Skip to content

DxTerity™ DXT5JL35 (UDI 00763000394165)Recalls & reported problems

Medtronic, Inc. · CATH DXT5JL35 DXT 5F 5PK 100CM JL35 · Class 2

All models of DxTerity (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00763000394165), received between Jun 12, 2024 and Jun 12, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K161287); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Difficult to Advance1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 12, 2024Detachment of Device or Device Component; Difficult to Advance; Material Deformation; Detachment of Device or Device Component; Difficult to Advance; Material Deformation

    AN ATTEMPT WAS MADE TO USE ONE DXTERITY DIAGNOSTIC CATHETER DURING A PROCEDURE. THE DEVICE WAS REMOVED FROM THE PACKAGING AND PREPPED PER IFU WITH NO ISSUES. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED. IT WAS REPORTED THAT THERE WAS A TEAR/HOLE IN THE TIP OF THE CATHETER. ONCE THE TEAR/HOLE WAS NOTICED THE DEVICE WAS REMOVED AND REPLACE WITH ANOTHER CATHETER. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19517961 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K161287. One submission often covers many models, so check the product and lot details in each record.

  • Z-3253-2018 · Aug 16, 2018 · Terminated
    Medtronic DxTerity(TM) TRA 5F, Diagnostic Catheter, REF U5TRAN35
  • Z-1479-2018 · Mar 13, 2018 · Terminated
    Medtronic DxTerity Diagnostic Catheter, JL 4.0, REF DXT5JL40, marked with 6F color code on box.

Other Catheter, intravascular, diagnostic devices

All Catheter, intravascular, diagnostic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page