Skip to content

Becker® 25120Recalls & reported problems

Medtronic Ps Medical, Inc. · EDMS II 25120 BECKER W/1-WAY VALVE · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00763000333430), received between May 19, 2022 and Sep 8, 2025 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Break” (9 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 1, 2024Open, Classified

    Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).

    Z-0661-2025 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Break9
  • Fracture9
  • Fluid/Blood Leak6
  • Disconnection3
  • Complete Blockage1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 8, 2025Disconnection; Disconnection

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORY SYSTEM. IT WAS REPORTED THAT THE DRAIN TUBING BROKE OFF THE Y PORT ON THE PRESSURE TUBING CONNECTED TO THE DRAIN. THE 3-WAY STOP COCK PROXIMAL TO PATIENT WAS IMMEDIATELY CLOSED TO THE PATIENT, AND THE DRAIN CLAMPED. A NEW DRAIN SET UP WAS OBTAINED AND SET UP, CONNECTED TO PATIENT, DRAIN RELEVELED AND ZEROED WITH GOOD WAVE FORM DISPLAYED ON THE MONITOR, ICP [INTRACRANIAL PRESSURE] 14.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22992485 ↗
  • MalfunctionSep 4, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT WHILE USING THE SYSTEM, THE DOCTOR NOTICED A LEAK OF CEREBROSPINAL FLUID FROM THE PATIENT LINE AT THE VALVE LEVEL. THE SYSTEM WAS REPLACED. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20135401 ↗
  • MalfunctionSep 4, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT WHILE USING THE SYSTEM, THE DOCTOR NOTICED A LEAK OF CEREBROSPINAL FLUID FROM THE PATIENT LINE AT THE VALVE LEVEL. THE SYSTEM WAS REPLACED. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20135179 ↗
  • MalfunctionJun 26, 2024Break; Fluid/Blood Leak; Fracture; Insufficient Device Problem Information; Break; Fluid/Blood Leak; Fracture; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THERE WAS A FAILURE OF THE SYSTEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19613820 ↗
  • MalfunctionMar 27, 2024Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE COLLECTION COLLECTION CHAMBER AND THE TUBING BELOW THAT WAS CLAMPED WAS NOTED TO BE LEAKING. ON FURTHER INSPECTION IT APPEARED THAT THE TUBING HAD SOME CRACKS IN IT CAUSING THE LEAK. THE MANUFACTURER REPRESENTATIVE WAS INFORMED THAT THE TUBING IN THAT LOCATION OF THE DEVICE FELT RIGID. THE ONE THAT WAS USED TO REPLACE THE ORIGINAL ONE HAD SOFTER PLIABLE TUBING AND APPEARED TO BE FUNCTIONING APPROPRIATELY. THERE WAS A PROCEDURE DELAY OF 15 MINUTES. THE PATIENT'S STATUS WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18988365 ↗
  • MalfunctionDec 1, 2023Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT ON (B)(6) 2023 REGISTERING PRESSURE READING. THE SYSTEM WAS LEAKING AND HAD COME APART BETWEEN THE COLLECTION CHAMBER AND DRAINAGE BAG. THE SYSTEM WAS CHANGED TO ANOTHER SYSTEM. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18246265 ↗
  • MalfunctionOct 31, 2023Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THE SYSTEM WAS IN USE ON (B)(6) 2023, AND THEN ON (B)(6) 2023, THE SYSTEM WAS BROKEN WHERE THE BREAKAGE OF THE DRAINAGE WAS AT THE SET AT THE HEIGHT OF THE CLAMP DOWNSTREAM OF THE COLLECTION WELL. THE DEVICE WAS CHANGED FOR ANOTHER ONE. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18038183 ↗
  • MalfunctionOct 25, 2023Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THERE WAS BREAKAGE OF THE DRAINAGE SET AT THE HEIGHT OF THE TAP PROXIMAL TO THE CATHETER CONNECTION. THE SYSTEM WAS REPLACED. THE PATIENT'S STATUS WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18004909 ↗
  • MalfunctionOct 25, 2023Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THERE WAS BREAKAGE OF THE DRAINAGE SET AT THE HEIGHT OF THE CLAMP DOWNSTREAM OF THE COLLECTION WELL. THE SYSTEM WAS REPLACED. THERE WAS A PROCEDURE DELAY OF 60 MINUTES. THE PATIENT'S STATUS WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18004418 ↗
  • MalfunctionOct 23, 2023Break; Fracture; Break; Fracture

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE WAS A PROCEDURE DELAY OF 60 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17986278 ↗
  • MalfunctionMay 10, 2023Complete Blockage; Complete Blockage

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING DEVICE. IT WAS REPORTED THAT AT 18:06 IN THE EVENING, THE DOCTOR REPORTED THAT THE DEVICE WAS NOT DRAINING CSF. THE REPRESENTATIVE ASKED IF THE SYSTEM WAS PRIMED BEFORE CONNECTING TO THE PATIENT LINE AND THE DOCTOR REPLIED YES. THEY QUESTIONED IF IT WAS PRIMED SEQUENTIALLY FOLLOWING THE MANUALS RECOMMENDATIONS AND REVIEWED THE STEPS. THE DOCTOR CONFIRMED THAT THE SYSTEM WAS PRIMED CORRECTLY. THE DOCTOR STOPPED USING THIS DEVICE AND IT WAS REPLACED WITH A DIFFERENT SYSTEM, AND DID SO AT THE PATIENT'S BEDSIDE IN THE PICU.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16902427 ↗
  • MalfunctionNov 21, 2022Break; Disconnection; Fracture; Break; Disconnection; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM (EDMS) USED DURING A PROCEDURE. IT WAS REPORTED THAT THERE WAS A RUPTURE WITH DETACHMENT OF THE DEVICE UNDER THE LIQUID COLLECTION PUMP. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THERE WAS NO REPORTED DELAY TO THE PROCEDURE. THERE WAS NO REPORTED IMPACT TO THE PATIENT. THE ISSUE WAS CONFIRMED TO HAVE OCCURRED ABOVE THE DRAINAGE BAG. ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THIS IS THE SECOND OCCURRENCE OF THIS ISSUE. THE CAUSE OF THE ISSUE WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15834183 ↗
  • MalfunctionNov 16, 2022Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL MONITORING DEVICE. IT WAS REPORTED THAT THERE WAS A RUPTURE WITH DETACHMENT OF THE DEVICE UNDER THE LIQUID COLLECTION SUMP. THERE WAS NO REPORTED IMPACT TO PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15806099 ↗
  • MalfunctionSep 24, 2022Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING A BECKER DEVICE. IT WAS REPORTED THAT THERE CEREBROSPINAL FLUID (CSF) LEAKAGE FROM THE CSF COLLECTION WELL. THE DEVICE WAS CHANGED FOR ANOTHER DEVICE AND THE ISSUE WAS RESOLVED WITH NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15476756 ↗
  • MalfunctionMay 19, 2022Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING A VALVE SYSTEM. IT WAS REPORTED THAT THERE WAS A BREAKAGE OF THE DRAINAGE SYSTEM OF THE CSF DERIVATION UNDER THE COLLECTION CHAMBER, UNDER THE CLAMP WITH CONSEQUENT CONSEQUENT CREATION OF AN OPEN CIRCUIT. IT WAS NECESSARY TO REPLACE THE OPEN CIRCUIT WITH A NEW CLOSED SYSTEM. THE PROCEDURE WAS COMPLETED WITH A BACK UP PRODUCT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14434279 ↗

Other Device, monitoring, intracranial pressure devices

All Device, monitoring, intracranial pressure devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.