Becker® 27931Recalls & reported problems

Medtronic Ps Medical, Inc. · EDMS 27931 BECKER GS TUBING SMARTSITES · Class 2

All models of Becker (28)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00763000333669), received between Aug 9, 2023 and Jan 19, 2024 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 7 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 1, 2024Open, Classified

    Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).

    Z-0661-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak8
  • Break6
  • Fracture6
  • Disconnection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 19, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THERE WAS LEAKING NEAR THE MAIN STOPCOCK. THE DRAIN WAS REPLACED WITH A DIFFERENT DRAIN. NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18544447 ↗
  • MalfunctionJan 19, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THERE WAS LEAKING NEAR THE MAIN STOPCOCK. THE DRAIN WAS REPLACED WITH A DIFFERENT DRAIN. NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18544436 ↗
  • MalfunctionJan 19, 2024Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THERE WAS LEAKING NEAR THE MAIN STOPCOCK. THE DRAIN WAS REPLACED WITH A DIFFERENT DRAIN. NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18544424 ↗
  • MalfunctionJan 4, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE DEVICE HAD A CRACK IN THE TUBING NEAR THE PATIENT LINE STOPCOCK. ANOTHER DEVICE OF THE SAME MODEL WAS OPENED TO REPLACE IT, HOWEVER, IT WAS ALSO CRACKED. THE DEVICE WAS REPLACED WITH A DIFFERENT MODEL TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18444331 ↗
  • MalfunctionJan 4, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE DEVICE HAD A CRACK IN THE TUBING NEAR THE PATIENT LINE STOPCOCK. ANOTHER DEVICE OF THE SAME MODEL WAS OPENED TO REPLACE IT, HOWEVER, IT WAS ALSO CRACKED. THE DEVICE WAS REPLACED WITH A DIFFERENT MODEL TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18444309 ↗
  • MalfunctionJan 4, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE DEVICE HAD A CRACK IN THE TUBING NEAR THE PATIENT LINE STOPCOCK. ANOTHER DEVICE OF THE SAME MODEL WAS OPENED TO REPLACE IT, HOWEVER, IT WAS ALSO CRACKED. THE DEVICE WAS REPLACED WITH A DIFFERENT MODEL TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18444282 ↗
  • MalfunctionJan 4, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE DEVICE HAD A CRACK IN THE TUBING NEAR THE PATIENT LINE STOPCOCK. ANOTHER DEVICE OF THE SAME MODEL WAS OPENED TO REPLACE IT, HOWEVER, IT WAS ALSO CRACKED. THE DEVICE WAS REPLACED WITH A DIFFERENT MODEL TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18444249 ↗
  • MalfunctionDec 11, 2023Break; Fluid/Blood Leak; Fracture; Break; Fluid/Blood Leak; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM (EDMS) USED DURING A PROCEDURE. IT WAS REPORTED THAT THE PRODUCT WAS CRACKED AT THE SAMPLE PORT. WHEN THE STOPCOCK WAS OPENED TO TAKE CEREBROSPINAL FLUID (CSF) FROM THE CHAMBER, CSF AND BLOOD LEAKED OUT OF THE STOPCOCK AND APPEARED TO BE CRACKED. NO DELAY OR PATIENT IMPACT INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18301362 ↗
  • MalfunctionNov 16, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE HOSPITAL USED TWO OF THE DEVICES AND EXPERIENCED A LEAK AT THE STOPCOCK CLOSEST TO THE PATIENT WITH BOTH OF THEM WHILE CONNECTED TO THE PATIENT. THE SITE DISCONNECTED THE TUBING FROM THE PATIENT AND TESTED EACH DEVICE SEPARATELY WITH SALINE AND IDENTIFIED A LEAK AT THE STOPCOCK CLOSEST TO THE PATIENT. A THIRD ONE WAS OPENED AND DID NOT LEAK. THE PRODUCTS WERE USED IN THE EMERGENCY ROOM. THE PATIENT'S STATUS WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18147922 ↗
  • MalfunctionNov 16, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT THE HOSPITAL USED TWO OF THE DEVICES AND EXPERIENCED A LEAK AT THE STOPCOCK CLOSEST TO THE PATIENT WITH BOTH OF THEM WHILE CONNECTED TO THE PATIENT. THE SITE DISCONNECTED THE TUBING FROM THE PATIENT AND TESTED EACH DEVICE SEPARATELY WITH SALINE AND IDENTIFIED A LEAK AT THE STOPCOCK CLOSEST TO THE PATIENT. A THIRD ONE WAS OPENED AND DID NOT LEAK. THE PRODUCTS WERE USED IN THE EMERGENCY ROOM. THE PATIENT'S STATUS WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18147914 ↗
  • MalfunctionNov 13, 2023Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT WHILE ONE OF THE NEUROSURGERY NURSE PRACTITIONERS WAS ADMINISTERING INTRATHECAL MILRINONE, THEY NOTICED AIR BUBBLES IN THE TUBING AND WHEN INVESTIGATING FURTHER, IT WAS NOTED THAT THERE WAS A CRACK IN THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18119800 ↗
  • MalfunctionOct 10, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT FLUID LEAKED AT THE CONNECTION TO THE 3-WAY STOPCOCK WHICH WAS ONE THAT CANNOT BE MANIPULATED. IT WAS ATTEMPTED TO BE TIGHTENED WITHOUT RESOLUTION. THE REPORTED EVENT RESULTED IN A PROCEDURE DELAY OF 30 MINUTES. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17904840 ↗
  • MalfunctionOct 10, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING SYSTEM. IT WAS REPORTED THAT FLUID LEAKED AT THE CONNECTION TO THE 3-WAY STOPCOCK WHICH WAS ONE THAT CANNOT BE MANIPULATED. IT WAS ATTEMPTED TO BE TIGHTENED WITHOUT RESOLUTION. THE REPORTED EVENT RESULTED IN A PROCEDURE DELAY OF 30 MINUTES. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17904801 ↗
  • MalfunctionAug 9, 2023Disconnection; Disconnection

    MEDTRONIC RECEIVED INFORMATION REGARDING AN EXTERNAL DRAINAGE AND MONITORING DEVICE. IT WAS REPORTED THAT THE TUBING NEAR THE MAIN STOP COCK DISENGAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17492143 ↗

Other Device, monitoring, intracranial pressure devices

All Device, monitoring, intracranial pressure devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page