Skip to content

Vectris™ SureScan® 977A260 (UDI 00763000324278)Recalls & reported problems

Medtronic, Inc. · LEAD 977A260 VECTRIS MRICS COMPACT · Class 3 · Implant

All models of Vectris SureScan (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This is a Medtronic Vectris SureScan spinal cord stimulation lead, a thin wire implanted near the spine to deliver pain-relieving signals to the lower back or legs. The 70 reports mostly describe leads breaking, cracking, or deforming, and trouble positioning or threading the lead during or after the implant procedure, along with high impedance readings. 14 reports involve injuries; no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
69
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 69 adverse event reports that list this exact device (DI 00763000324278), received between Jun 2, 2022 and May 1, 2026 — classified as 14 injuries, 55 malfunctions.

The most frequently reported specific device problem is “Break” (22 reports).

9 other recalls cover devices cleared under the same FDA submission (P840001); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 4 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 3 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 3 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 1 reportsNov 2025: 3 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break22
  • Positioning Problem15
  • Material Deformation11
  • Adverse Event Without Identified Device or Use Problem10
  • High impedance9
  • Insufficient Device Problem Information9
  • Difficult to Insert8
  • Difficult to Remove4

Patient problems

  • No Clinical Signs, Symptoms or Conditions60
  • Pain4
  • Insufficient Information3
  • Cerebrospinal Fluid Leakage2
  • Perforation2
  • Headache1
  • Scar Tissue1

Recent adverse event reports

The 25 most recent of 69 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 1, 2026Break; Difficult to Insert

    THE MANUFACTURER REPRESENTATIVE (REP) REPORTED THAT SURGEON WAS NOT ABLE TO REINSERT THE STYLET ONCE IT WAS REMOVED. REP SAID SURGEON HAD TO USE ANOTHER LEAD. REP TO SEND LEAD BACK FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25056109 ↗
  • MalfunctionApr 9, 2026Unsealed Device Packaging; Device Damaged Prior to Receipt by User

    IT WAS REPORTED THAT, PRIOR TO USE, THE IMPLANTABLE LEAD WAS FOUND WITH CRACKED PACKAGING AND THE DEVICE WAS NEVER IMPLANTED. DUE TO QUESTIONABLE STERILITY OF THE PRODUCT INSIDE, THEY DID NOT USE THE PRODUCT. THE LEAD COMPONENTS INSIDE THE PACKAGING WERE DISCARDED BY THE CUSTOMER, AND THE CRACKED PACKAGING WAS RETAINED FOR RETURN. THERE WAS NO PATIENT INVOLVED. THE MANUFACTURER REPRESENTATIVE (REP) INDICATED THEY WOULD SEND ANY FURTHER INFORMATION AS THEY BECOME AWARE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24831952 ↗
  • MalfunctionMar 9, 2026Difficult to Insert; Insufficient Device Problem Information

    IT WAS REPORTED THAT AFTER INITIAL PLACEMENT OF THE LEAD, THE PHYSICIAN ATTEMPTED TO REINSERT THE STYLET TO ASCEND THE LEAD TIP HIGHER AND DISCOVERED A BLOCKAGE AT THE DISTAL END OF THE LEAD, WHICH WAS POTENTIALLY A BLOOD CLOT. MULTIPLE ATTEMPTS WERE MADE TO THREAD THE STYLET, BUT THESE WERE UNSUCCESSFUL. THE LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE ISSUE WAS RESOLVED. THE MANUFACTURER REPRESENTATIVE (REP) INDICATED THEY WOULD SEND ANY FURTHER INFORMATION AS THEY BECOME AWARE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24546270 ↗
  • MalfunctionJan 26, 2026Impedance Problem; Insufficient Device Problem Information

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT IMPLANTED WITH A NEUROSTIMULATOR (INS). IT WAS REPORTED THAT THE DOCTOR WAS PLACING THORACIC PERC LEADS AND THE ENTIRE LEAD WAS LIT UP RED WHEN CHECKING THE CONNECTION IN THE WIRELESS EXTERNAL NEUROSTIMULATOR (WENS). THIS WAS A DEVICE ISSUE WHEN CONNECTING A NEW LEAD TO A NEW WENS, ALL CONNECTIONS WERE RED. THE PATIENT DIDN¿T EXPERIENCE AN ISSUE. THE ISSUE WAS A LEAD ISSUE OUT OF THE BOX. AFTER TESTING THE LEAD CONNECTION IN THE WENS, THE LEAD WAS MOVED TO THE OTHER SIDE OF THE WENS (0-7 TO 8-15) AND THE LEAD CONNECTIONS WERE ALL STILL RED. REP OPENED A NEW LEAD AND THEIR WERE NO ISSUES AT ALL. ALL CONNECTIONS WERE GOOD ON THE SAME WENS. THE CAUSE WAS NOT KNOWN. THE ISSUE WAS RESOLVED. THE DEVICE WAS IN REP'S POSSESSION. WILL BE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24183682 ↗
  • MalfunctionDec 30, 2025High impedance; High impedance

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL (HCP) REGARDING A PATIENT'S IMPLANT PROCEDURE. IT WAS REPORTED THAT THERE WERE HIGH IMPEDANCES. MANUFACTURER REPRESENTATIVE (REP) CHECKED CONNECTIONS WITH LEAD IN BOTH CHANNELS (0-7 AND 8-15) OF EXTERNAL NEUROSTIMULATOR AND THE SAME ELECTRODE SHOWED THE HIGH IMPEDANCE VALUE ON ELECTRODE 4 OR 12 DEPENDING ON THE CHANNEL. THE LEAD WAS THEN PLACED INTO THE NEUROSTIMULATOR AND CONTINUED TO SHOW HIGH IMPEDANCE VALUE ON ONE ELECTRODE SO THE PHYSICIAN REQUESTED A NEW LEAD. CAUSE IS NOT KNOWN, BUT THE ISSUE WAS RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23923544 ↗
  • MalfunctionDec 9, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER ATTEMPTING TO PLACE THE LEAD 977A260 5MM COMPACT 1X8 MAGNETIC RESONANCE IMAGING (MRI) AND REMOVING IT FROM THE STYLET, THE LEAD WAS FOUND TO BE FRACTURED, DAMAGED, AND BROKEN. THE DEVICE WAS NEVER IMPLANTED. THE ISSUE WAS RESOLVED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE MANUFACTURER REPRESENTATIVE (REP) INDICATED THEY WOULD SEND ANY FURTHER INFORMATION AS THEY BECOME AWARE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23756437 ↗
  • InjuryNov 24, 2025Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE (REP) REGARDING A PATIENT (PT) WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). REP REPORTS THERE WAS AN ANCHOR ISSUE. REP REPORTS A LEAD MIGRATION OR DISLODGEMENT OCCURRED. PRODUCTS DID NOT MIGRATE AROUND OR ENTANGLE INTERNAL ORGANS. NO STIMULATION ISSUES WERE REPORTED AS A RESULT OF THIS ISSUE. LEADS WERE EXPLANTED. REP REPORTS THE ISSUE WAS RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23636265 ↗
  • InjuryNov 24, 2025Migration or Expulsion of Device; Migration or Expulsion of Device

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE (REP) REGARDING A PATIENT (PT) WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). REP REPORTS THERE WAS AN ANCHOR ISSUE. REP REPORTS A LEAD MIGRATION OR DISLODGEMENT OCCURRED. PRODUCTS DID NOT MIGRATE AROUND OR ENTANGLE INTERNAL ORGANS. NO STIMULATION ISSUES WERE REPORTED AS A RESULT OF THIS ISSUE. LEADS WERE EXPLANTED. REP REPORTS THE ISSUE WAS RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23636248 ↗
  • MalfunctionNov 12, 2025High impedance; Adverse Event Without Identified Device or Use Problem; High impedance; Adverse Event Without Identified Device or Use Problem

    INS IS NEW AND IMPLANTED NOTHING WAS WRONG WITH THAT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23533996 ↗
  • InjuryOct 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE (REP) REPORTING THE LEAD SERIAL NUMBERS.

    Patient problems: Perforation

    MAUDE report 23300431 ↗
  • InjurySep 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE (REP) REPORTING THE LEAD SERIAL NUMBERS.

    Patient problems: Perforation

    MAUDE report 23109729 ↗
  • MalfunctionSep 12, 2025Material Deformation; Material Twisted/Bent; Positioning Problem; Material Deformation; Material Twisted/Bent; Positioning Problem

    IT WAS REPORTED THAT THE LEAD WAS NOT IMPLANTED. MANUFACTURER REPRESENTATIVE (REP) STATED THAT THE LEAD WAS BENT WHILE ATTEMPTING TO GAIN ACCESS TO EPIDURAL SPACE TO IMPLANT AND UNABLE TO STEER EVEN WITH A DIFFERENT STYLET. THEY HAD TO OPEN A NEW LEAD TO USE WHICH WORKED FINE. ISSUE RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23039545 ↗
  • MalfunctionSep 2, 2025Break; Break

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22945305 ↗
  • MalfunctionJun 27, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    INFORMATION WAS RECEIVED FROM MANUFACTURER REPRESENTATIVE (REP) WHO REPORTED WHEN THE LEAD KIT WAS OPENED BY THE OPERATION ROOM STAFF, THE SCRUB TECHNICIAN NOTICED A SMUDGE ON THE INSIDE OF THE STERILE PLASTIC COVER. AFTER EXAMINATION IT APPEARED TO BE THE GLUE THAT IS USED ON THE PAPER. CAUSE IS UNKNOWN. NO SYMPTOMS WERE REPORTED. IT WAS REPORTED THAT PRODUCT WILL BE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22350422 ↗
  • MalfunctionJun 12, 2025Break; Positioning Problem; Break; Positioning Problem

    IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING THE LEAD 977A260 5MM COMPACT 1X8 MAGNETIC RESONANCE IMAGING (MRI), THE LEAD WAS KINKED AND COULD NOT BE ADVANCED, WITH SIGNIFICANT RESISTANCE ENCOUNTERED DURING ATTEMPTED ADVANCEMENT. THE HEALTHCARE PROFESSIONAL CHANGED THE STYLET IN AN ATTEMPT TO RESOLVE THE ISSUE BUT WAS STILL UNABLE TO ADVANCE THE LEAD. THE ISSUE WAS RESOLVED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE MANUFACTURER REPRESENTATIVE (REP) INDICATED THEY WOULD SEND ANY FURTHER INFORMATION AS THEY BECOME AWARE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22194290 ↗
  • MalfunctionMay 7, 2025Break; Break

    AFTER OPENING LEAD FOR PROCEDURE HOSPITAL STAFF NOTICED THE LEAD COULD BE FRAYED. THEREFORE LEAD WAS NOT IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21977000 ↗
  • MalfunctionApr 18, 2025Device Damaged Prior to Receipt by User; Device Damaged Prior to Receipt by User

    2025-MAR-19: INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) WHO REPORTED WHEN THE LEAD KIT WAS OPENED THEY NOTICED IT WAS COMPROMISED. THEY REPORTED THE THERE WAS AN OUT OF BOX FAILURE AS THE LEAD TRAY WAS BROKEN. NO SYMPTOMS WERE REPORTED. (B)(6) 2025: MANUFACTURER REPRESENTATIVE (REP) STATED THE LEAD KIT TRAY WAS OPENED IN THE PACKAGE. REP TO SEND BACK PRODUCT WITH THE TRAY. TECHNICAL SERVICES TOLD REP THEY CAN SEND IT BACK WITHOUT USING THE RETURN KIT MAILER SINCE THE TRAY WON'T FIT IN THE MAILER. OOB ISSUE. 2025-MAR-21: ADDITIONAL INFORMATION WAS RECEIVED FROM MANUFACTURER REPRESENTATIVE (REP) WHO REPORTED THE PHYSICAL TRAY WHICH HELD THE LEAD COMPONENTS WAS VISUALLY COMPROMISED. THEY PROVIDED TRACKING DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21865897 ↗
  • InjuryApr 15, 2025Break; Difficult to Insert; Break; Difficult to Insert

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THAT DURING SURGERY INVOLVING THE LEAD, THE STYLET WAS CHANGED MULTIPLE TIMES, AND ON THE FOURTH CHANGE, THE STYLET WOULD NOT GO IN ALL THE WAY. IT WAS NOTED THAT DAMAGE WAS CAUSED OR DISCOVERED DURING SURGERY. THE PRODUCT WAS EXPLANTED. NO TROUBLESHOOTING WAS PERFORMED. THE ISSUE WAS RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21839223 ↗
  • MalfunctionApr 9, 2025Unsealed Device Packaging; Device Contaminated During Manufacture or Shipping; Unsealed Device Packaging; Device Contaminated During Manufacture or Shipping

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED LEAD WAS OPENED TO SCRUB TECH AND THE INTERNAL STERILE PACKAGING HAD MOISTURE IN IT. CAUSE WAS UNKNOWN; WILL BE RETURNING THE LEAD IN ITS PACKAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21808695 ↗
  • MalfunctionFeb 12, 2025Unsealed Device Packaging; Device Damaged Prior to Receipt by User; Unsealed Device Packaging; Device Damaged Prior to Receipt by User

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THAT WHEN THE REP WHEN TO USE THE PRODUCT FOR AN IMPLANT THEY NOTICED THE INTERNAL PACKAGE WAS COMPLETELY DESTROYED AND NOT STERILE. THE OUTSIDE PACKAGE WAS UNDAMAGED. THE DEVICE WAS NOT USED WITH A PATIENT SO THERE WAS NO PATIENT INVOLVEMENT. THE CAUSE IS UNKNOWN, BUT THE DEVICE WILL BE RETURNED. THE REP NOTED THEY WOULD SUBMIT ANY NEW INFORMATION THAT BECOMES AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21363517 ↗
  • MalfunctionJan 28, 2025Break; Material Deformation; Positioning Problem; Break; Material Deformation; Positioning Problem

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THAT DURING AN ATTEMPT TO IMPLANT A LEAD, THE LEAD WAS FRACTURED, DAMAGED, OR BROKEN. THE PATIENT HAD A SIGNIFICANT AMOUNT OF SCAR TISSUE, WHICH CONTRIBUTED TO DIFFICULTIES IN STEERING THE LEAD, RESULTING IN A BEND. TROUBLESHOOTING STEPS INCLUDED CHANGING STYLETS AND APPLYING LIGHT FORCE TO IMPLANT THE LEAD, BUT THE LEAD COULD NO LONGER BE STEERED. THE ISSUE REMAINS ONGOING. THE MANUFACTURER REPRESENTATIVE (REP) INDICATED THEY WOULD SEND ANY FURTHER INFORMATION AS THEY BECOME AWARE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Scar Tissue

    MAUDE report 21252033 ↗
  • MalfunctionJan 27, 2025Break; Material Deformation; Break; Material Deformation

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED LEAD WAS OPENED ON THE STERILE FIELD, PHYSICIAN EXAMINED LEAD AND DETERMINED IT WAS BENT MORE THAN NORMAL IN THE OPPOSITE DIRECTION OF ORIGINAL BEND. NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21246262 ↗
  • MalfunctionJan 17, 2025Positioning Problem; Insufficient Device Problem Information; Positioning Problem; Insufficient Device Problem Information

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). PHYSICIAN CLAIMED THE LEAD FELT DIFFERENT TO PREVIOUS LEADS AND HE HAD MORE TROUBLE DRIVING IT THAT USUAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21178580 ↗
  • MalfunctionDec 9, 2024Positioning Problem; Positioning Problem

    INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THAT DURING A PROCEDURE, THE LEAD WAS NOT DRIVING WELL AND WAS DIFFICULT TO STEER. ADDITIONALLY, THE LOCKING MECHANISM OF THE LEAD ONTO THE BEVEL WOULD NOT STAY SECURE. UPON REMOVING THE PRE-THREADED BLUE TIP STYLET AND REPLACING IT WITH A GREEN TIP STYLET, THE STYLET WAS FOUND TO BE BENT. THE LEAD AND STYLET WERE NOT USED DURING THE CASE AND WERE NEVER IMPLANTED. TROUBLESHOOTING WAS PERFORMED, WHICH INCLUDED AN ATTEMPTED USE OF THE LEAD. THE ISSUE WAS RESOLVED AS THE LEAD WAS NOT IMPLANTED. THE MANUFACTURER REPRESENTATIVE (REP) INDICATED THEY WOULD SEND ANY FURTHER INFORMATION AS THEY BECOME AWARE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20880806 ↗
  • InjuryNov 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE (REP). REP REPORTED THAT THE LEADS WERE CONNECTED, BUT WHILE MAPPING WAS BEING PERFORMED, THE PATIENT FELT THE STIMULATION, THEY DID NOT LIKE IT AND WANTED IT OUT. THE CAUSE OF THE ISSUE WAS NOT DETERMINED. REP REPORTED NOTHING WAS IMPLANTED AND THE PREVIOUS DEVICE WAS REMOVED. ALL DEVICES HAVE BEEN DISCARDED. THE INFORMATION HAS BEEN CONFIRMED WITH THE PHYSICIAN.

    Patient problems: Insufficient Information; Pain

    MAUDE report 20737558 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P840001 on Nov 30, 1984. Submitted by Medtronic Neuromodulation as “Itrel(R) Totally Implantable Spinal Cord Stim. Sys”. The FDA review took 322 days. The FDA has since approved 615 supplements (changes to the device or its labeling), most recently on Nov 13, 2026. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Recalls under the same FDA submission

These recalls involve devices cleared under P840001. One submission often covers many models, so check the product and lot details in each record.

  • Z-0372-2025 · Sep 18, 2024 · Open, Classified
    Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
  • Z-1691-2024 · Mar 28, 2024 · Open, Classified
    Intellis AdaptiveStim Product Number 97715, Spinal Cord Stimulator
  • Z-0780-2024 · Dec 19, 2023 · Open, Classified
    Model A710 Intellis Clinician Programmer Application, version 2.0.97
  • Z-2159-2021 · Jul 9, 2021 · Open, Classified
    Medtronic Vectris SureScan MRI, Catalog Numbers: a) 977A260 b) 977A275 c) 977A290
  • Z-1926-2016 · May 3, 2016 · Terminated
    Medtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of the trunk and/or li
  • Z-1147-2015 · Jan 9, 2015 · Terminated
    The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
  • Z-1524-2014 · Feb 27, 2014 · Terminated
    Medtronic neurostimulators. Restore (37711), RestoreADVANCED (37713), RestoreADVANCED SureScan MRI (97713), RestoreSENSOR (37714), RestoreSENSOR SureScan MRI (97714), PrimeAdvanced
  • Z-0278-2014 · Sep 30, 2013 · Terminated
    Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.
  • Z-1536-06 · Aug 2, 2006 · Terminated
    Medtronic 3777 Lead Kit, 1x8 Low Impedance Lead Kit for Spinal Cord Stimulation (SCS). Manufactured at: Medtronic, Inc., Villalba, Puerto Rico USA. Medtronic, Inc., Minneapolis

Common questions about the Vectris™ SureScan® 977A260

Has the Vectris™ SureScan® 977A260 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00763000324278), based on the FDA recall file dated Nov 27, 2026. However, 9 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Vectris™ SureScan® 977A260?

The FDA's MAUDE database lists 69 adverse event reports for this device, received between Jun 2, 2022 and May 1, 2026. In those reports, the most common specific device problems are “Break” (22), “Positioning Problem” (15) and “Material Deformation” (11); the most common patient problems are “Pain” (4), “Cerebrospinal Fluid Leakage” (2) and “Perforation” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Vectris™ SureScan® 977A260?

None of the reports are classified as deaths, 14 as injuries and 55 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Vectris™ SureScan® 977A260 FDA approved?

It was approved through premarket approval (PMA) P840001 on Nov 30, 1984.

Is the Vectris™ SureScan® 977A260 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Vectris™ SureScan® 977A260, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand Vectris™ SureScan®, model 977A260). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy devices

All Stimulator, spinal-cord, totally implanted for relief of pain due to diabetic neuropathy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page