MIDAS REX MR8 PM800 (UDI 00763000431013)Recalls & reported problems
Medtronic Powered Surgical Solutions · MOTOR PM800 MR8 PNEUMATIC · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00763000431013), received between Jun 9, 2025 and Jun 24, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Material Integrity Problem2
- No Apparent Adverse Event2
- Overheating of Device2
- Fluid/Blood Leak1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Insufficient Information1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 24, 2026Fluid/Blood Leak
IT WAS REPORTED THAT THE PRIOR TO THE PROCEDURE, LEAK WAS IDENTIFIED IN MOTOR. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25605817 ↗MalfunctionJun 11, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING POST OP HOSE DETACHED FROM THE MOTOR. IT WAS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT OR COMPLICATIONS AS A RESULT OF THE EVENT.ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25470341 ↗MalfunctionJun 9, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE DEVICE WAS NOT WORKING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT HOSE DETACHED WERE FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25439022 ↗MalfunctionJun 8, 2026Material Integrity Problem
IT WAS REPORTED THAT THE MOTOR CABLE WAS FOUND TO BE BROKEN PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25430181 ↗MalfunctionMay 29, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE DEVICE HOSE IS ALMOST COMPLETELY RIPPED OFF OF THE PLUG FOR THE MACHINE. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333476 ↗MalfunctionNov 10, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT WHILE THE MOTOR WAS BEING USED DURING A PROCEDURE, THE HOSE RUPTURED AND A LOUD EXPLOSION OCCURRED. THIS RESULTED IN AN AIR LEAK AND A 30-MINUTE DELAY IN THE PROCEDURE. BROKEN PIECES OF THE PRODUCT REMAINED INSIDE THE PATIENT'S BODY AND WERE SUBSEQUENTLY RETRIEVED. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23507840 ↗MalfunctionAug 21, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED.
Patient problems: Insufficient Information
MAUDE report 22867352 ↗MalfunctionAug 13, 2025Material Integrity Problem; Material Integrity Problem
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF CABLE DAMAGED.IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22792764 ↗MalfunctionJul 3, 2025Overheating of Device; Overheating of Device
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING.IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22398230 ↗MalfunctionJun 9, 2025No Apparent Adverse Event; No Apparent Adverse Event
REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF NOT A COMPLAINT. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. REPAIR IS BEING ESCALATED TO PRODUCT EVENT DUE TO REASON FOR THE RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22169263 ↗
Other MOTOR, DRILL, PNEUMATIC devices
- NA 05.001.080Synthes GmbH · MOTOR, DRILL, PNEUMATIC432 reports
- NA CSR60Synthes GmbH · MOTOR, DRILL, PNEUMATIC221 reports
- Midas Rex® AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC163 reports
- Maestro Air 5407-200-000Stryker Corporation · MOTOR, DRILL, PNEUMATIC112 reports
- Maestro 5400210060Stryker Corporation · MOTOR, DRILL, PNEUMATIC95 reports
- MIDAS REX™ AF02Medtronic Powered Surgical Solutions · MOTOR, DRILL, PNEUMATIC91 reports
- CORE, Maestro 5400200000Stryker Corporation · MOTOR, DRILL, PNEUMATIC75 reports
- Maestro Air 5407-200-00HStryker Corporation · MOTOR, DRILL, PNEUMATIC56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page