Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00763000335793), received between Jul 8, 2024 and Jun 24, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 24, 2025Fracture; Fracture

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF DISSECTING TOOL BREAKAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22317573 ↗
  • MalfunctionJun 16, 2025Fracture; Fracture

    IT WAS REPORTED THAT THERE WAS TOOL BROKE. NO PATIENT IMPACT REPORTED. IT WAS ALSO REPORTED THAT THERE WAS DELAY OF 10MINS, NO BROKEN PIECES OF THE REPORTED PRODUCT REMAIN INSIDE THE PATIENT'S BODY, PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT(S). ON FOLLOWUP IT WAS CONFIRMED WITH THE PHYSICIAN THE DRILL FIRED OFF AT HIGH SPEED AND ALMOST HIT ANESTHESIA.

    Patient problems: Insufficient Information

    MAUDE report 22209892 ↗
  • MalfunctionJul 8, 2024

    DURING CASE, A STERILE, SINGLE USE 23 CM 3.0 MEDTRONIC TWIST DRILL WAS OPENED. OR TEAM IMMEDIATELY NOTED THAT DRILL BIT WAS DISCOLORED AND APPEARED TO BE CONTAMINATED. ITEM WAS IMMEDIATELY GIVEN TO OR CASE MANAGER AND COMPANY VENDOR NOTIFIED. NO HARM TO PATIENT. DRILL BIT NEVER REACHED THE PATIENT OR THE STERILE FIELD. DRILL BIT REF #MR8-23TD3030, LOT #0223111521. DEVICE AVAILABLE IN CASE MANAGER'S OFFICE. COMPANY VENDOR WILL REPORT TO MEDTRONIC CASE MANAGER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19689083 ↗

FDA clearance and classification

Cleared through 510(k) K202552 on Nov 2, 2020 (decision: Substantially Equivalent). Submitted by Medtronic Powered Surgical Solutions (Mpss) as “Midas Rex Attachments and Dissecting Tools”. The FDA review took 60 days. FDA record ↗

The FDA regulates product code OLO (orthopedic stereotaxic instrument) under 21 CFR 882.4560 as a Class II device. 21 CFR 882.4560 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Midas Rex MR8-23TD3030

Has the Midas Rex MR8-23TD3030 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00763000335793), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Midas Rex MR8-23TD3030?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jul 8, 2024 and Jun 24, 2025. In those reports, the most common specific device problem is “Fracture” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Midas Rex MR8-23TD3030?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Midas Rex MR8-23TD3030 FDA approved?

It was cleared through 510(k) K202552 on Nov 2, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Midas Rex MR8-23TD3030, and is it still sold?

The FDA's device database lists Medtronic Powered Surgical Solutions as the company responsible for it (brand Midas Rex, model MR8-23TD3030). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.