SYNCHROMED®II 8637-20 (UDI 00763000422608)Recalls & reported problems

Medtronic, Inc. · PUMP 8637-20 SM II 20ML EMAN SYMBL GRC · Class 3 · Implant

All models of SYNCHROMEDII (4)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This is the Medtronic SynchroMed II infusion pump, implanted to deliver medication. The 1,628 reports often describe communication failures, flow problems, and the pump moving out of place. Some mention pain and infection. The records include 2 deaths, 989 injuries and 637 malfunctions. There is 1 open recall, from November 2023, over the pump possibly not restarting after an MRI scan.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,613
Deaths reported
2
Injuries reported
984
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1,613 adverse event reports that list this exact device (DI 00763000422608), received between Jul 6, 2021 and Sep 29, 2026 — classified as 2 deaths, 984 injuries, 627 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (225 reports).

1 FDA recall record name this device's identifier, 1 still open.

39 other recalls cover devices cleared under the same FDA submission (P860004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 19 reportsJan 2024: 26 reportsFeb 2024: 22 reportsMar 2024: 18 reportsApr 2024: 20 reportsMay 2024: 21 reportsJun 2024: 20 reportsJul 2024: 24 reportsAug 2024: 17 reportsSep 2024: 31 reportsOct 2024: 33 reportsNov 2024: 18 reportsDec 2024: 11 reportsJan 2025: 21 reportsFeb 2025: 29 reportsMar 2025: 28 reportsApr 2025: 14 reportsMay 2025: 23 reportsJun 2025: 23 reportsJul 2025: 28 reportsAug 2025: 26 reportsSep 2025: 19 reportsOct 2025: 12 reportsNov 2025: 18 reportsDec 2025: 17 reportsJan 2026: 33 reportsFeb 2026: 54 reportsMar 2026: 45 reportsApr 2026: 15 reportsMay 2026: 20 reportsJun 2026: 11 reportsJul 2026: 4 reportsAug 2026: 2 reportsSep 2026: 4 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202654
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 13, 2023Class IIOpen, Classified

    Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.

    Z-0614-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem465
  • Communication or Transmission Problem225
  • Insufficient Device Problem Information218
  • Infusion or Flow Problem217
  • Migration or Expulsion of Device207
  • Insufficient Flow or Under Infusion138
  • Obstruction of Flow116
  • Electromagnetic Interference110

Patient problems

  • No Clinical Signs, Symptoms or Conditions790
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available244
  • Pain235
  • Insufficient Information162
  • Unspecified Infection151
  • Cerebrospinal Fluid Leakage48
  • Muscular Rigidity47
  • Seroma39

Recent adverse event reports

The 25 most recent of 1,613 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Obstruction of Flow; Insufficient Device Problem Information

    Information was received from a consumer regarding a patient with an implantable pump containing Morphine (unknown) for non-malignant pain. It was reported that they had a pump and catheter implant (b)(6) 2019, and the catheter got tied up with scare tissue and catheter got kinked, scar tissue grew around the pump and the pump stopped and the pump had to be replaced on (b)(6) 2022. Patient reported the healthcare provider re-established the catheter in a different position.

    Patient problems: Scar Tissue

    MAUDE report 50941289 ↗
  • MalfunctionSep 29, 2026Charging Problem; Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939799 ↗
  • InjurySep 18, 2026Migration or Expulsion of Device; Insufficient Device Problem Information

    Information was received from a friend/family member regarding a patient (pt) who was receiving Unknown Drug via an implantable pump for non-malignant pain indications for use. It was reported that the patient's pump had not been implanted for a while. Caller said that when the pump was removed, the catheter was left implanted. Caller said that the pump was removed on (b)(6)2022. Caller said that the pt got a cyst where the pump was removed from, and the cyst was removed around (b)(6)2026 since the cyst had gotten so hard that it was causing the pt pain. Caller said that the catheter was not tied strong enough, so the pt had spinal fluid leaking from the catheter. Caller said that the healthcare provider (HCP) wanted to cut the catheter shorter and put fatty tissue over the catheter to stop the spinal fluid from leaking. Caller said that since the prior week, the spinal fluid was leaking quite a bit. Caller said that the pt had been having migraines since the cyst was removed. Caller said that they thought the catheter had moved since the cyst was removed. Caller said that the pt was admitted to the hospital last night. It was reported that a neurosurgeon was supposed to meet with

    Patient problems: Cerebrospinal Fluid Leakage; Cyst(s); Headache; Pain

    MAUDE report 50824708 ↗
  • MalfunctionSep 14, 2026Pumping Stopped

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50765425 ↗
  • MalfunctionAug 14, 2026Ambient Temperature Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50260203 ↗
  • InjuryAug 3, 2026Battery Problem

    Information was received from a patient's representative regarding an implantable infusion pump. The reason for call was caller stated that the patient had the pump replaced 2 weeks ago because the battery was going to die.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50115279 ↗
  • MalfunctionJul 29, 2026Electromagnetic Interference

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50073206 ↗
  • InjuryJul 29, 2026Insufficient Device Problem Information

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50067363 ↗
  • MalfunctionJul 27, 2026Application Program Problem; Communication or Transmission Problem

    Information was received from a patient who was receiving Unknown Drug via an implantable pump for non-malignant pain. It was reported that the communicator was not syncing up with the personal therapy manager (PTM) and the only way to reset it was to restart the PTM. The communicator would flash and not connect saying communicator not found. They were also getting the 156 error at least twice a day now. They noted the issue started probably close to a year ago when connecting to the PTM app and it got to a point where it was time consuming because they would have to restart and reload the PTM back up; it took 16 minutes to connect to PTM app and then another 17 minutes to process the bolus. Their PTM was replaced years ago and once they got this one they always lagged at 90%. During call agent walked them through clearing data, resetting communicator, and had them close all applications. They were able to connect to PTM app like normal. They will call back if issues persist.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50044046 ↗
  • MalfunctionJul 2, 2026Electromagnetic Interference

    INFORMATION WAS RECEIVED FROM A MANUFACTURER'S REPRESENTATIVE (REP) REGARDING A PATIENT RECEIVING UNKNOWN BACLOFEN AND DILAUDID VIA AN IMPLANTABLE INFUSION PUMP FOR NON-MALIGNANT PAIN. IT WAS REPORTED THAT THE PATIENT HAD AN MRI (MAGNETIC RESONANCE IMAGING) EARLIER THAT DAY AND THEY WERE AT THE FACILITY TO READ THE PUMP AFTER THE MRI. THE REP STATED THAT THEY GOT A SERVICE CODE 99 FROM (B)(6) 2026 THAT SAID, 'POSSIBLE PUMP DAMAGE, THE PUMP IS STOPPED FOR 1092 HOURS'. THE REP ALSO MENTIONED THE PUMP STARTED TO ALARM TODAY BECAUSE IS NEEDS A REFILL AND A LOW RESERVOIR ALARM. THE REP WAS ADVISED TO CHECK LOGS FOR A MOTOR STALL RECOVERY THAT WOULD SHOW IN LOGS TODAY. THE ISSUE WAS NOT RESOLVED THROUGH TROUBLESHOOTING. ADDITIONAL INFORMATION WAS RECEIVED FROM A MANUFACTURER'S REPRESENTATIVE (REP) AND CONFIRMED BY THE HEALTHCARE PROVIDER (HCP) REGARDING THE PATIENT RECEIVING BACLOFEN (20MCG/ML AT 3.375MCG/DAY) AND DILAUDID (2MG/ML AT .3375MG/DAY) VIA THEIR IMPLANTABLE INFUSION PUMP. IT WAS REPORTED THAT THE CAUSE OF THE ISSUE WAS UNKNOWN. THE PATIENT STATED THEIR PUMP WAS NOT INTERROGATED AFTER THEIR LAST MAGNETIC RESONANCE IMAGING (MRI) FOR UNKNOWN REASONS. THE HCP COMPLETED A REFILL (B

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25691366 ↗
  • MalfunctionJun 24, 2026Excess Flow or Over-Infusion; Insufficient Flow or Under Infusion

    INFORMATION WAS RECEIVED FROM A PATIENT WHO WAS RECEIVING DILAUDID (HYDROMORPHONE) AND BUPIVACAINE VIA AN IMPLANTABLE PUMP FOR NON-M ALIGNANT PAIN. IT WAS REPORTED THAT THE PATIENT (PT) HAD LEFT OVER MEDICATION WHEN IT WAS TIME TO REFILL THE PUMP. PT REPORTED THAT IN (B)(6) THERE WERE SUPPOSED TO HAVE 11.7 MG LEFT IN THE PUMP BUT INSTEAD HAD 11. PT REPORTED IN (B)(6) THEY WERE SUPPOSED TO HAVE 4.5 MG BUT THEY HAD 5. IN (B)(6) SUPPOSED TO BE 4.5 MG BUT WAS 7 AND ON (B)(6) IT WAS SUPPOSED TO BE 3.6 BUT WAS AT 6 MG. THE ISSUE WAS NOT RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25612556 ↗
  • InjuryJun 23, 2026Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, HEALTHCARE PROVIDER) REGARDING A PATIENT WHO WAS RECEIV ING COMPOUNDED BACLOFEN (250 MCG/ML AT 111.77 MCG/DAY) AND DILAUDID (HYDROMORPHONE) (5 MG/ML AT 2.2355 MG/DAY) VIA AN IMPLANTABLE PUMP FOR SPINAL PAIN INDICATIONS. IT WAS REPORTED THAT PATIENT HAS BEEN REPORTING SUBOPTIMAL PAIN RELIEF. IMAGING REVEALED THAT HER CATHETER TIP WAS IMPLANTED AT ANTERIOR T7, SO THE PHYSICIAN PLANNED TO RETRACT THE CATHETER BACK WITHOUT REMOVING THE ANCHOR. PATIENT WAS TAKEN TO THE OPERATING ROOM FOR PLANNED CATHETER REVISION. THE PHYSICIAN BEGAN BY IDENTIFYING THE PRE-EXISTING ANCHOR AND CATHETER TIP UNDER FLUOROSCOPIC GUIDANCE. THE CATHETER TIP WAS AT ANTERIOR T7. THE PHYSICIAN PROCEEDED TO OPEN UP THE MIDLINE, LUMBAR INCISION AND DISSECTED DOWN TO THE EXISTING ANCHOR AND SPINAL SEGMENT. ONCE THAT WAS LOCATED, HE THEN PROCEEDED TO RETRACT THE CATHETER BACK WITHOUT REMOVING THE EXISTING ANCHOR. THE CATHETER TIP WAS PLACED AT T10 AND WAS CONFIRMED TO BE IN A POSTERIOR POSITION. INCISIONS WERE THEN IRRIGATED AND CLOSED. THE ISSUE IS RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25600312 ↗
  • MalfunctionJun 22, 2026Application Program Problem; Communication or Transmission Problem

    INFORMATION WAS RECEIVED FROM A PATIENT WHO WAS RECEIVING UNKNOWN DRUG VIA AN IMPLANTABLE PUMP FOR SPINAL PAIN. IT WAS REPORTED THAT THE HANDSET TOOK A LONG TIME TO CONNECT AND WAS STUCK AT 90%. PATIENT STATED THEY ARE GETTING SYSTEM ERROR THAT NEED TO RESTART THE PROGRAM BUT UNABLE TO RECALL SOME OF THE ERROR MESSAGE. AGENT REVIEWED DEVICE FUNCTION AND ASSISTED THE PATIENT WITH CLEARING THE DATA. PATIENT WAS ABLE TO CONNECT TO PUMP WITHOUT ANY LAG. THE TROUBLESHOOTING STEPS THAT WERE TAKEN ON THE CALL RESOLVED THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25578761 ↗
  • MalfunctionJun 19, 2026Charging Problem

    INFORMATION WAS RECEIVED FROM A CONSUMER REGARDING A PATIENT RECEIVING AN UNKNOWN DRUG VIA AN IMPLANTABLE PUMP. THE INDICATION FOR USE WAS CANCER CHEMOTHERAPY/OTHER CARCINOMA. THE PATIENT REPORTED THAT FOR THE LAST COUPLE DAYS THE COMMUNICATOR KEPT SAYING THE BATTERY WAS LOW AND NEEDED TO BE RECHARGED. THE PATIENT STATED THAT THEY KEPT PLUGGING THE CONTROLLER INTO THE AC POWER SUPPLY, BUT THAT DID NOT RESOLVE THE ISSUE. THE PATIENT THEN STATED THAT THE COMMUNICATOR WAS NOW COMPLETELY DEAD AND THEY HAVE NOT BEEN ABLE TO DELIVER A BOLUS SINCE YESTERDAY MORNING. AN EMAIL WAS SENT TO THE REPAIR DEPARTMENT FOR A REPLACEMENT COMMUNICATOR AS IT WON'T CHARGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25562487 ↗
  • InjuryJun 18, 2026Adverse Event Without Identified Device or Use Problem

    THE PATIENT'S WIFE CALLED BACK AND ASKED HOW LONG THE PATIENT WOULD HAVE WITHDRAWALS AND EXCRUCIATING PAIN. ADDITIONAL INFORMATION WAS RECEIVED FROM THE PATIENT REPRESENTATIVE STATING THAT NOTHING WAS DONE TO RESOLVE THE PATIENT'S SYMPTOMS WHEN THE PATIENT WAS IN THE HOSPITAL.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Pain

    MAUDE report 25551905 ↗
  • MalfunctionJun 17, 2026Infusion or Flow Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER'S REPRESENTATIVE (REP) REGARDING A PATIENT RECEIVING UN KNOWN MORPHINE DRUG VIA AN IMPLANTABLE INFUSION PUMP FOR UNKNOWN INDICATION FOR USE. IT WAS REPORTED THAT THE PATIENT HAD NOT BEEN GETTING PAIN RELIEF RECENTLY. IT WAS UNKNOWN IF ANY ENVIRONMENTAL, EXTERNAL, OR PATIENT FACTORS WERE CAUSAL OR CONTRIBUTORY TO THIS EVENT. DIAGNOSTICS INCLUDED A DYE STUDY, BUT THEY WERE UNABLE TO ASPIRATE AT THE CATHETER ACCESS PORT (CAP). INTERVENTIONS INCLUDE THAT THE PLAN IS TO REPLACE THE CATHETER, BUT NOTHING WAS SCHEDULED AS OF THE TIME OF THE REPORT. THE ISSUE WAS NOT RESOLVED AT THE TIME OF THE REPORT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25541120 ↗
  • MalfunctionJun 15, 2026Communication or Transmission Problem

    INFORMATION WAS RECEIVED FROM A FRIEND/FAMILY MEMBER REGARDING A PATIENT WHO WAS RECEIVING UNKNOWN MORPHINE DRUG VIA AN IMPLANTABLE PUMP FOR NON-MALIGNANT PAIN. IT WAS REPORTED THAT THE PATIENT WAS GETTING A WEIRD MESSAGE ON THE HANDSET WITHIN THE LAST WEEK THAT SAYS ALERT CONNECTION INTERRUPTED, CONNECTION TO THE PUMP WAS INTERRUPTED, MUST EXIT AND RESTART THE APP, SERVICE CODE 338. CALLER STATED THE SCREEN ALWAYS SEEMS TO GET STUCK AT 91% WHEN THEY WERE TRYING TO CONNECT WITH THE CONTROLLER, AND THEY HAD TO RESTART THE APPLICATION, AND IT TOOK A LONG TIME TO CONNECT TO THE PUMP. AGENT ASKED CLARIFYING QUESTIONS. AGENT ASSISTED CALLER WITH CLOSING OUT OF ALL RUNNING APPLICATIONS AND CLEARING THE DATA ON THE HANDSET AND DEVICE CONNECTED TO PUMP VERY FAST. AGENT REVIEWED DEVICE FUNCTION. AGENT REVIEWED LONGEVITY AND RECOMMEND PATIENT CONSULT WITH THEIR HEALTHCARE PROVIDER (HCP). THE TROUBLESHOOTING STEPS THAT WERE TAKEN ON THE CALL RESOLVED THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25512353 ↗
  • MalfunctionJun 15, 2026Disconnection

    INFORMATION WAS RECEIVED FROM A PATIENT (PT) WHO WAS RECEIVING DILAUDID (HYDROMORPHONE) VIA AN IMPLANTABLE PUMP FOR NON-MALIGNANT PAIN. IT WAS REPORTED THAT THE PT HAD BACK SURGERY, AND THAT THEIR HEALTHCARE PROVIDER (HCP) DISCONNECTED THE TUBE THAT GOES INTO THEIR SPINE TO PUT THE MEDICATION IN. PT STATED THAT THEIR PUMP IS NOW OFFLINE AND REPORTED EXPERIENCING EXTREME PAIN AFTER THE SURGERY. FOR THE EVENT DATE, PT STATED (B)(6) 2025, THEY THOUGHT.

    Patient problems: Pain

    MAUDE report 25502881 ↗
  • MalfunctionJun 12, 2026Charging Problem

    INFORMATION WAS RECEIVED FROM A CONSUMER REGARDING A PATIENT WITH AN IMPLANTABLE PUMP CONTAINING UNKNOWN DRUG FOR NON-MALIGNANT PAIN. IT WAS REPORTED THAT THE COMMUNICATOR WILL NO LONGER CHARGE. PATIENT STATED THEY WERE ABLE TO USE IT THIS MORNING BUT IT NEEDS TO BE CHARGED. PATIENT MENTIONED THEY HAVE TRIED ALL DIFFERENT CORDS AND THE DEVICE STILL WAS NOT TAKING A CHARGE. PATIENT DID NOT HAVE THE DEVICE WITH THEM AT THE TIME OF THE CALL. PATIENT WILL CALL BACK WITH COMMUNICATOR PRESENT FOR FURTHER TROUBLESHOOTING. PATIENT CALLED BACK STATING THAT THE COMMUNICATOR STILL WON'T TAKE A CHARGE. THE AGENT SENT REQUEST TO REPAIR FOR REPLACEMENT COMMUNICATOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25487409 ↗
  • MalfunctionJun 11, 2026Charging Problem

    INFORMATION WAS RECEIVED FROM A CONSUMER REGARDING A PATIENT RECEIVING FENTANYL VIA AN IMPLANTED PUMP. THE INDICATION FOR PUMP USE WAS FAILED BACK SURGERY SYNDROME AND NON-MALIGNANT PAIN. THE PATIENT REPORTED THAT THEY WERE HAVING A PROBLEM WITH THE COMMUNICATOR. THE PATIENT STATED THAT THE DEVICE WAS STARTING TO MALFUNCTION, AND THEY THOUGHT THEY NEEDED A NEW ONE. THE PATIENT SAID THAT WHEN THEY CONNECTED THE REMOTE TO THE WALL, THEY HAD TO MANEUVER IT SO IT COULD CHARGE AND IF THEY DIDN'T IT WOULD NOT CHARGE. THE AGENT ASKED, ¿IF THE PATIENT WAS NOT PLUGGING BOTH DEVICES IN AT THE SAME TIME AND PATIENT SAID NO¿. THE PATIENT CONFIRMED THEY DID GET A CONNECTION AT TIMES IF THEY WIGGLED THE CORD. THE ISSUE WAS NOT RESOLVED THROUGH TROUBLESHOOTING. A REPLACEMENT COMMUNICATOR WAS REQUESTED FROM THE REPAIR DEPARTMENT BECAUSE THE COMMUNICATOR WOULD NOT CHARGE. IT WAS NOTED THAT THE PATIENT HAD PURCHASED NEW CORDS AND IT STILL DIDN¿T WORK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25477366 ↗
  • MalfunctionJun 9, 2026Filling Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) REGARDING A PATIENT RECEIVING AN UNKNOWN DRUG VIA AN IMPLANTABLE INFUSION PUMP FOR CANCER CHEMOTHERAPY. IT WAS REPORTED THAT THE HCP WAS NOT ABLE TO ACCESS THE PATIENT'S PUMP TO REFILL IT. THE HCP STATED THEY HAD STUCK THE PATIENT 3 TIMES ALREADY. THE ISSUE WAS NOT RESOLVED THROUGH TROUBLESHOOTING, AND THE HCP WAS TRANSFERRED TO THE NATIONAL ANSWERING SERVICE (NAS) TO HAVE A LOCAL MANUFACTURER'S REPRESENTATIVE (REP) PAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25451568 ↗
  • InjuryMay 29, 2026Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) REGARDING A PATIENT RECEIVING FENTANYL AND BACLOFEN VIA AN IMPLANTABLE INFUSION PUMP FOR INTRACTABLE SPASTICITY. IT WAS REPORTED THAT THE PATIENT HAD AN ISSUE WITH THEIR PUMP WHERE IT HAD STARTED ERODING AT LEAST 6 MONTHS AGO OR SO. THEIR HCP HAD TO DO A REVISION BECAUSE IT WAS UNCOMFORTABLE AND NEVER HEALED.

    Patient problems: Discomfort; Erosion; Impaired Healing

    MAUDE report 25340464 ↗
  • MalfunctionMay 26, 2026Communication or Transmission Problem

    INFORMATION WAS RECEIVED FROM A PATIENT REGARDING AN EXTERNAL DEVICE TO AN IMPLANTABLE PUMP INFUSING AN UNKNOWN DRUG. IT WAS REPORTED THAT THE PATIENT WAS HAVING TROUBLE WITH THE COMMUNICATOR BECAUSE THEY CANNOT GET THE CHARGING CORD TO PLUG INTO THE PORT OF THE COMMUNICATOR. THE PATIENT STATED THAT THE CHARGING CORD WAS THE RIGHT SIZE AND PLUG IN BUT STILL WOULDN'T GO IN. THEY HAD BEEN THROUGH 20 DIFFERENT WAYS AND TRIED MULTIPLE CHARGING CORDS BUT THAT DIDN'T RESOLVE THE ISSUE. THE PATIENT WAS THINKING IF THERE WAS DAMAGE INSIDE THE COMMUNICATOR, BUT WHEN THEY CHECKED THE PORT, NO VISIBLE DAMAGE WAS SEEN AND THE PORT LOOKED ALIGNED. THE ISSUE HA DBEEN OCCURRING FOR 2 WEEKS; WITHIN THAT TIME THEY'VE BEEN ABLE TO GET A LITTLE BIT OF CHARGE FROM ONE OF THE CORDS, BUT THEY HAD TO PUT 'BOOKS UNDERNEATH'. THE PATIENT ALSO SAID THEY WOULD TURN THE COMMUNICATOR ON AND OFF VERY FAST TO SAVE THE BATTERY. THE AGENT SENT A REQUEST TO REPLACE THE COMMUNICATOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25284073 ↗
  • MalfunctionMay 22, 2026Communication or Transmission Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED FROM A PATIENT (CON) REGARDING AN EXTERNAL DEVICE TO AN IMPLANTABLE PUMP INFUSING UNKNOWN DRUG FOR NON-MALIGNANT PAIN. IT WAS REPORTED THE PATIENT REPORTED THE HANDSET WAS TAKING A LONG TIME TO CONNECT TO THE PUMP. THE PATIENT STATED THAT THEY NEEDED TO RUB THE COMMUNICATOR AGAINST THEIR SKIN, TO THE POINT THAT IT WOULD BECOME SORE, FOR IT TO WORK. THE AGENT REVIEWED THE DATA AND ASSISTED THE PATIENT IN DELETING IT. AFTER THAT; THE PATIENT WAS ABLE TO CONNECT TO THE PUMP WITHOUT LAG. THE AGENT ALSO REVIEWED BEST PRACTICES AND THE PROPER PLACEMENT OF THE COMMUNICATOR WHILE CONNECTING TO THE PUMP. IT WAS REPORTED THAT THEY WERE ALLOWED 4 BOLUSES A DAY.

    Patient problems: Discomfort; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25260801 ↗
  • MalfunctionMay 18, 2026Electromagnetic Interference

    ADDITIONAL INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA COMPANY REPRESENTATIVE REPORTING THAT THE PUMP DID RECOVER MORE THAN 2 HOURS AFTER IT STALLED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25204027 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P860004 on Mar 14, 1988. Submitted by Medtronic, Inc. as “Medtronic(R) Synchromed(Tm) Pump & Infusion System”. The FDA review took 770 days. The FDA has since approved 445 supplements (changes to the device or its labeling), most recently on Oct 30, 2026. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Recalls under the same FDA submission

These recalls involve devices cleared under P860004. One submission often covers many models, so check the product and lot details in each record.

  • Z-2201-2026 · Apr 8, 2026 · Open, Classified
    Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900
  • Z-1506-2025 · Jan 9, 2025 · Open, Classified
    A820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
  • Z-2352-2024 · May 10, 2024 · Open, Classified
    Medtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)
  • Z-0615-2024 · Nov 13, 2023 · Open, Classified
    Medtronic SynchroMed II, Model 8637-40, Programmable pump
  • Z-0522-2024 · Jul 13, 2020 · Completed
    Medtronic SynchroMed, Model A10
  • Z-0508-2020 · Oct 11, 2019 · Terminated
    Synchromed II programmable pump, Model numbers 8637-20 and 8637-40
  • Z-0537-2018 · Sep 21, 2017 · Terminated
    Medtronic Ascenda(R) Intrathecal Catheter, Models: (a) 8780, UDI: 00643169202122, 00643169783027 (b) 8781, UDI: 00643169202139 (c) 8784, UDI: 00643169202153 The Ascenda Intrath
  • Z-0896-2018 · Aug 9, 2017 · Open, Classified
    Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of a
  • Z-0788-2017 · Oct 3, 2016 · Terminated
    Medtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer, and Model 8637 Sy
  • Z-1271-2016 · Feb 9, 2016 · Terminated
    Medtronic, Model 8781 Ascenda" Intrathecal Catheter kit, Rx Only, Sterile Ethylene Oxide.
  • Z-1681-2015 · Apr 10, 2015 · Terminated
    Medtronic SynchroMed II Implantable Infusion Pumps, models 8637-20 and 8637-40. Sterile. Product Usage: The implantable Medtronic SynchroMed II programmable pumps are part of an
  • Z-1060-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed II (Model No: 8637) implantable drug pump The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the follow
  • Z-1061-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed implantable drug pump Refill Kits (Model No: 8551, 8555, 8561, 8562, 8564, 8565, 8566 The Model 8551, 8555, 8561, 8562, 8564, 8565, and 8566 Refill Kits ar
  • Z-1062-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed EL (Model No: 8626 and 8627) implantable drug pump The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion
  • Z-2276-2009 · Jul 1, 2009 · Terminated
    Medtronic SynchroMed II Programmable Pump, 8637, Rx Only, Sterilized by ethylene oxide gas, Medtronic inc., Minneapolis, MN 55432-5604. (The SynchroMed II Pump is supplied in 20 m

…and 24 more.

Common questions about the SYNCHROMED®II 8637-20

Has the SYNCHROMED®II 8637-20 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0614-2024 (Class II), began on Nov 13, 2023: Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the SYNCHROMED®II 8637-20?

The FDA's MAUDE database lists 1,613 adverse event reports for this device, received between Jul 6, 2021 and Sep 29, 2026. In those reports, the most common specific device problems are “Communication or Transmission Problem” (225), “Infusion or Flow Problem” (217) and “Migration or Expulsion of Device” (207); the most common patient problems are “Pain” (235), “Unspecified Infection” (151) and “Cerebrospinal Fluid Leakage” (48). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SYNCHROMED®II 8637-20?

2 reports are classified as deaths, 984 as injuries and 627 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SYNCHROMED®II 8637-20 FDA approved?

It was approved through premarket approval (PMA) P860004 on Mar 14, 1988.

Is the SYNCHROMED®II 8637-20 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the SYNCHROMED®II 8637-20, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand SYNCHROMED®II, model 8637-20). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Pump, infusion, implanted, programmable devices

All Pump, infusion, implanted, programmable devices →