Skip to content

NA CRANI-A-G1Recalls & reported problems

Synthes GmbH · 15 mm Craniotome Attachment, G1 · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Synthes craniotome attachment is a drill attachment used to cut bone in skull surgery. The 358 reports mostly describe overheating, parts coming loose and bent tips. The reports record 2 injuries, 356 malfunctions and no deaths. Reports fell from 33 in the year before to 9 in the last year. 1 open recall, started in December 2022, covers ball bearings that came out of the attachment during surgery.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
358
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 358 adverse event reports that list this exact device (DI 00845384016410), received between Nov 28, 2017 and Jun 15, 2026 — classified as 2 injuries, 356 malfunctions.

The most frequently reported specific device problem is “Device Damaged by Another Device” (180 reports).

1 FDA recall record name this device's identifier, 1 still open.

10 other recalls cover devices cleared under the same FDA submission (K133604); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 7 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 5 reportsJul 2024: 4 reportsAug 2024: 4 reportsSep 2024: 1 reportsOct 2024: 7 reportsNov 2024: 5 reportsDec 2024: 2 reportsJan 2025: 5 reportsFeb 2025: 6 reportsMar 2025: 3 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 3 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 3 reportsDec 2025: 4 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 7, 2022Open, Classified

    Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

    Z-0918-2023 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Device Damaged by Another Device180
  • Excessive Heating63
  • Detachment of Device or Device Component45
  • Material Twisted/Bent39
  • Vibration26
  • Break24
  • Appropriate Device Problem Term/Code Not Available22
  • Complete Loss of Power22

Patient problems

  • No Clinical Signs, Symptoms or Conditions227
  • No Consequences Or Impact To Patient115
  • No Known Impact Or Consequence To Patient13
  • Foreign Body In Patient2
  • Hemorrhage/Blood Loss/Bleeding1
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 358 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 15, 2026Device Damaged by Another Device; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE NEURO-TIP OF THE CRANIOTOME DEVICE WAS BENT/DAMAGED. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT PROPERLY CONNECT TO THE HANDPIECE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2026. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25506094 ↗
  • MalfunctionJan 22, 2026Device Damaged by Another Device

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE NEURO-TIP OF THE CRANIOTOME DEVICE WAS BENT/DAMAGED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED; HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24145529 ↗
  • MalfunctionDec 22, 2025Break; Device Damaged by Another Device; Difficult or Delayed Separation; Break; Device Damaged by Another Device; Difficult or Delayed Separation

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CRANIOTOME DEVICE NEURO TIP WAS BENT/DAMAGED. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS DAMAGED WHEN IT WAS APPLIED WITH A LITTLE FORCE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23866728 ↗
  • MalfunctionDec 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE MOTOR DEVICE OVERHEATED. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THE EXACT DATE OF THIS EVENT WAS UNKNOWN BUT WAS NOTED TO HAVE OCCURRED IN 2025. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23862330 ↗
  • MalfunctionDec 17, 2025Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CRANIOTOME DEVICE NEURO-TIP WAS BENT/DAMAGED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23828798 ↗
  • MalfunctionDec 12, 2025Device Damaged by Another Device; Noise, Audible; Device Damaged by Another Device; Noise, Audible

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CRANIOTOME DEVICE NEURO TIP WAS BENT/DAMAGED. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE HAD ABNORMAL NOISE WHEN CUTTING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23790755 ↗
  • MalfunctionNov 19, 2025Device Damaged by Another Device; Complete Loss of Power; Excessive Heating; Device Damaged by Another Device; Complete Loss of Power; Excessive Heating

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE NEURO-TIP OF THE CRANIOTOME DEVICE WAS BENT/DAMAGED AND THE DEVICE PRODUCED HEAT. IT WAS FURTHER DETERMINED THAT THE DEVICE HAD VIBRATION. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT, TEMPERATURE VERIFICATION AND VIBRATION ASSESSMENT. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT FUNCTION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23597237 ↗
  • MalfunctionNov 17, 2025Break; Break

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE NEURO-TIP ON THE ADULT CRANIOTOME ATTACHMENT DEVICE WAS BENT/DAMAGED. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE HAD INNER BEARING DAMAGE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23575535 ↗
  • MalfunctionNov 10, 2025Device Damaged by Another Device; Device-Device Incompatibility; Device Damaged by Another Device; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CRANIOTOME DEVICE EXHIBITED VIBRATION AND NEURO TIP WAS BENT/DAMAGED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT AND VIBRATION ASSESSMENT. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS UNABLE TO CONNECT TO THE GREEN CUTTER DEVICE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23508674 ↗
  • MalfunctionOct 20, 2025Device Damaged by Another Device; Difficult or Delayed Separation; Device Damaged by Another Device; Difficult or Delayed Separation

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CRANIOTOME DEVICE HAD BEARING DAMAGE, THE COLOR BAND WAS CHIPPING/FADING, THE NEURO TIP WAS BENT/DAMAGED, AND THE CUTTER COULD NOT BE INSERTED. IT WAS NOTED THAT THE DURA GUARD HAD SCRATCHES, THE COLOR LABEL WAS MISSING, BURR INSERTION WAS IMPOSSIBLE (BEARING MISSING). IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT, AND CUTTER INSERTION. IT WAS NOTED IN THE SERVICE ORDER THAT THE ATTACHMENT DEVICE WAS DIFFICULT TO REMOVE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. AL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23335524 ↗
  • MalfunctionOct 17, 2025Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRE-TESTING, IT WAS DISCOVERED THAT THE NEURO-TIP OF THE CRANIOTOME DEVICE WAS BENT/DAMAGED. IT WAS FURTHER DETERMINED THAT THE DEVICE MET THE VISUAL ASSESSMENT CRITERIA. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23328450 ↗
  • MalfunctionAug 28, 2025Device Damaged by Another Device; Complete Loss of Power; Device Damaged by Another Device; Complete Loss of Power

    IT WAS REPORTED THAT THE CRANIOTOME DEVICE WAS NOT WORKING. DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE DEVICE CRANIOTOME NEURO TIP WAS FRACTURED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22911509 ↗
  • MalfunctionAug 15, 2025Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING SERVICE AND REPAIR/PRE-TESTING, IT WAS OBSERVED THAT THE CRANIOTOME DEVICE HEATED UP. THE DEVICE ALSO FAILED PRETEST FOR TEMPERATURE VERIFICATION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22814293 ↗
  • MalfunctionJul 3, 2025Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT TWO ADULT CRANIOTOME DEVICES, ONE SHORT DEVICE AND ONE MEDIUM ATTACHMENT DEVICE HAD OVERHEATING ISSUES. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPOT 2 OF 4 OF (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22393999 ↗
  • MalfunctionJul 3, 2025Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT TWO ADULT CRANIOTOME DEVICES, ONE SHORT DEVICE AND ONE MEDIUM ATTACHMENT DEVICE HAD OVERHEATING ISSUES. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 1 OF 4 OF (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22393993 ↗
  • MalfunctionJun 26, 2025Appropriate Device Problem Term/Code Not Available; Excessive Heating; Appropriate Device Problem Term/Code Not Available; Excessive Heating

    IT WAS REPORTED THAT THE ADULT CRANITOME DEVICE HAD AN UNDETERMINED MALFUNCTION. DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE DEVICE PRODUCED HEAT, HAD COMPONENT DAMAGE AND THE LOCKING MECHANISM WAS STIFF/STUCK. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT AND LOCK OPERATION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22336473 ↗
  • MalfunctionJun 10, 2025Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE ADULT CRANIOTOME DEVICE WAS BENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22174296 ↗
  • MalfunctionJun 4, 2025Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRE-TESTING, IT WAS DISCOVERED THAT THE CRANIOTOME DEVICE GENERATED HEAT. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR TEMPERATURE VERIFICATIONS. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22143036 ↗
  • MalfunctionMay 20, 2025Complete Loss of Power; Complete Loss of Power

    IT WAS REPORTED THAT THE CRANIOTOME DEVICE WAS NOT WORKING. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE COULD NOT INSERT THE CUTTER AND THE CRANIOTOME NEURO-TIP WAS BENT/DAMAGED. THE DEVICE ALSO FAILED PRETESTS FOR VISUAL ASSESSMENT AND CUTTER INSERTION. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22055690 ↗
  • MalfunctionMay 14, 2025Break; Excessive Heating; Break; Excessive Heating

    IT WAS REPORTED THAT THE CRANIOTOME DEVICE JOINT ASSEMBLY WAS HOT (THE INNER BEARING IS DAMAGED). IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THE EXACT DATE OF THIS EVENT WAS UNKNOWN BUT WAS NOTED TO HAVE OCCURRED IN 2025. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22021337 ↗
  • MalfunctionApr 30, 2025Overheating of Device; Noise, Audible; Overheating of Device; Noise, Audible

    IT WAS REPORTED THAT THE CRANIOTOME DEVICE HAD EXCESSIVE NOISE AND HEAT. DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD EXCESSIVE NOISE, HEAT AND FAILED PRE-TEST FOR TEMPERATURE VERIFICATION AND VIBRATION ASSESSMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21935093 ↗
  • MalfunctionMar 24, 2025Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CRANIOTOME DEVICE HAD HEAT AND FAILED PRE-TEST FOR TEMPERATURE VERIFICATION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21677016 ↗
  • MalfunctionMar 14, 2025Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE NEURO-TIP OF THE CRANIOTOME DEVICE WAS BENT/DAMAGED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21606240 ↗
  • MalfunctionMar 13, 2025Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ADULT CRANIOTOME DEVICE COULD NOT ENGAGE THE ATTACHMENT AND THE NEURO TIP WAS BENT/DAMAGED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETESTS FOR VISUAL ASSESSMENT AND LOCK OPERATION. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS NOT ABLE TO LOCK WITH THE HANDPIECE DEVICE AND CUTTER DEVICES.. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21597583 ↗
  • MalfunctionFeb 25, 2025Device Damaged by Another Device; Complete Loss of Power; Device Damaged by Another Device; Complete Loss of Power

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ADULT CRANIOTOME DEVICE NEURO TIP WAS BENT/DAMAGED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT. IT WAS NOTED IN THE SERVICE ORDER THAT DURING PRE-SURGERY, IT WAS OBSERVED THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21457600 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133604. One submission often covers many models, so check the product and lot details in each record.

  • Z-0922-2023 · Dec 7, 2022 · Open, Classified
    7.5 cm Large Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-L-R Intended for cutting and shaping bone including the spine and cranium
  • Z-0919-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-R intended for cutting and shaping bone including the spine and cranium
  • Z-0923-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Pediatric Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-P Intended for cutting and shaping bone including the spine and cranium
  • Z-0924-2023 · Dec 7, 2022 · Open, Classified
    Pediatric Craniotome Ref:CRANI-P-G1
  • Z-0917-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Craniotome, Thin Foot Plate Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-01 intended for cutting and shaping bone including the spine and crani
  • Z-0921-2023 · Dec 7, 2022 · Open, Classified
    Adult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including the spine and cranium
  • Z-0916-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems. Ref: CRANI-A Intended for cutting and shaping bone including the spine and cranium
  • Z-0920-2023 · Dec 7, 2022 · Open, Classified
    7.5 cm Large Craniotome Attachment Use with XMax, microMax, and eMax Systems. Ref: CRANI-L Intended for cutting and shaping bone including the spine and cranium
  • Z-2253-2016 · Jun 15, 2016 · Terminated
    Anspach G1 High Speed Electric Handpiece Air Cooled (eG1A) Use for cutting and shaping bone including bones of the cranium and spine.
  • Z-1037-2015 · Dec 10, 2014 · Terminated
    16cm Minimally Invasive Attachment. For use with Motor Systems for cutting and shaping bones, including the spine and cranium. Model MIA16-G1

Other MOTOR, DRILL, ELECTRIC devices

All MOTOR, DRILL, ELECTRIC devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.