Skip to content

MIDAS REX MR8 EM800 (UDI 00763000430917)Recalls & reported problems

Medtronic Powered Surgical Solutions · MOTOR EM800 MR8 ELECTRIC · Class 2

Quick summary

This Medtronic Midas Rex surgical drill motor is an electric motor for surgical drills. Its 186 reports include no deaths and 3 injuries. Most describe the motor overheating, while others mention rust, corrosion, or bearings coming loose or being damaged. Reports in the last 12 months (22) are down from 120 in the prior 12 months, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
186
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 186 adverse event reports that list this exact device (DI 00763000430917), received between Jun 13, 2023 and Apr 28, 2026 — classified as 3 injuries, 183 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (150 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 3 reportsMar 2024: 0 reportsApr 2024: 5 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 5 reportsSep 2024: 19 reportsOct 2024: 6 reportsNov 2024: 8 reportsDec 2024: 6 reportsJan 2025: 8 reportsFeb 2025: 9 reportsMar 2025: 28 reportsApr 2025: 3 reportsMay 2025: 9 reportsJun 2025: 3 reportsJul 2025: 18 reportsAug 2025: 8 reportsSep 2025: 10 reportsOct 2025: 10 reportsNov 2025: 10 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 5 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202628
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device150
  • No Apparent Adverse Event15
  • Corroded10
  • Detachment of Device or Device Component9
  • Noise, Audible7
  • Device Displays Incorrect Message5
  • Difficult to Insert5
  • Difficult to Remove5

Patient problems

  • No Clinical Signs, Symptoms or Conditions176
  • Insufficient Information14
  • Burn(s)4

Recent adverse event reports

The 25 most recent of 186 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 28, 2026Device Displays Incorrect Message; Material Integrity Problem

    IT WAS REPORTED THAT THE DEVICE HAS ERROR CODE DISPLAYED AND CONNECTOR BROKEN ISSUE. THERE WAS PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION RECEIVED THAT THERE WAS NO PATIENT IMPACT REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25010839 ↗
  • MalfunctionApr 16, 2026Corroded; Loss of Power

    IT WAS REPORTED THAT THE MOTOR WAS NOT WORKING PROPERLY AND EXHIBITED RUST AND CORROSION. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24894132 ↗
  • MalfunctionApr 6, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING USE THE BALL BEARING CAME OFF. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24789553 ↗
  • MalfunctionMar 30, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24710000 ↗
  • MalfunctionMar 17, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE MOTOR IS NOT WORKING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DAMAGED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24616217 ↗
  • MalfunctionMar 13, 2026Loss of Power

    IT WAS REPORTED THAT THE MOTOR WAS NOT WORKING PROPERLY. IT WAS REPORTED THAT THERE IS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24598280 ↗
  • MalfunctionMar 10, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE MOTOR WAS NOT WORKING PROPERLY. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING WAS DAMAGED DUE TO CORROSION INSIDE THE EQUIPMENT FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24555221 ↗
  • MalfunctionMar 5, 2026Difficult to Remove

    IT WAS REPORTED THAT THE DEVICE WOULD NOT DETACH FROM ITS ATTACHMENT. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24515663 ↗
  • MalfunctionFeb 26, 2026Loss of Power

    IT WAS REPORTED THAT THE MOTOR WAS NOT WORKING PROPERLY. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DAMAGED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24451398 ↗
  • MalfunctionFeb 25, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE DEVICE WAS NOT WORKING PROPERLY. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24438796 ↗
  • MalfunctionFeb 10, 2026Difficult to Remove

    IT WAS REPORTED THAT BUR WON'T BE REMOVED. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24305198 ↗
  • MalfunctionDec 3, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT MOTOR WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23712119 ↗
  • MalfunctionNov 25, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE MOTOR WAS OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23642531 ↗
  • MalfunctionNov 25, 2025Overheating of Device; Physical Resistance/Sticking; Overheating of Device; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE PRODUCT WAS INSTALLED IN SEPTEMBER BUT FACED BOTH HEATING AND LOCKING ISSUE. REPAIRED MOTOR CAME BUT STILL IT IS FACING HEATING ISSUES. THERE WAS NO PATIENT INVOLVED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23642253 ↗
  • MalfunctionNov 19, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE MOTOR WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23601262 ↗
  • MalfunctionNov 19, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE WAS OVERHEATING. IT WAS REPORTED THAT THERE IS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23591837 ↗
  • MalfunctionNov 18, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE MOTOR WAS OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23580411 ↗
  • MalfunctionNov 12, 2025Overheating of Device; Noise, Audible; Overheating of Device; Noise, Audible

    IT WAS REPORTED THAT DURING A PROCEDURE, THE MOTOR WAS OBSERVED TO OVERHEAT AND MAKE EXCESSIVE NOISE DURING USE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23533223 ↗
  • MalfunctionNov 11, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE MOTOR WAS OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23519563 ↗
  • MalfunctionNov 10, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING PRIOR TO USE THE MOTOR EXPERIENCED OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23508045 ↗
  • MalfunctionNov 7, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT PRIOR TO USE, THE MOTOR EXHIBITED HEATING DURING PRE-OPERATIVE PREPARATION. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23494794 ↗
  • MalfunctionNov 3, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE MOTOR WAS OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23448010 ↗
  • MalfunctionOct 29, 2025Overheating of Device; Device Displays Incorrect Message; Overheating of Device; Device Displays Incorrect Message

    IT WAS REPORTED THAT AS SOON AS IT STARTED TURNING ON, THE MOTOR STARTED TO OVERHEAT AND GIVES AN ERROR. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23413217 ↗
  • MalfunctionOct 27, 2025Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. REPAIR REQUEST WAS ESCALATED TO A PRODUCT EVENT BASED ON REASON FOR RETURN AND RETURN OF THE PART IN LESS THAN 90 DAYS FROM THE PURCHASE OR REPAIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23388981 ↗
  • MalfunctionOct 27, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING PRE-OPERATIVE PREPARATION FOR USE, THE MOTOR OF THE PRODUCT WAS GETTING HOT. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23388523 ↗

Other MOTOR, DRILL, ELECTRIC devices

All MOTOR, DRILL, ELECTRIC devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.