CONCERTO VERSA™ CV-4-15 (UDI 00763000321260)Recalls & reported problems
Micro Therapeutics, Inc. · Coil CV-4-15 Concerto Versa V02 · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000321260), received between Mar 16, 2021 and Sep 1, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Difficult or Delayed Positioning1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 1, 2021Difficult or Delayed Positioning; Detachment of Device or Device Component; Material Deformation; Difficult or Delayed Positioning; Detachment of Device or Device Component; Material Deformation
MEDTRONIC RECEIVED A REPORT THAT THERE WAS RESISTANCE AND PREMATURE DETACHMENT WITH A CONCERTO COIL IT WAS REPORTED THAT THIS INCIDENT OCCURRED DURING A BENCHTOP STUDY IN A GLASS AND SILICONE MODEL WITH PHYSICIAN -NO PATIENT INVOLVEMENT. PHYSICIAN WAS DEPLOYING COIL INTO 4MM GLASS VESSEL MODEL AT THE END OF A TORTUOUS PARTIAL SPLENIC MODEL THROUGH 4F GLIDECATHETER. MODEL HAD APPROXIMATELY 300 ML/MIN FLOW OF BLOOD MIMICKING FLUID. PHYSICIAN WAS HAVING TROUBLE GETTING COIL TO ANCHOR IN THIS SITUATION SO BEGAN ROTATING THE CATHETER TO ATTEMPT TO TWIST COIL AND THEREBY GET IT TO "GRAB" THE VESSEL WALL. A KINK IN THE CATHETER WAS NOTED AT THE CELIAC TAKEOFF POINT FROM THE AORTA, POSSIBLY RELATED TO THE TWISTING. THE PHYSICIAN WAS MANIPULATING THE COIL AND NOTED THAT IT STOPPED RESPONDING (WITHDRAWING) WITH THE PUSHER AND SAID THAT IT HAD PREMATURELY DETACHED. PREMATURE DETACHMENT/BREAK OCCURRED WITH COIL PARTIALLY OUT OF CATHETER AND PARTIALLY IN. PHYSICIAN WAS ABLE TO REMOVE THE CATHETER AND COIL TOGETHER. THE COIL WAS PARTIALLY INSIDE THE CATHETER AND PARTIALLY OUT; IT WAS PHOTOGRAPHED AT THE TIME AND THEN REMOVED FROM THE CATHETER BY MEDTRONIC PERSONNEL AT THE SCENE FOR SAFE SHIPPING
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12406696 ↗MalfunctionMar 16, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
ADDITIONAL INFORMATION RECEIVED INDICATED THE DEVICE WAS PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU), WITH THE EXCEPTION OF USING A CONTINUOUS FLUSH. THE PATIENT WAS BEING TREATED FOR A 3.5 MM GDA (GASTRODUODENAL ARTERY) WITH NORMAL VESSEL TORTUOSITY. THE REMAINDER OF THE PROCEDURE WAS COMPLETED WITHOUT ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11494818 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page