CONCERTO VERSA™ CV-12-50 (UDI 00763000321680)Recalls & reported problems
Micro Therapeutics, Inc. · Coil CV-12-50 Concerto Versa V02 · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00763000321680), received between Aug 31, 2021 and Sep 1, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Physical Resistance/Sticking2
- Material Deformation1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 1, 2021Detachment of Device or Device Component; Physical Resistance/Sticking; Detachment of Device or Device Component; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THERE WAS RESISTANCE AND PREMATURE DETACHMENT WITH A CONCERTO COIL IT WAS REPORTED THAT MALFUNCTION HAPPENED IN A SILICONE MODEL - NO PATIENT INVOLVED. PHYSICIAN DEPLOYED THIRD COIL INTO SILICONE ANEURYSM MODEL (TWO PREVIOUS COILS). PER PROTOCOL, PHYSICIAN WAS ASKED TO RETRIEVE COIL AND RE-DEPLOY IT AS IF THE COIL WERE THE WRONG SIZE. PHYSICIAN HAD BEGUN WITHDRAWING COIL BACK INTO CATHETER AND NOTED THAT THERE WAS SOME RESISTANCE DURING THE WITHDRAWAL. THERE WAS NO APPARENT REASON FOR RESISTANCE, SO PHYSICIAN CONTINUED WITHDRAWING, AND THE COIL PREMATURELY DETACHED. SUBSEQUENT COIL WAS DELIVERED THROUGH CATHETER WITH NO PROBLEMS OR RESISTANCE NOTED. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES PREPARED AS INDICATED IN THE IFU. THERE WAS NO SURGICAL OR MEDICAL INTERVENTION REQUIRED. THE PUSHWIRE WAS NOT BENT OR BROKEN. THERE WAS FRICTION/DIFFICULTY DURING DELIVERY. THE COIL WAS REPOSITIONED IN THE PROCESS OF WITHDRAWING IT THE FIRST TIME. THE PHYSICIAN DID NOT ATTEMPT TO DETACH THE COIL. THE PHYSICIAN DID NOT ROTATE THE DELIVERY PUSHER DURING THE PROCEDURE, AND A CONTINUOUS FLUSH WAS ADMINISTERED DURING THE PROCEDURE. ANCILLARY DEVICES INCLUDE A 6F
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12406806 ↗MalfunctionAug 31, 2021Separation Failure; Detachment of Device or Device Component; Material Deformation; Physical Resistance/Sticking; Separation Failure; Detachment of Device or Device Component; Material Deformation; Physical Resistance/Sticking
MEDTRONIC RECEIVED INFORMATION REGARDING A CONCERTO VERSA COIL THAT FAILED TO DETACH DURING A BENCHTOP TESTING USING A SILICONE MODEL. IT WAS REPORTED THE DEVICES WERE NOT PREPARED PER THE INSTRUCTIONS FOR USE (IFU); THERE WAS NO INSPECTION AND NO FLUSHING USED. THE PHYSICIAN DEPLOYED THE COIL PER THEIR TYPICAL PRACTICE. THE MANUFACTURER REPRESENTATIVE (REP) ASKED THE PHYSICIAN TO INSERT, FULLY WITHDRAW, AND THEN INSERT THIS COIL AGAINST. THE PHYSICIAN NOTED SOME FRICTION DESCRIBED AS THE COIL FELT "SCRATCHY" DURING INSERTION THROUGH THE WAVELOCK SHEATH. NO OTHER ISSUES WERE NOTED DURING DEPLOYMENT. HOWEVER, WHEN DETACHMENT WAS ATTEMPTED WITH THE INSTANT DETACHER (ID), THE COIL DID NOT DETACH AND COULD BE SEEN RETRACTING WITH THE PUSHER THROUGH THE CATHETER. THE COIL WAS REPOSITIONED IN THE ANEURYSM AND THE REP ATTEMPTED TO DETACH IT AGAINST WITH THE ID BUT THE COIL STILL DID NOT DETACH AND WAS PULLED BACK WITH THE DELIVERY SYSTEM DURING WITHDRAWAL. IT WAS NOTED THERE WAS NO ISSUE WITH THE INSTANT DETACHER. MANUAL DETACHMENT WAS THEN ATTEMPTED. THE PUSHER BROKE AT THE HBI AND THE PROXIMAL SECTION OF THE PUSHER WAS PULLED BACK A SUBSTANTIAL AMOUNT (MORE THAN 2CM) TO TRY TO RETRACT T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12393789 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page