ON JUNE 8TH AND DURING THE OPERATION THEY ASKED THE NEUROSURGEON TO MOVE THE STIMULATOR TO A PLACE CLOSER TO THE SIDE SO IT'S EASIER FOR THEM TO CHARGE THE IMPLANT. PATIENT SAID SINCE THEN THE CONTROLLER SAYS SOMETHING LIKE IT CANNOT CONTROL THE INTENSITY OR FIND IT OR SOMETHING IS OFF WITH IT SO IT NEEDS TO BE RESET. PATIENT STATED WHEN THEY TURN THERAPY ON THERE IS ONE OR TWO SETTINGS THAT ENDS UP BEING KIND OF PAINFUL, NOT VERY PAINFUL BUT SLIGHTLY LIKE A SHARP PAIN. PATIENT STATED THEY HAD A LOBECTOMY AND THE MUSCLES ARE HEALING BECAUSE THEY CUT THROUGH THEM. PATIENT STATED THEY HAVE NOT CHARGE THE IMPLANT SINCE THE SURGERY AND THEY ARE WARY OF DOING ANYTHING AT THIS POINT. AGENT ASKED PATIENT TO CONNECT WITH THE CONTROLLER AND THE CONTROLLER IS SHOWING SETTINGS NOT AVAILABLE, CANNOT PROVIDE YOUR DESIRED INTENSITY SETTING. AGENT CONFIRMED PATIENT IS GETTING THE SAME MESSAGE ON ALL THE PROGRAMS AND GROUPS. AGENT CONFIRMED THE CONTROLLER BATTERY IS AT 100% AND THE IMPLANT WAS AT 50%. AGENT ASKED CLARIFYING QUESTION A
Patient problems: Pain
MAUDE report 25838059 ↗MalfunctionJul 14, 2026High impedance
IT WAS REPORTED THAT A PATIENT HAD HIGH IMPEDANCE READINGS. HIGH IMPEDANCE OF 40,000 OHMS WAS REPORTED ON MULTIPLE ELECTRODES (#0, #1, #2, #3, #4, AND #8), WITH GUIDANCE NOTED TO AVOID OR NOT USE THE AFFECTED ELECTRODES, INCLUDING ELECTRODE #8 SHOWING 40,000 OHMS WHILE USING ELECTRODES 1, 3, OR 4. TROUBLESHOOTING WAS PERFORMED BY REPROGRAMMING THE DEVICE USING THE RIGHT LEAD, AND STIMULATION WAS REPORTED TO CAPTURE THE PATIENT¿S LOW BACK PAIN. THE ISSUE WAS REPORTED AS ONGOING AND NOT RESOLVED, AND THE CAUSE WAS REPORTED AS UNKNOWN. THE MANUFACTURER REPRESENTATIVE (REP) INDICATED THEY WOULD SEND ANY FURTHER INFORMATION AS THEY BECOME AWARE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25814610 ↗InjuryJul 13, 2026Adverse Event Without Identified Device or Use Problem
709243654-INS DEEP, HARD TO CHARGE]: INFORMATION WAS RECEIVED FROM A PATIENT REGARDING AN IMPLANTABLE NEUROSTIMULATOR. THE REASON FOR CALL WAS THE PATIENT NEEDED TO GET AN MRI AND NEEDED TO KNOW THEIR MRI ELIGIBILITY. THE PATIENT SAID THEY WERE GETTING ANOTHER MRI DONE BECAUSE THEY HAD THEIR IMPLANTED NEUROSTIMULATORS REVISED AND THE HEALTHCARE PROVIDER PUT THE DEVICE IN DEEPER ON (B)(6) OF LAST YEAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25799108 ↗MalfunctionJul 10, 2026High impedance
IT WAS REPORTED THAT AN IMPLANTED INTELLIS PAIN IMPLANTABLE NEUROSTIMULATOR AND IMPLANTED LEAD 977A260 5MM COMPACT 1X8 MRI LEADS HAD IMPEDANCE IRREGULARITIES IDENTIFIED DURING CONNECTION CHECK AND IMPEDANCE CHECK TESTING. A CONNECTION CHECK SHOWED A RED X ON CONTACTS 0 AND 3, AND IMPEDANCE CHECK VALUES WERE REPORTED AS 0 AND >40000, WITH CONTACTS 1, 2, 3, AND 9 REPORTED AS OPEN/SHORT WITH >40000. THE ISSUE WAS REPORTED AS ONGOING FOR TWO MONTHS. THE PATIENT DENIED FALLS, TRIPS, OR ACCIDENTS. THE DEVICE WAS REPROGRAMMED AND THE PHYSICIAN WAS UPDATED. THE ISSUE WAS REPORTED AS RESOLVED, AND THE PATIENT WAS ALIVE WITH NO INJURY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE MANUFACTURER REPRESENTATIVE (REP) INDICATED THEY WOULD SEND ANY FURTHER INFORMATION AS THEY BECOME AWARE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25779126 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P840001 on Nov 30, 1984. Submitted by Medtronic Neuromodulation as “Itrel(R) Totally Implantable Spinal Cord Stim. Sys”. The FDA review took 322 days. The FDA has since approved 615 supplements (changes to the device or its labeling), most recently on Nov 13, 2026. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Recalls under the same FDA submission
These recalls involve devices cleared under P840001. One submission often covers many models, so check the product and lot details in each record.
- Z-0372-2025 · Sep 18, 2024 · Open, ClassifiedMedtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
- Z-1691-2024 · Mar 28, 2024 · Open, ClassifiedIntellis AdaptiveStim Product Number 97715, Spinal Cord Stimulator
- Z-0780-2024 · Dec 19, 2023 · Open, ClassifiedModel A710 Intellis Clinician Programmer Application, version 2.0.97
- Z-2159-2021 · Jul 9, 2021 · Open, ClassifiedMedtronic Vectris SureScan MRI, Catalog Numbers: a) 977A260 b) 977A275 c) 977A290
- Z-1926-2016 · May 3, 2016 · TerminatedMedtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of the trunk and/or li
- Z-1147-2015 · Jan 9, 2015 · TerminatedThe Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
- Z-1524-2014 · Feb 27, 2014 · TerminatedMedtronic neurostimulators. Restore (37711), RestoreADVANCED (37713), RestoreADVANCED SureScan MRI (97713), RestoreSENSOR (37714), RestoreSENSOR SureScan MRI (97714), PrimeAdvanced
- Z-0278-2014 · Sep 30, 2013 · TerminatedMedtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.
- Z-1536-06 · Aug 2, 2006 · TerminatedMedtronic 3777 Lead Kit, 1x8 Low Impedance Lead Kit for Spinal Cord Stimulation (SCS). Manufactured at: Medtronic, Inc., Villalba, Puerto Rico USA. Medtronic, Inc., Minneapolis
Common questions about the Intellis™ 97716
Has the Intellis™ 97716 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00763000315474), based on the FDA recall file dated Nov 27, 2026. However, 9 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the Intellis™ 97716?
The FDA's MAUDE database lists 1,118 adverse event reports for this device, received between Jun 4, 2021 and Sep 28, 2026. In those reports, the most common specific device problems are “Charging Problem” (228), “Migration or Expulsion of Device” (200) and “High impedance” (177); the most common patient problems are “Pain” (341), “Electric Shock” (98) and “Burning Sensation” (74). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Intellis™ 97716?
None of the reports are classified as deaths, 357 as injuries and 761 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Intellis™ 97716 FDA approved?
It was approved through premarket approval (PMA) P840001 on Nov 30, 1984.
Is the Intellis™ 97716 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Intellis™ 97716, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand Intellis™, model 97716). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy devices
- WaveWriter Alpha™ Prime SC-1432Boston Scientific Neuromodulation Corporation · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy7 deaths970 reports
- WaveWriter Alpha™ Prime 16 SC-1416Boston Scientific Neuromodulation Corporation · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy15 deaths927 reports
- Intellis™ 97716Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy633 reports
- Intellis™ Pro 977118Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy188 reports
- Vectris™ SureScan® 977A260Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy69 reports
- N/A A710Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy1 recall0 reports
- Infinion™ Pro SC-2318-50Boston Scientific Neuromodulation Corporation · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy43 reports
- NA 97725Medtronic, Inc. · Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy16 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page